Who Has a Foreign Manufacturer Registered in ÜTS?
A manufacturer established abroad is not the party that opens a company registration in ÜTS in its own name. The party that registers in ÜTS as a company and places the device on the market in Türkiye is the importer established in Türkiye…

A manufacturer established abroad is not the party that opens a company registration in ÜTS in its own name. The party that registers in ÜTS as a company and places the device on the market in Türkiye is the importer established in Türkiye. The foreign manufacturer is defined in the “Manufacturer Institution/Firm” field during device and document registration, and the importer firm making the registration enters this definition as well.
How a foreign manufacturer is defined in ÜTS
In device registration, the “Imported/Manufactured Information” field is decisive. If the manufacturing firm (legal manufacturer) that issued the declaration of conformity to which the device is linked is established in Türkiye, this field is selected as “Manufactured”; if it is not located in Türkiye, it is selected as “Imported”. When “Imported” is selected, filling in the country imported from and the GTIP number becomes mandatory and the manufacturer firm must be selected.
- The relevant foreign manufacturer firm is selected from the “Select Manufacturer Institution” screen that appears under the “Manufacturer Institution/Firm” field.
- If there is no record for the firm on this screen, the firm is saved by entering its details in the “Add Manufacturer Institution/EU Authorized Representative” section on the same screen, and is then selected from the “Select Manufacturer Institution” screen.
- If the manufacturer is established outside Türkiye or the EU member states, the EU Authorized Representative appointed by the manufacturer must also be selected. If there is no record for this firm either, it is saved by entering its details in the same add section.
Authorized representative requirement
According to the Medical Device Regulation, if the manufacturer is not established in Türkiye or an EU member state, the device may be placed on the market only on condition that the manufacturer appoints an authorized representative. The appointment of the authorized representative becomes valid when a mandate is issued and accepted in writing by the authorized representative, and covers as a minimum all devices of the same generic device group. Among the tasks under the mandate are complying with registration obligations and verifying that the manufacturer has complied with its registration obligations.
The importer's verification obligation
The Regulation also ties the accuracy of the registration to the importer. Within 2 weeks of the placing on the market of a device other than a custom-made device, importers verify that the manufacturer or authorized representative has entered the relevant information in the electronic system. If the information is missing or incorrect, they inform the relevant authorized representative or manufacturer and add their own information to the relevant entry. Economic operators update the data within 1 week of any change to this information.
A similar structure exists for custom-made medical devices: the ÜTS application for custom-made medical devices other than custom-made orthopedic orthoses and prostheses is made by the firms that manufacture or import the device concerned and whose activity area registered in ÜTS is “Manufacturer / Importer / Distributor / Exporter”.
Common mistakes
- Assuming that a foreign manufacturer can open a company registration in ÜTS in its own name in Türkiye.
- Entering the manufacturer's name as free text; the manufacturer is selected from the “Select Manufacturer Institution” screen and, if there is no record, added first.
- Leaving the EU authorized representative information blank; this selection is mandatory for manufacturers established outside Türkiye or the EU.
- Marking the registration as “Manufactured” although the legal manufacturer is abroad and skipping the country imported from and GTIP fields.
Who it concerns: importer firms, EU authorized representatives established in Türkiye, and distributors working with manufacturers abroad. İthalatçı firmalar, Türkiye'de yerleşik AB yetkili temsilcileri ve yurt dışı imalatçıyla çalışan distribütörler.
Common mistake: instead of trying to open a separate company registration in ÜTS for the foreign manufacturer, failing to make the required definition within the device registration. Yabancı imalatçı için ÜTS'de ayrı bir firma kaydı açmaya çalışmak yerine gereken tanımı cihaz kaydı içinde yapmamak.
Basis: Guide on Registration Processes of Medical Devices in the Product Tracking System (TCKKD-KLVZ-03, Rev.00, effective 13.05.2026) ARTICLE 27 and ARTICLE 29; Medical Device Regulation (RG 02.06.2021-31499 bis) ARTICLE 3, ARTICLE 11, ARTICLE 30 and ARTICLE 31. Tıbbi Cihazların Ürün Takip Sistemine Kayıt Süreçlerine İlişkin Kılavuz (TCKKD-KLVZ-03, Rev.00, yürürlük 13.05.2026) MADDE 27 ve MADDE 29; Tıbbi Cihaz Yönetmeliği (RG 02.06.2021-31499 mükerrer) MADDE 3, MADDE 11, MADDE 30 ve MADDE 31.
Before Registration