Notified Body and ÜTS Document Registration: Certificate Rules
How do the designation scope and identification number of a notified body affect ÜTS document registration? The language, content and validity of the certificate, suspension and restriction.
There is a four-digit number on your certificate, and when you start the document record in ÜTS you cannot find that body in the list. Or the body is in the list and the certificate looks valid, but after the record you receive a 45-day revision decision. Both situations have the same source: the notified body's own status is a status that can change independently of your certificate and is monitored continuously; ÜTS also places the certificate on top of this status.
This page explains the notified body from the point of view of the firm making the ÜTS record: by whom and how the body is designated, what the designation scope means, where the identification number comes from and where it must appear, what must be inside the certificate and what is not looked for, how long a certificate is valid, and what happens to your certificate if the body's designation is suspended or withdrawn. We do not go into the relationship of certificate types to conformity assessment routes here.
The sources are the Medical Device Regulation (Tıbbi Cihaz Yönetmeliği) published in the Official Gazette No. 31499 (bis) of 2/6/2021 and the Manual on Registration Procedures of Medical Devices in the Product Tracking System (Tıbbi Cihazların Ürün Takip Sistemine Kayıt Süreçlerine İlişkin Kılavuz).
A notified body is not an inspection firm but a designated body
The authority of a notified body comes not from its own commercial decision but from an act of designation. Article 35(1) of the Medical Device Regulation says that the Agency is the authority responsible for planning the process for designating a conformity assessment body as a notified body to carry out conformity assessment activities under the Regulation, or for designating a notified body, and that the Agency is also responsible for monitoring notified bodies, including their subcontractors and subsidiaries and representative offices.
Article 38(1) of the same Regulation says that conformity assessment bodies established in Turkey apply to the Agency for designation. Point (a) of paragraph 2 of the same article requires the application to specify the conformity assessment activities and the device types for which the body applies to be designated, and to be supported by documentation demonstrating compliance with Annex VII. Point (b) of the same paragraph requires the submission of a valid accreditation certificate issued by the Turkish Accreditation Agency (Türk Akreditasyon Kurumu), and the relevant assessment report, concerning the organisational and general requirements and the quality management requirements set out in points 1 and 2 of Annex VII.
The qualities the body must have are also listed in the Regulation. Article 36(1)(a) of the Medical Device Regulation says that notified bodies fulfil the tasks for which they are designated and, meeting the organisational and general requirements and the quality management, resource and process requirements needed for these tasks, comply in particular with Annex VII. Point (b) of the same paragraph says that the personnel referred to in points 3.2.3 and 3.2.7 of Annex VII are employed by the notified body itself and cannot be external experts or subcontractors.
The point most often overlooked is when a designation takes effect. Article 42(10) of the Medical Device Regulation says that the designation becomes valid on the day after the notification is published in NANDO, and that the published notification states the scope of the lawful conformity assessment activity of the notified body. Paragraph 11 of the same article completes this: the conformity assessment body concerned may carry out notified body activities only after the designation has become valid.
The designation scope draws the limit of the certificate
A notified body cannot issue a certificate for every device. Article 42(3) of the Medical Device Regulation says that the notification clearly states the designation scope, showing the conformity assessment activities and the device types the notified body is authorised to assess, and the conditions associated with the designation. Paragraph 12 of the same article says that the codes and list of device types prepared by the Commission are used as the basis in determining the designation scope.
This scope also limits the manufacturer's freedom of choice. Article 53(1) of the Medical Device Regulation says that, where a notified body must be involved in the conformity assessment procedure, the manufacturer may apply to a notified body of its choice, provided that the body is designated for the conformity assessment activities relating to the device types concerned. The same paragraph also contains a prohibition: the manufacturer cannot lodge an application in parallel with another notified body for the same conformity assessment procedure. Paragraph 3 of the same article says that, when applying, the manufacturer declares whether it has withdrawn an earlier application to another notified body before that body's decision, and submits information about any previous application for the same conformity assessment that was refused by another body.
A body is not free to hand its tasks to others either. Article 37(1) of the Medical Device Regulation says that, where a notified body subcontracts specific tasks or uses a subsidiary or representative office, it verifies that the subcontractor or the subsidiary or representative office meets the applicable requirements set out in Annex VII and informs the Agency. Paragraph 2 of the same article says that notified bodies take full responsibility for the tasks performed on their behalf by these units; paragraph 4 says that conformity assessment activities may be subcontracted, or carried out by a subsidiary or representative office, provided that the person who made the application is informed accordingly.
The practical result is this: which device types your certificate covers depends not on the body's own statement but on the designation scope. A certificate issued for a device outside the scope is not a valid certificate, even if it looks formally flawless.
The identification number is the body's identity in ÜTS
What lets you find the notified body in ÜTS is not its name but its number. Article 43(1) of the Medical Device Regulation says that the Commission assigns an identification number to each notified body whose notification has become valid under Article 42(10), and that a single identification number is assigned even if the body is notified under more than one piece of legislation. The same paragraph also contains a continuity rule: bodies approved under the repealed Medical Device Regulation and the repealed Active Implantable Medical Devices Regulation retain the identification numbers assigned to them under those Regulations when they are designated under this Regulation. This is why the number you know from the MDD period appears the same on the MDR certificate.
The number is visible on the device as well. Article 20(5) of the Medical Device Regulation says that, where a notified body is responsible for the conformity assessment procedures referred to in Article 52, the CE marking is followed by the identification number of the notified body, and that this number is also shown in any promotional material which mentions that the device fulfils the requirements for CE marking.
In the document fields of ÜTS the number appears in three separate documents. Article 7(2)(d) of the Manual looks for the name and identification number of the notified body in the EU certificate; point (e) of the same paragraph looks for the address of the body. Article 6(2)(d) and (e) of the Manual repeat the same two fields for the EC certificate. Article 11(2)(ı) of the Manual carries the number to the declaration: EU declarations of conformity, other than those whose scope consists only of Class-I Other or Class A devices, must contain the name and identification number of the notified body, a description of the conformity assessment procedure carried out, and the identification of the EU certificate or certificates issued.
The same number must appear in all three documents. The last sentence of Article 7(2) of the Manual makes it a condition that the information entered into ÜTS is the same as the information in the certificate; the last sentence of Article 6(2) of the Manual sets the same rule for the EC certificate.
If the body is not defined in ÜTS, a request is opened before the document record
Your certificate may be valid while the body is not registered in ÜTS. Article 7(7) of the Manual regulates this case separately: if the notified body that issued the EU certificate is not defined in ÜTS as a notified body, then before the ÜTS document registration of the EU certificate concerned, a request is opened, using the "Report Request" ("Talep Bildir") screen in ÜTS, through the "Notified/Conformity Assessment Body Procedures" ("Onaylanmış/Uygunluk Değerlendirme Kuruluş İşlemleri") field, for the body to be added to ÜTS or for an authorisation to be defined under the relevant legislation.
The same paragraph also writes how the request is concluded: if, as a result of queries made on the NANDO Notified Bodies search site upon the request, it is established that the conformity assessment body is authorised as a notified body under the relevant legislation, the body is added to ÜTS and defined in ÜTS according to the New Medical Device Regulations under which it is authorised, that is, the MDR and/or the IVDR. Article 6(7) of the Manual defines the same route for EC certificates and states there that the definition will be made according to the MDD, the IVDD and/or the AIMDD.
The detail here is important: the body is added to ÜTS not "in general" but on the basis of the regulation under which it is authorised. A body may be defined in ÜTS under the MDD but not under the MDR. If your certificate is an MDR certificate and the body is not defined under the MDR, you still need to open this request before starting the document record.
On the Regulation side, where the notification is published points to the same place. Article 42(2) of the Medical Device Regulation says that the conformity assessment bodies designated by the Agency are notified by the Ministry of Trade to the Commission and the EU member states through NANDO. Paragraph 6 of the same article says that, if no objection arises, the notification will be published in NANDO within 42 days of the notification being made.
What the certificate names, and what it does not have to name
The information that the certificate you register in ÜTS must carry varies by document type. Article 7(2)(h) of the Manual splits the product information and scope field for EU certificates into two. Sub-point (1) of the same point looks, for EU technical documentation assessment certificates, EU type-examination certificates and EU product verification certificates, for a clear identification of the device, including name, model and type, its intended purpose, its risk class and the Basic UDI-DI. Sub-point (2) of the same point looks, for EU quality management system certificates and EU production quality assurance certificates, for the identification of the devices or device groups, their risk class and, for Class-IIb devices, the intended purpose information.
The wording of this second sub-point is clear: in certificates that look at the quality system, the "identification of the device group" alone is sufficient; the device name and the Basic UDI-DI are not looked for. On the old regulation side no such distinction is made. Article 6(2)(i) of the Manual looks only for "product information / scope" in the EC certificate; a separate rule on whether this field will contain the product name or the product group, according to the annex under which the certificate was issued, is not regulated in this Manual. Therefore an EC certificate issued under the MDD Annex II full quality assurance system, which carries a product group and not a device name, is not considered incomplete in terms of the fields the Manual looks for; what is decisive is whether the scope covers your device.
The non-identifying fields of the certificate are listed in the same paragraphs. Article 7(2)(c) of the Manual looks for the certificate number, point (ç) the manufacturer's Single Registration Number if already issued, point (f) the regulation it belongs to, point (g) the annex under which it was issued, that is, reference to the annex under which the conformity assessment was carried out, and point (ğ) the issue and validity dates of the document. Point (i) of the same paragraph asks for reference to the examinations and tests carried out, for example the relevant common specifications, harmonised standards, test reports and audit reports; point (k) for information on surveillance carried out by the notified body, where applicable; and point (l) for the results of the notified body's conformity assessment relating to the relevant annex. Article 6(2)(j) and Article 7(2)(n) of the Manual look, in both certificate types, for the legally binding signature of the notified body in accordance with the applicable national legislation.
How the scope of the certificate is uploaded is also bound by a rule. Article 7(6) of the Manual requires the EU certificate to be uploaded and submitted to the Agency in full, with all its annexes and additional pages; Article 6(6) of the Manual requires the same for the EC certificate. Since the scope list sits in the annex in most certificates, these two paragraphs relate directly to proof of scope.
The language, content and duration of the certificate are bound by the Regulation
The formal framework of the certificate comes from the Regulation. Article 56(1) of the Medical Device Regulation says that certificates issued by notified bodies under Annex IX, Annex X and Annex XI will be in Turkish and/or in an official EU language determined by the Agency or, failing such a determination, in an official EU language acceptable to the notified body and/or in Turkish, and that the minimum content of the certificates will be as set out in Annex XII.
On the duration side, paragraph 2 of the same article is clear: certificates remain valid for the period they indicate, which does not exceed 5 years. The same paragraph says that, on application by the manufacturer, the validity may be extended, on the basis of a re-assessment under the applicable conformity assessment procedures, for further periods, each not exceeding 5 years; and that any supplement to a certificate remains valid as long as the certificate it supplements is valid.
The Manual repeats the same period on its own side. Article 7(3) of the Manual says that EU certificates issued under the New Medical Device Regulations remain valid for the period indicated in the certificate, not exceeding 5 years, as long as no change or update is made to the certificate within this period. Article 6(3) of the Manual says that EC certificates, other than the EC Verification Certificate and the EC Type Verification Certificate, have a validity period, which cannot be more than 5 years. For the quality management document the limit is shorter: Article 20(5) of the Manual says that for documents of the ISO 13485 type the validity period cannot be more than 3 years.
The date field of the certificate is filled in ÜTS by a mechanical rule. Article 7(4) of the Manual says that, if the document has been revised, the revision date is entered as the document start date; if it has not been revised and the start date is not clearly stated in the certificate, the most current issue date in the certificate is entered. Paragraph 5 of the same article says that, if the revision number is shown together with the certificate number, the certificate is recorded together with its revision number.
Restricting and suspending a certificate is within the body's authority
A certificate is not an untouchable document once it has been issued. Article 56(3) of the Medical Device Regulation says that notified bodies may impose restrictions on the intended purpose of a device for certain groups of patients or require manufacturers to undertake specific post-market clinical follow-up studies under Annex XIV Part B. These restrictions are written into the text of the certificate itself and, under Article 7(2)(m) of the Manual, are among the fields checked in the ÜTS review under the heading of conditions or limitations on the validity of the certificate.
Paragraph 4 of Article 56 of the same Regulation defines a heavier authority: where the notified body finds that the requirements of the Regulation are no longer met by the manufacturer, and compliance is not ensured through corrective action by the manufacturer within an appropriate deadline set by the body, it, taking account of the principle of proportionality, suspends or withdraws the certificate issued or imposes restrictions on it; the body gives the reasons for its decision.
These decisions do not remain secret. Paragraph 5 of the same article says that the notified body enters into the electronic system referred to in Article 57 all information concerning certificates issued, including amendments and supplements, and concerning certificates suspended, reinstated, withdrawn or refused and restrictions imposed on certificates, and that this information is accessible to the public. Article 57(1)(g) of the same Regulation lists this information among the contents of the electronic system.
The body itself is also monitored continuously
The status of a notified body does not freeze on the day of its designation. Article 44(1) of the Medical Device Regulation says that notified bodies notify the Agency without delay, and no later than within 15 days, of all changes that may affect their compliance with the requirements in Annex VII or their ability to carry out the conformity assessment activities relating to the devices for which they have been designated.
Article 44(4)(a) of the same Regulation says that the Agency re-assesses, at least once a year, the notified bodies established in the country and, where necessary, their subsidiaries, representative offices and subcontractors, as to whether they continue to meet the requirements of Annex VII, and that this assessment includes an on-site inspection. Paragraph 5 of the same article says that the monitoring includes witnessed audits carried out while the body's personnel perform quality management system assessments at a manufacturer's premises. Paragraph 7 states that, in addition to regular monitoring or on-site assessments, the Agency may conduct short-notice, unannounced or "for-cause" reviews.
There is also a long-interval check. Article 44(10) says that, 5 years after the notification of a notified body and every 5 years thereafter, a full re-assessment is carried out to determine whether the body continues to meet the requirements set out in Annex VII. The same paragraph states that the Agency may also carry out a full re-assessment before these dates, at the body's request or where it has concerns based on the results of the annual assessments.
The monitoring goes down to the files on which the certificates rest. Article 45(1) of the Medical Device Regulation says that, as part of its continuous monitoring, the Agency reviews, on an appropriate sample size, the notified bodies' assessments of manufacturers' technical documentation and in particular of the clinical evaluation documentation; these reviews are carried out both off-site and on-site. Paragraph 2 of the same article requires the sampling of the files to be reviewed to be representative of the types and risk of devices certified by the body, in particular high-risk devices.
Suspension of the designation directly affects your certificate
The most critical scenario is the restriction, suspension or withdrawal of the body's own designation. Article 46(4) of the Medical Device Regulation says that, where the Agency determines that a notified body no longer meets the requirements of Annex VII or has not fulfilled its obligations or has not implemented the necessary corrective measures, it suspends, restricts or fully or partially withdraws the designation, depending on the seriousness. The same paragraph states that the suspension does not exceed 1 year and may be renewed once for the same period.
It is an obligation that the manufacturer becomes aware of this. Paragraph 5 of the same article says that, in the event of suspension, restriction or full or partial withdrawal of its designation, the notified body informs the manufacturers concerned within 10 days at the latest. Paragraph 7(a) of the same article says that the Agency assesses the impact on the certificates issued by the body, and point (c) says that the Agency asks the body to suspend or withdraw, within a reasonable period determined by the Agency, any certificates unduly issued, in order to ensure the safety of devices on the market.
For duly issued certificates a conditional continuity is provided. Paragraph 8 of the same article says that, where a designation has been suspended or restricted, certificates remain valid if one of two conditions is met. Point (a) of the same paragraph requires the Agency to confirm, within 1 month of the suspension or restriction, that there is no safety issue in relation to the certificates affected, and to set out the actions foreseen for the remedy with a timeline. Point (b) of the same paragraph requires the Agency to confirm that no certificates will be issued, amended or re-issued during the suspension, and to determine whether the body has the capability to continue monitoring existing certificates; if it is determined that it does not, manufacturers submit to the competent authority, within 3 months of the suspension or restriction, a written confirmation that another competent notified body is temporarily assuming the functions.
In the case of withdrawal of the designation the period is more definite. Paragraph 9 of the same article says that, except for certificates unduly issued, certificates remain valid for a period of 9 months under two conditions: that the competent authority confirms that there is no safety issue associated with the devices, and that another notified body confirms in writing that it will assume immediate responsibilities and that its assessment will be completed within 12 months of the withdrawal of the designation. The same paragraph states that, under these conditions, the certificates may be provisionally extended by further periods of 3 months, the total not exceeding 12 months.
A body's withdrawal on its own decision is regulated separately. Paragraph 3 of the same article says that, where a notified body decides to cease its conformity assessment activities, in the case of a planned cessation it informs the Agency and the manufacturers concerned 1 year before ceasing its activities. The same paragraph states that certificates may remain valid for a temporary period of 9 months after the cessation of activities, provided that another notified body has confirmed in writing that it will assume responsibilities, and that the new body completes a full assessment of the devices by the end of this period before issuing new certificates.
A change of body ties the expiry date of the certificate to a contract
A manufacturer may also change body of its own accord. Article 58(1) of the Medical Device Regulation says that, where a manufacturer terminates its contract with a notified body and enters into a contract with another body for the conformity assessment of the same device, the detailed arrangements for the change are clearly defined in an agreement between the manufacturer, the incoming body and, where practicable, the outgoing body.
The same paragraph lists the minimum content of this agreement. Point (a) covers the date on which the certificates issued by the outgoing body become invalid; point (b) the date until which the identification number of the outgoing body may be indicated in the information supplied by the manufacturer, including promotional material; point (c) the transfer of documents, including confidentiality aspects and proprietary rights; point (ç) the date after which the tasks of the outgoing body are assigned to the incoming body; and point (d) the last serial or lot number for which the outgoing body is responsible. Paragraph 2 of the same article says that the outgoing notified body withdraws the certificates it issued for the device concerned on the date on which they become invalid.
On the ÜTS side this means not a new document record after a change of body, but setting up correctly the date line between updating the existing record and the closing of the old certificate. The date until which the identification number may be shown in promotional material is also written in the same agreement, and this directly affects label and instructions for use updates.
ÜTS queries the validity of the certificate after the record as well
Registering the certificate in ÜTS does not mean the process is over. Article 39(18) of the Manual says that, where a document application contains a notification of the document types mentioned in Articles 6 and 7 of the Manual and the assessment has a positive outcome, the validity and currency of the certificates that become registered are queried by the Agency through the notified body that issued the certificate or the competent authority that authorised that body.
Point (a) of the same paragraph says that a 45-day revision action is applied to certificates determined not to be current; as an exception, only for EU certificates under the MDR and IVDR, and only for certificates revised because of an expansion of product scope, a 120-day extension action is applied instead of the 45-day revision action. Point (b) says that certificates determined to be invalid for reasons other than having lost currency are directly rejected, and that for documents determined to be forged or tampered with, a market surveillance and inspection process is also started after the rejection. Both points state that these actions are carried out for all certificates with the same document number that are registered in ÜTS and not current.
That is, a status change on the notified body side can have consequences in ÜTS even without touching your application. Monitoring the status of a registered certificate is a continuation of the care you showed at the time of registration.
What we do on the certificate side, and what we cannot do
Within medical device ÜTS registration consultancy, our concrete outputs on the notified body side are these:
Body and number verification: We compare the body name, identification number and address in the certificate with Article 7(2)(d) and (e) of the Manual, and check, before the document record, whether the body is defined in ÜTS under the relevant regulation.
Opening a request: If the body is not defined in ÜTS, we prepare the "Report Request" ("Talep Bildir") route in Article 7(7) or Article 6(7) of the Manual, selecting the correct request field.
Scope reading: Based on Article 7(2)(h) of the Manual, we distinguish whether the device name is looked for according to the type of certificate, and assess together with the annexes whether the scope includes your device.
Duration and revision calendar: Taking as the basis the 5-year upper limit in Article 56(2) of the Medical Device Regulation and the 3-year limit in Article 20(5) of the Manual, we track document expiry dates and revision numbers.
Status change scenario: If the body's designation is suspended or withdrawn, we determine which of the validity conditions in paragraphs 8 and 9 of Article 46 of the Medical Device Regulation operates, and present in writing what needs to be done for the record in ÜTS.
What we cannot do is clear as well. We cannot intervene in the decision of a notified body, expand the scope of a certificate, or provide a certificate for a device outside the body's designation scope; Article 42(3) of the Medical Device Regulation ties the designation scope to the notification, and Article 53(1) ties the choice of body to the manufacturer. Nor can we change a decision to suspend or withdraw a certificate; Article 56(4) of the same Regulation gives this authority to the notified body. The decision to add a notified body to ÜTS also belongs to the Agency; Article 7(7) of the Manual ties the addition to the result of the NANDO query.
Our way of working is this: the firm defines a user authorisation for us from its own ÜTS firm page, we create the certificate records and the document application with the "Medical Device Registration Operator" ("Tıbbi Cihaz Kayıt Elemanı") authorisation; the person who electronically signs the application remains the firm's signatory.