PRRC: The Person Responsible for Regulatory Compliance (Article 15)

Who is the PRRC (person responsible for regulatory compliance), and what are the expertise requirements and duties? The micro and small enterprise exception, whether a consultant can be the PRRC, and the link with ÜTS.

During a notified body audit or a customer enquiry you are asked "who is your PRRC?" and nobody in the company holds that title. Or you write down your quality manager's name, but there is no file showing which of the qualifications sought by the Regulation that person's diploma and experience correspond to. There is a third possibility too: you are a micro-sized manufacturer, you cannot hire anyone, and you are looking at whether it is possible to buy the service from outside.

This page reads the person responsible for regulatory compliance from a single provision: Article 15 of the Medical Device Regulation (Tıbbi Cihaz Yönetmeliği). Paragraph by paragraph it covers the routes of qualification, the concrete tasks the article places on this person, whether the person must be within the organisation, the rule that changes for micro and small enterprises, and the separate obligation for a responsible person placed on the authorised representative. It also answers the question "can a consultant be my PRRC?" on the basis of what the article actually says about permanent and continuous service and about position within the organisation.

The sources are the Medical Device Regulation published in the Official Gazette No. 31499 (bis) of 2/6/2021, the Manual on Registration Procedures of Medical Devices in the Product Tracking System (Tıbbi Cihazların Ürün Takip Sistemine Kayıt Süreçlerine İlişkin Kılavuz) and the Medical Device Sales, Advertising and Promotion Regulation (Tıbbi Cihaz Satış, Reklam ve Tanıtım Yönetmeliği) published in the Official Gazette No. 29001 of 15/5/2014.

The obligation lies with the manufacturer, not in the person's title

The design of the article is this: the manufacturer is the one under the obligation, and the person is the means by which that obligation is met. Article 15(1) of the Medical Device Regulation states that manufacturers shall have within their organisation at least one person responsible for regulatory compliance who possesses the requisite expertise in the field of medical devices.

The phrase "at least one person" sets a floor, not a ceiling. One person is enough, but there can also be more than one; in that case paragraph 4 of the article comes into play and is dealt with separately below.

The "requisite expertise" is defined within the article itself. It is not a concept left to your discretion; the same paragraph lists through which qualifications the expertise is to be proven. So what you need to produce when naming your PRRC is not a title but a documented file that corresponds to one of these qualifications.

The routes for proving expertise are alternatives, not cumulative

The points of Article 15(1) begin with the words "any of the following qualifications". So meeting one of the three routes is enough; there is no need to meet all three.

  • Diploma plus one year: Point (a) requires a diploma, certificate or other formal evidence of qualification awarded on completion of a university-level course of study, or of a course recognised as equivalent by the Council of Higher Education, in law, medicine, dentistry, pharmacy, engineering or another relevant scientific discipline, together with at least one year of professional experience in regulatory affairs or in quality management systems relating to medical devices. Four years without a diploma: Point (b) accepts four years of professional experience in regulatory affairs or in quality management systems relating to medical devices as sufficient on its own. Two years for custom-made devices: Point (c) accepts, without prejudice to national provisions on professional qualifications, at least two years of professional experience in the relevant field of manufacturing, for manufacturers of custom-made devices.

Note two details. First, the list of diplomas in point (a) is not closed; the phrase "another relevant scientific discipline" opens the list, but the burden of showing that the discipline is relevant remains with you. Second, the one year in point (a) and the four years in point (b) are not just any experience; the point requires the experience to have been gained in regulatory affairs or in quality management systems relating to medical devices. Four years spent on a production line do not correspond to this point.

The third route is open only to manufacturers of custom-made devices. Article 3(s) of the Regulation defines a custom-made device as a device specifically made in accordance with a written prescription issued by a person authorised under Law No. 1219 of 11/4/1928 by virtue of his or her professional qualifications, under that person's responsibility, intended for the sole use of a particular patient exclusively to meet their individual conditions and needs. If you manufacture in series, point (c) is not open to you.

The tasks the article places on the responsible person are listed

This person is not a general "compliance officer"; Article 15(3) lists, as a minimum, what the person is responsible for ensuring.

  • Conformity check before release: Point (a) requires ensuring that the conformity of the devices is appropriately checked, in accordance with the quality management system under which the devices are manufactured, before a device is released. This is a task tied to the quality management system that Article 10(9) orders the manufacturer to establish, document, implement and maintain. Technical documentation and EU declaration of conformity: Point (b) requires ensuring that the technical documentation and the EU declaration of conformity are drawn up and kept up to date. The content of the technical documentation is what is set out in Annexes II and III under Article 10(4); the EU declaration of conformity is regulated in Article 19. Post-market surveillance: Point (c) requires ensuring compliance with the post-market surveillance obligations under Article 10(10). That paragraph places on the manufacturer the obligation to implement and keep up to date a post-market surveillance system in accordance with Article 81.

  • Vigilance reporting: Point (ç) requires ensuring that the reporting obligations referred to in Articles 85 to 88 and in the acts adopted by the Commission are fulfilled. This is the part that extends from serious incident and field safety corrective action reporting to trend reporting and the analysis of vigilance data.

  • Investigational device statement: Point (d) requires ensuring, in the case of investigational devices, that the statement referred to in section 4.1 of Chapter II of Annex XV is issued.

None of these five headings is "signing documents"; all are built on the verb "being responsible for ensuring". So the PRRC, even if not required to do the work personally, is obliged to be sure it is done. This distinction is decisive when writing down the person's authority within the company: a person without access to information and without the ability to stop processes cannot satisfy this paragraph.

The source does not reconcile "within its organisation" with "whether or not an employee of the organisation"

Article 15(1) states that manufacturers shall have this person "within their organisation". Paragraph 5 of the same article provides that the person responsible for regulatory compliance shall not suffer any disadvantage within the manufacturer's organisation in relation to the proper fulfilment of their duties, "whether or not they are employees of the organisation".

Paragraph 1 asks for the person to be within the organisation, while paragraph 5 assumes that the person may not be an employee of the organisation. The Regulation does not expressly reconcile these two statements. The text has neither a sentence limiting the "within the organisation" requirement of paragraph 1 to an employment contract, nor a sentence expressly accepting any type of contract.

The only place where this uncertainty is explicitly resolved is paragraph 2: there, only for micro and small enterprises, it is expressly written that the person need not be kept within the organisation. That this exception was written separately indicates that paragraph 1 should be read more strictly for other enterprises. The text says this much, and we write this much.

The rule changes for micro and small enterprises

Article 15(2) states that micro and small enterprises within the scope of the Regulation on the Definition, Characteristics and Classification of Small and Medium-Sized Enterprises (Küçük ve Orta Büyüklükteki İşletmelerin Tanımı, Nitelikleri ve Sınıflandırılması Hakkında Yönetmelik), put into force by Council of Ministers Decision No. 2005/9617 of 19/10/2005, need not have the person responsible for regulatory compliance within their own organisation; however, those enterprises shall have such a person permanently and continuously at their disposal, in accordance with the procedures and principles determined by the Agency.

This paragraph has three consequences. First, the scope of the exemption depends on the definition of size, and the definition is not in the Regulation itself but in the SME Regulation to which it refers; medium-sized enterprises are not within this paragraph. Second, the exemption does not mean "you may go without a PRRC"; the obligation to obtain the service continues, and the nature of the service is described as "permanent and continuous". Support that is project-based, one-off or taken only in the audit period does not meet this definition.

Third, and most often overlooked: the paragraph requires the service to comply with the procedures and principles determined by the Agency. There is no regulation on these procedures and principles in the Manual on Registration Procedures of Medical Devices in the Product Tracking System; the Manual does not regulate the person responsible for regulatory compliance at all. So the text you will rely on when structuring your service contract is the Regulation provision dealt with on this page itself.

Can a consultant be your PRRC

The short and honest answer to this question requires placing two paragraphs of the article side by side.

For micro and small enterprises the answer is directly in the text: Article 15(2) states that these enterprises need not keep the person within their own organisation and will obtain the service from outside permanently and continuously. Here, obtaining the service from outside is not a flexibility; it is the route the paragraph envisages.

For other enterprises the text is stricter. Paragraph 1 requires the person to be within the organisation; paragraph 5 assumes that the person may not be an employee but still places the person "within the manufacturer's organisation". Because the Regulation does not reconcile these two statements, it cannot be concluded with certainty from the text whether a service obtained from outside satisfies paragraph 1. Closing this uncertainty with "a consultant can be the PRRC" or "cannot be" would be making up a provision that is not in the source.

What is certain is this: whichever route you choose, you need to set up a position from which the five tasks in paragraph 3 can actually be carried out. Paragraph 5 requires that the person not suffer any disadvantage in relation to the proper fulfilment of their duties; this includes being able to stop an unsuitable shipment and not being threatened with termination of their contract as a result. An external service provider whose contract can be terminated unilaterally at any moment and who has no access to the company's quality records meets neither the "permanent and continuous" criterion of paragraph 2 nor the protection in paragraph 5.

We do not provide this service in the capacity of PRRC under the name of medical device ÜTS registration consultancy. What we do is carry out the ÜTS-side registration and document work needed for your PRRC to be able to perform the tasks in paragraph 3, and structure the person's qualification file according to the points of paragraph 1.

If there is more than one responsible person, the scope must be in writing

In growing firms, tasks are divided: one person looks after the technical file, another after vigilance. The Regulation allows this but sets one condition. Article 15(4) states that where several persons are jointly responsible for regulatory compliance in accordance with paragraphs 1, 2 and 3, their respective areas of responsibility shall be stated in writing.

The words "in writing" are decisive here. It is not enough for both persons to have "regulatory compliance" in their job descriptions; it must be broken down which person is responsible for which point of paragraph 3. The question asked in an audit is usually not "who is responsible" but "who is responsible for this point".

The responsible person cannot be put under pressure because of the task

Article 15(5) states that the person responsible for regulatory compliance shall not suffer any disadvantage within the manufacturer's organisation in relation to the proper fulfilment of their duties, whether or not they are employees of the organisation.

This provision is not a courtesy rule; it is a safeguard placed so that the system works. Point (a) of paragraph 3 ties the check of conformity before a device is released to this person; when the person gives a negative result on that check, the consequence is that the shipment stops. The paragraph prevents that consequence from being used against the person. Its corporate counterpart is that the PRRC's reporting line is tied to senior management rather than to commercial units, and that this link is in writing.

The authorised representative has a separate responsible-person obligation

Article 15 does not regulate only the manufacturer. Paragraph 6 of the article states that authorised representatives shall have permanently and continuously at their disposal at least one person responsible for regulatory compliance who possesses the requisite expertise regarding the regulatory requirements for medical devices, and lists in two points the qualifications by which the requisite expertise is proven: in point (a), a diploma, certificate or other formal evidence of qualification awarded on completion of a university-level course of study, or of a course recognised as equivalent, in law, medicine, dentistry, pharmacy, engineering or another relevant scientific discipline, together with at least one year of professional experience in regulatory affairs or in quality management systems relating to medical devices; in point (b), four years of professional experience in regulatory affairs or in quality management systems relating to medical devices.

Two differences stand out at once. First, for the authorised representative the verb is not "has within its organisation" but "has permanently and continuously at its disposal"; so for the representative, obtaining the service from outside is expressly accepted in the paragraph itself. Second, the two-year experience route specific to custom-made devices does not appear in this paragraph; for the representative there are only two qualification routes.

This paragraph produces a concrete question to ask when buying authorised representative services: who is the representative's person responsible for regulatory compliance, and under which point is that person regarded as qualified. Since Article 11(5) of the Regulation states that where the manufacturer fails to comply with the obligations in Article 10, the authorised representative will be jointly and severally liable for defective devices together with the manufacturer, whether the representative takes this staffing seriously concerns you directly.

Do not confuse it with the responsible manager of a sales centre

In Türkiye there are two separate "responsible persons" in the same sentence, and they arise from different legislation with different obligations.

Article 10(1) of the Medical Device Sales, Advertising and Promotion Regulation states that a responsible manager meeting the specified conditions shall be present at the sales centre. Paragraph 2 of the same article requires the responsible manager to be a graduate of a higher vocational school providing professional education and training at least at associate-degree level, or a graduate of an institution abroad whose diploma equivalence has been accepted; paragraph 3 provides for the documentation of the training under Article 35; paragraph 4 for assuming responsibility for all the activities of the sales centre under that Regulation; and paragraph 5 that the responsible manager cannot work at more than one sales centre.

The content of the role is also different. Paragraph 9 of the same article states that the responsible manager will immediately notify the Agency and the manufacturer or importer of the relevant device of adverse events reported in relation to the devices sold. This is a notification duty at the point of distribution. The person responsible for regulatory compliance in Article 15 of the Medical Device Regulation, by contrast, is placed within the manufacturer's or authorised representative's organisation and has a scope extending from the technical documentation to vigilance reporting. The qualification requirements of the two roles and the legislation they rest on are also separate. If they are combined in the same person, the requirements of the two separate pieces of legislation have to be met separately.

There is no PRRC field in ÜTS, but the PRRC's work is reflected in ÜTS

The Manual on Registration Procedures of Medical Devices in the Product Tracking System does not regulate a field or notification concerning the person responsible for regulatory compliance. The "responsible manager" in Article 5(7) of the Manual is information tied to the firm registration and states that if this information is updated in ÇKYS, the change will be updated automatically in ÜTS; this is not the person in Article 15 of the Medical Device Regulation.

On the other hand, the output of the PRRC's tasks in paragraph 3 goes directly into ÜTS. The obligation in point (b) to keep the EU declaration of conformity up to date is met by keeping the ÜTS record of the "EU Declaration of Conformity (MDR/IVDR)" type document, regulated in Article 11 of the Manual, up to date. The technical documentation side of the same point shows up in the consistency of the documents linked to the device under the document business rules in Article 66 of the Manual. The post-market surveillance and reporting obligations in points (c) and (ç) require the device record to be kept up to date; Article 34(1) of the Regulation states that, without prejudice to obligations relating to EUDAMED, economic operators established in Türkiye will fulfil their obligations relating to ÜTS.

The practical result: your PRRC has no box in ÜTS, but every inconsistency in ÜTS falls within the PRRC's field of duty.

What we do on PRRC, and what we cannot do

Within medical device ÜTS registration consultancy, the outputs we produce under this heading are as follows:

  • Qualification file: We arrange the diploma, equivalence and experience records of the person you have designated so that they show which point of Article 15(1) they correspond to, and we list any missing evidence. Draft job description: We prepare a task and authority description that meets points (a) to (d) of paragraph 3 one by one and shows which decision the person takes in which process. Separation of multiple responsibilities: If more than one person is to be designated, we break down by point the written individual areas of responsibility that paragraph 4 requires. Size determination: We assess whether you can benefit from the exemption in paragraph 2 according to the definition of micro and small enterprises in the SME Regulation referred to, and, if the exemption applies, we structure the contract so that it shows "permanent and continuous" service. Authorised representative enquiry: We ask about the responsible person of the authorised representative you work with, or are considering working with, under paragraph 6, and the qualification point under which that person is regarded as qualified, and we file the answer.

  • ÜTS reflection: We check whether the document currency for which the PRRC is responsible has a counterpart in the ÜTS records, and report deviations.

What we cannot do is clear too. We do not take on the capacity of person responsible for regulatory compliance on your behalf. We cannot issue a document stating that a person is qualified; qualification is proven under the points of Article 15(1) and the evidence is held by you. We cannot produce technical documentation, a clinical evaluation or a quality management system in place of the manufacturer; if Article 11(4) states that these obligations cannot be delegated even by a mandate, they cannot be delegated at all by a consultancy contract. Nor do we give an interpretation of the procedures and principles in paragraph 2 that is not in the source.

Our way of working on the ÜTS side is this: the firm assigns us a user authorization from its own ÜTS firm page, and we create the document and device records with the "Medical Device Registration Operator" ("Tıbbi Cihaz Kayıt Elemanı") authorization. Article 39(13) of the Manual states that an application in the "Awaiting Signature" ("İmza Bekliyor") status will be electronically signed by the firm's signatory who is the applicant; the signing authority stays with the firm, and responsibility for regulatory compliance also continues to rest with the manufacturer.

For the qualifications, tasks, exception and ÜTS registration of the person responsible for regulatory compliance, you can look at our guide page on the person responsible for regulatory compliance (PRRC).

Tıbbi Cihaz ÜTS Kayıt Danışmanlığı

Request an ÜTS registration consultancy quote

Ehlibeyt Mahallesi, Tekstilciler Caddesi No: 35/7 06520 Çankaya / Ankara - TÜRKİYE

Cevizli Mahallesi, Zuhal Caddesi Ritim İstanbul B Blok, No:44 Maltepe / İstanbul - TÜRKİYE

info@medexmedikal.com
+90 505 580 4622
+90 312 389 1354

ÜTS Guides · Legislation Calendar · Glossary · ÜTS Guide Index · Frequently Asked Questions

Services: ÜTS Consultancy · Who Carries Out the ÜTS Registration? · ÜTS Document Registration Consultancy · ÜTS Product Registration Consultancy · Importer ÜTS Registration Consultancy · Sales Centre Authorisation Certificate · EUDAMED Registration Consultancy · MDR/IVDR Declaration of Conformity · Technical File Preliminary Review · Label and Instructions for Use Check · PRRC Service · PMS and PSUR · Clinical Evaluation · EMDN and GMDN Code Determination · Certificate of Free Sale · EUDAMED XML Generator · ÜTS Registration Training · Correction of a Rejected ÜTS Application

KVKK and Privacy Policy | Cookie Policy | Accessibility Statement

© 2026 Medex Kurumsal Danışmanlık Ltd. Şti.

bottom of page
WhatsApp Hemen Ara