When Is a Notified Body Agreement Required?
A written agreement with a notified body is one of the conditions for benefiting from the transitional provisions, and its dates vary by legislation.

A written agreement with a notified body is one of the conditions for benefiting from the transitional provisions, and its dates vary by legislation. On the MDR side, the manufacturer must have made a formal application to a notified body by 26.05.2024 and signed the written agreement by 26.09.2024. On the IVDR side, the agreement date is 26.09.2025, 26.09.2026 or 26.09.2027, depending on the device's risk class.
The two gates on the MDR side
Under sub-paragraph (5) of paragraph (c) of the fifth paragraph of PROVISIONAL ARTICLE 1 of the Medical Device Regulation:
- By 26.05.2024, the manufacturer or authorised representative must have made a formal application to a notified body for conformity assessment of the device concerned or of the device intended to replace it.
- By 26.09.2024, the notified body and the manufacturer must have signed a written agreement.
If either of these two conditions is not met, the device cannot benefit from the right to be placed on the market until 31.12.2027 or 31.12.2028. The same dates are also required for Class III custom-made implantable devices.
Agreement for devices whose certificate expired early
For certificates that expired before 20.03.2023 but were valid on 26.05.2021, there is an additional condition: before the certificate expired, a written agreement must have been signed between the manufacturer and a notified body for conformity assessment of the device covered by the expired certificate or the device intended to replace it. Alternatively, it is required that the Agency has accepted an exemption under Article 59, or has requested the conformity assessment procedure to be carried out under Article 94.
Dates on the IVDR side
For IVD devices that did not require a notified body under the old legislation but do under the new legislation, the application and agreement dates are as follows:
- Class D group: application before 26.05.2025, agreement before 26.09.2025.
- Class C group: application before 26.05.2026, agreement before 26.09.2026.
- Class B and Class A Sterile group: application before 26.05.2027, agreement before 26.09.2027.
Surveillance responsibility and certification
- The notified body that signs the written agreement becomes responsible for surveillance of the devices covered by the agreement on the MDR side by 26.09.2024 at the latest. If the agreement covers a device intended to replace an existing device, surveillance is carried out according to the replacing device.
- In an extension application made in Türkiye, the confirmation letter issued by the notified body must include a statement that the application was received, a statement that the written agreement was signed, and, where applicable, the date of the agreement and a clear and descriptive list of the devices covered by the agreement.
- Apostille or notarisation is not required for these documents; for documents in a foreign language, the procedure set out in the Guide is followed.
Common mistakes
- Meeting the application date but missing the agreement date; the two conditions are required separately.
- Assuming the list of devices in the agreement covers the entire product portfolio.
- Confusing the MDR and IVDR agreement dates.
- Submitting the notified body confirmation letter without the minimum content requirements.
Who it concerns: Manufacturers benefiting from the transitional provisions, authorised representatives, importers and companies carrying out the extension application. Geçiş hükümlerinden yararlanan imalatçılar, yetkili temsilciler, ithalatçılar ve süre uzatım başvurusunu yürüten firmalar.
Common mistake: Treating the application to the notified body as sufficient and not tracking the written agreement date. Onaylanmış kuruluşa başvuruyu yeterli sayıp yazılı anlaşma tarihini takip etmemek.
Basis: Medical Device Regulation (OG 02.06.2021-31499 bis) PROVISIONAL ARTICLE 1, ARTICLE 59 and ARTICLE 94; TİTCK ÜTS Registration Guide (TCKKD-KLVZ-03, Rev.00) ARTICLE 46, ARTICLE 47, ARTICLE 49 and ARTICLE 50. Tıbbi Cihaz Yönetmeliği (RG 02.06.2021-31499 mükerrer) GEÇİCİ MADDE 1, MADDE 59 ve MADDE 94; TİTCK ÜTS Kayıt Kılavuzu (TCKKD-KLVZ-03, Rev.00) MADDE 46, MADDE 47, MADDE 49 ve MADDE 50.
MDR and IVDR Transition