Self-Declaration for Class I Medical Devices and Uploading Documents to ÜTS
Self-declaration for Class I medical devices: which devices stay with self-declaration, the EU declaration of conformity fields, the documents ÜTS asks for, and the difference between a domestic manufacturer and an importer.
You have a Class I device, you have no notified body certificate, and you open the ÜTS record thinking "since I have no certificate, I have nothing to upload". Then the system asks you for a chain of documents. It is true that Class I is the "easy class", but this ease does not mean no documents; it is a route in which the whole of the responsibility stays with the manufacturer, while ÜTS checks the declaration to the letter.
This page explains the self-declaration route for Class I: which Class I device can really be placed on the market on its own declaration, which ones bring a notified body into the process for a limited subject, what ÜTS expects for this class, and which fields the EU declaration of conformity must carry. The sources are the Manual on Registration Procedures of Medical Devices in the Product Tracking System (Tıbbi Cihazların Ürün Takip Sistemine Kayıt Süreçlerine İlişkin Kılavuz) and the Medical Device Regulation (Tıbbi Cihaz Yönetmeliği) published in the Official Gazette No. 31499 (bis) of 2/6/2021.
Class I means not the absence of a certificate, but a declaration that takes the certificate's place
Article 52(6) of the Medical Device Regulation states that, for Class I devices other than custom-made or investigational devices, manufacturers declare the conformity of their products by issuing the EU declaration of conformity referred to in Article 19 after drawing up the technical documentation set out in Annexes II and III. The order is clear: first the technical documentation, then the declaration.
Article 61(4) of the Manual describes the same route in ÜTS language. Manufacturers of Class I devices draw up technical documentation as set out in MDR Annexes II and III in order to be able to place their devices on the market or put them into service and to affix the CE marking under the MDR; after the technical documentation has been prepared, the manufacturer declares the conformity of its devices by drawing up a declaration of conformity as referred to in MDR Annex IV and in Article 11 of the Manual.
The legal weight of the declaration is set by Article 19(3) of the Medical Device Regulation: by drawing up the EU declaration of conformity, the manufacturer assumes responsibility for compliance with the requirements of this Regulation and all other legislation applicable to the device. Paragraph 1 of the same article requires the declaration to be kept up to date and to contain, as a minimum, the information set out in Annex IV; for devices placed on the domestic market, if the declaration is drawn up in another language, a Turkish translation is to be attached.
So in Class I there is no third-party audit; the manufacturer's signature takes the place of the audit. This is why ÜTS review looks so strictly at the content of the declaration in this class.
Which Class I devices really stay on self-declaration
Class I is not a single block. Article 52(6) of the Medical Device Regulation states that, if devices are placed on the market in sterile condition, have a measuring function, or are reusable surgical instruments, the manufacturer will apply the procedures set out in Chapters I and III of Annex IX or in Part A of Annex XI. These three cases bring the notified body into the process.
However, the body's role is limited, and the limit is written in the same paragraph:
Devices placed on the market in sterile condition: Point (a) of the same paragraph states that the notified body's involvement is limited to the aspects relating to establishing, securing and maintaining sterile conditions. Devices with a measuring function: Point (b) states that the involvement is limited to the aspects relating to the conformity of the devices with the metrological requirements. Reusable surgical instruments: Point (c) states that the involvement is limited to the aspects relating to the reuse of the device, in particular cleaning, disinfection, sterilisation, maintenance, functional testing and the related instructions for use.
Article 61(4) of the Manual matches these three cases with the class names in ÜTS: Class Is if the device is supplied sterile, Class Im if it has a measuring function, Class Ir if it is a reusable surgical instrument. Points (a), (b) and (c) of the same paragraph repeat that the notified body's involvement is limited to the same three subjects.
The remaining group is the group selected in ÜTS as "Class I Other (not sterile and without a measuring function)" and is the only group that really proceeds on the self-declaration route. Article 27(4)(a) of the Manual lists these four options explicitly and states that the "Class" field is a mandatory field in ÜTS.
A limited-scope certificate cannot be used like a full-scope certificate
If you hold an EU certificate for your Class Is, Class Im or Class Ir device, that certificate does not document the device's conformity with the general safety and performance requirements; it documents only one of the three limited subjects above. The document declaring the conformity of the device is still the EU declaration of conformity. This distinction causes problems when the certificate is uploaded to ÜTS without reading its scope text.
The fields the EU declaration of conformity must contain
Article 11(2) of the Manual lists, one by one, the fields that must be in the EU declaration of conformity and that are checked during document review in ÜTS; the same paragraph also requires that there be no difference between the fields as entered into ÜTS and as they appear in the document. The ones critical for Class I are the following:
Manufacturer and authorised representative information: Point (a) looks for the manufacturer's name or trade name and address, and point (b), where applicable, for the name or trade name and address of the EU Authorised Representative.
Statement of responsibility: Point (ç) requires the declaration to contain a statement that the EU declaration of conformity is issued under the sole responsibility of the manufacturer. In Class I this sentence is the backbone of the declaration.
Basic UDI-DI: Point (d) looks for the Basic UDI-DI information.
Identification of the device: Point (e) asks for the intended purpose, together with the product name and trade name, product code, catalogue number or other unambiguous reference allowing identification and traceability of the device covered by the declaration, such as a photograph where applicable.
Risk class: Point (g) requires the risk class of the device to be in the declaration.
Reference to the regulation: Point (ğ) looks for a statement that the device is in conformity with the Medical Device Regulation to which it belongs and, if applicable, with any other Union legislation that provides for the issuing of an EU declaration of conformity.
Dates: Point (j) asks for the document start (issue) date and point (k) for the document end (validity) date. Sub-point 1 of the same point states that if the document is of unlimited duration, "Unlimited Duration" ("Sınırsız Süreli") is selected on the ÜTS screens.
Signature: Point (m) looks for the name and surname of the authorised person together with their position and signature/stamp information; the same point states that an electronic signature is also valid.
The truly distinguishing provision for Class I is point (ı) of the same paragraph: for declarations other than those covering only Class I Other or Class A devices, the name and identification number of the notified body, a description of the conformity assessment procedure performed and the identification of the EU certificate or certificates issued are required. This point reads as follows: notified body information is not expected only in a Class I Other declaration. In Class Is, Class Im and Class Ir declarations, the notified body information must be in the declaration.
Article 11(3) of the Manual states that the EU Declaration of Conformity is drawn up by the legal manufacturer firm. Paragraph 6 of the same article requires the declaration to be uploaded in full, with all its annexes and additional pages. Paragraph 7 states that EU Declaration of Conformity templates, created especially to serve as examples for domestic manufacturer firms, are in Annex-7 and Annex-8 of the Manual.
What ÜTS asks of a Class I device
Article 66(1) of the Manual states that, for a device registration notification under the MDR, according to the type and risk class of the device, the appropriate one or ones of the documents mentioned in Articles 7, 11, 17, 20, 22, 23 and 24 of the Manual must be linked to the device. Paragraph 3 of the same article states that, for devices to be notified as "Class I Other", "Class Is", "Class Im" or "Class Ir", the appropriate one of the document business rules shown in Annex-88 of the Manual will be applied. The four Class I variants share the same set of business rules; the variant determines which rule applies.
The document to be linked must also be ready. Article 66(1) of the Manual requires the document to be in "Registered" status in ÜTS as a result of the application made under Article 39 of the Manual. Article 27(5)(a)(3) of the Manual ties the order: before document selection, the class of the relevant device must have been selected on the ÜTS registration screen, and the selection is made from among the documents that can be selected according to the class of the device.
On the instructions for use side, Class I has a privilege, but it is not absolute. Article 17(3) of the Manual states that uploading instructions for use is optional for Class I Other, Class Im, Class Is, Class Ir, Class A, Class A Sterile and IVD Other devices. The same paragraph goes on to state that the Agency may request the instructions for use of devices in this risk group to be uploaded where it deems necessary, for example if a doubt arises about product safety, or if the original label of the device refers to the instructions for use and some information needs to be verified from them.
The document upload route differs for domestic manufacturers and importers
A significant part of Class I products are registered by domestic manufacturers, and the Manual defines a separate route for these firms. Article 13(1)(a) of the Manual states that domestic manufacturer firms must apply by selecting "E-Signed Document Application for Manufacturer Firms" ("İmalatçı Firmalar için e-imzalı Belge Başvurusu") as the Document Application Type in their notifications of the EU Declaration of Conformity for their own products. Point (c) of the same paragraph states that documents whose original language is Turkish are accepted.
Point (ç) of the same paragraph states that in the "Original Document" field, the original of the document electronically signed by the person registered as "Responsible Manager" ("Mesul Müdür") or "Signatory" ("İmza Yetkilisi") on behalf of the relevant domestic manufacturer firm is uploaded in full. Sub-point 1 of the same point states that ÜTS allows the application only when the document has been electronically signed by one of these two authorities; sub-point 2 states that if the name and surname of the signing authority appear in writing on the document itself, this name and surname must match the name and surname of the authority who electronically signed the document.
On the importer side the route is completely different. Article 12(1) of the Manual states that importer firms, in notifications of declarations of conformity belonging to foreign manufacturers, upload to the "Original Document" field the apostille-certified version of the document, the version certified by the Republic of Türkiye embassy or consulate in the manufacturer's country, the original of a verifiable e-Apostille-certified document, or the original of a verifiable e-signed document. Sub-point 2 of point (a) of the same paragraph states that the original of the apostille-certified document must be uploaded and that there is no need to upload a copy issued by a Turkish notary or by a notary of another country.
Article 12(2) of the Manual states that in the "Turkish Document" ("Türkçe Doküman") field the Turkish translation of the original document, made by a sworn translator, is uploaded in full; point (b) of the same paragraph requires the translation to carry the translator's name and surname, stamp, signature or seal and a statement that the translation is true to the original.
The points that most often break an application in this class
In Class I, most rejection and revision decisions arise not from the existence of a document but from the difference between the information in the document and the information entered into ÜTS. The matters directly tied to sanctions in the source are these:
Barcode entered differently from the document: Article 27(2)(b)(3) of the Manual states that the barcode allocated under the standards the device holds must be entered into ÜTS exactly, without any change. Sub-point 4 of the same point states that a device application made under Article 40 of the Manual for a device record whose barcode has not been entered into ÜTS as it appears in its technical documents (label/sales packaging, EU declaration of conformity, EU certificate, etc.) will end with a negative result. Entering the carton barcode instead of the smallest package: Sub-point 6 of the same point states that device registration in ÜTS is made mainly with the barcode of the smallest package placed on the market, and that the barcodes of sales packages with more than one product are added, together with their content quantities, to the "Market Placement Information" ("Piyasaya Arz Bilgisi") field after registration is complete.
Leaving the brand information empty: Article 27(2)(e) of the Manual states that the "Brand Name" field is a mandatory field and that, for the sake of product safety, registration in ÜTS is not carried out for devices for which no brand is declared.
GMDN code contradicting the product characteristics: Article 27(4)(b)(5) of the Manual states that consistency is mandatory between the information in the description of the GMDN code to which the device is matched and product characteristics such as whether the device is single-use and/or sterile. Sub-point 6 of the same point states that the GMDN code description of a single-use device contains the expression single-use.
Selecting the EMDN code from a higher level: Article 27(4)(ç)(9) of the Manual states that only the most detailed EMDN code and terminal terms are added to ÜTS; less detailed codes are not added.
Sterile fields contradicting the device: Article 27(7)(r) of the Manual states that selecting the "Was it packaged sterile?" ("Steril paketlendi mi?") field is mandatory; sub-point 4 of the same point states that if the device was packaged sterile, this information must be on the label and/or sales packaging. If these fields are incompatible with a Class Is selection, the class itself becomes disputable.
Incomplete upload of the declaration: Article 11(6) of the Manual requires the EU Declaration of Conformity to be uploaded in full with all its annexes and additional pages.
Article 11(4) of the Manual points to a separate risk: if the manufacturer firm has the devices covered by the declaration produced by another manufacturer, the title and address information of that manufacturer is also expected to be in the EU declaration of conformity or EU certificate. If this information is not on the declaration, certificate, label or sales packaging, the Agency, where it deems necessary, requests that the technical file pages containing this information be uploaded to the "Technical Document of the Product(s) in the Document" field.
What is uploaded if the Agency asks for technical documentation
In Class I the whole technical file is not uploaded to ÜTS; what is asked for is a summary. Article 11(5) of the Manual lists the information to be uploaded to the "Technical Document of the Product(s) in the Document" ("Belgedeki Ürün(ler)e Ait Teknik Doküman") field where deemed necessary for any of the devices in the declaration of conformity: the name or trade name and general description of the device, the manufacturer's name and address, the UDI-DI number and the EU certificate number if any, the intended purpose, place of use, manner of contact, duration of use, the population to be used on, the mechanism of action, the class and classification rule, the device content and percentage composition information according to the form of the device, label information and the instructions for use.
The rest of the same paragraph states two things: these technical documents must be in the device's technical file, and, where deemed necessary, the entire technical file is kept ready to be delivered to the Agency. That in Class I "the notified body does not look" does not mean there is no technical file.
The obligation is also clear on the Regulation side. Article 10(4) of the Medical Device Regulation states that manufacturers of devices other than custom-made devices draw up and keep up to date technical documentation, and that this documentation contains the matters set out in Annexes II and III. Article 10(8)(a) of the same article requires the technical documentation and the EU declaration of conformity to be kept for at least 10 years after the last device covered by the declaration has been placed on the market, and to be made available to the Agency on request.
In Class I no obligation ends when registration ends
A Class I manufacturer has obligations that continue after the ÜTS registration too. Article 10(9) of the Medical Device Regulation states that, except for investigational devices, manufacturers of devices establish, document, implement and maintain a quality management system in a way that is proportionate to the risk class and the type of device. This obligation does not leave Class I out; it only ties it to the measure of "proportionate".
On the post-market side, Article 83(1) of the Medical Device Regulation defines a special report for Class I: manufacturers of Class I devices prepare a post-market surveillance report summarising the results and conclusions of the analyses of the data gathered as a result of the post-market surveillance plan referred to in Article 82, together with a rationale and description of any preventive and corrective actions taken. This report is updated when necessary and is made available to the Agency upon request.
There is also a Class I specific distinction on the CE marking side. Article 20(5) of the Medical Device Regulation states that, where a notified body is responsible for the conformity assessment procedures set out in Article 52, the CE marking will be followed by the identification number of the notified body. Because there is no such body for Class I Other devices, a four-digit number is not looked for next to the CE marking; for Class Is, Class Im and Class Ir devices it is. Whether this number is on the label must be consistent with the class variant you select in ÜTS.
What we do in a Class I file, and what we cannot do
Within medical device ÜTS registration consultancy, our concrete output on the Class I side is this:
Variant decision file: We verify, from the label, sales packaging and instructions for use, whether the device is Class I Other, Class Is, Class Im or Class Ir, and write, with reasons, which of the options in Article 27(4)(a) of the Manual is to be selected. Declaration field check: We check the EU declaration of conformity received from the manufacturer field by field against the list of points in Article 11(2) of the Manual, and report missing fields to the manufacturer together with the relevant article. Setting up the document upload route: Before the application, we prepare the e-signed document application route in Article 13 of the Manual for a domestic manufacturer, and the certification and sworn translation conditions in Article 12 of the Manual for an importer. Technical document summary: We compile the thirteen headings listed in Article 11(5) of the Manual so that they come out of the technical file. Field consistency scan: Before the application, we compare the barcode, brand, GMDN, EMDN and sterile fields with the document.
What we cannot do is clear too. We cannot draw up the EU declaration of conformity in the manufacturer's place; Article 11(3) of the Manual ties the declaration to the legal manufacturer firm, and Article 19(3) of the Medical Device Regulation places the responsibility on the manufacturer. We cannot produce the technical documentation, the risk management file, the clinical evaluation or the quality management system on the manufacturer's behalf; paragraphs 2, 3, 4 and 9 of Article 10 of the Medical Device Regulation give these obligations directly to the manufacturer. Nor can we widen the scope of a notified body's limited-scope certificate.
The way of working is this: the firm opens a user authorization for us from its own ÜTS firm page, and we prepare the document and device records with the "Medical Device Registration Operator" ("Tıbbi Cihaz Kayıt Elemanı") authorization. Article 40(5)(f) of the Manual leaves the "Approve" action only to users with this authorization, and point (g) of the same paragraph states that for the application to reach the Agency it must be electronically signed by the firm's signatory. The person who signs the declaration and the person who signs the application are both the firm's own authorised persons.