How Is an IVD Device Classified?

In vitro diagnostic medical devices are divided into four classes, Class A, Class B, Class C and Class D, taking into account the risks arising from their intended purpose and design…

ÜTS frequently asked questions: How Is an IVD Device Classified?

In vitro diagnostic medical devices are divided into four classes, Class A, Class B, Class C and Class D, taking into account the risks arising from their intended purpose and design. Classification under the IVDR is made according to a total of 7 rules, and the highest-risk group is Class D. Classification is carried out under the provisions of the In Vitro Diagnostic Medical Device Regulation (OG 02.06.2021-31499 bis).

Summary of the seven rules

  • Rule 1: Devices intended to detect the presence of, or exposure to, a transmissible agent in order to assess suitability for transfusion, transplantation or cell administration; to detect a transmissible agent that causes a life-threatening disease with a high or suspected high risk of propagation; or to determine the infective load of a life-threatening disease where monitoring is critical in patient management are Class D.
  • Rule 2: Devices intended for blood grouping, determining foetomaternal blood incompatibility or tissue typing are Class C. However, if they are intended to determine markers of the ABO, Rhesus, Kell, Kidd or Duffy systems, they are Class D.
  • Rule 3: Devices used to detect sexually transmitted agents, as companion diagnostics, for cancer screening, diagnosis or staging, for human genetic testing, for prenatal screening, for screening congenital disorders in newborns and for the other purposes listed in the rule are Class C.
  • Rule 4: Devices for self-testing are Class C; however, devices intended for detecting pregnancy, fertility testing, determining cholesterol level and detecting glucose, erythrocytes, leucocytes and bacteria in urine are Class B. Near-patient testing devices are classified in their own right.
  • Rule 5: General culture media and histological stains, products for general laboratory use, accessories without critical characteristics, buffer solutions, washing solutions, instruments intended for use in in vitro diagnostic procedures and specimen receptacles are Class A.
  • Rule 6: All other in vitro diagnostic devices are Class B.
  • Rule 7: Controls without a quantitative or qualitative assigned value are Class B.

General principles to observe in practice

  • Classification is based on the intended purpose of the device.
  • Accessories of an IVD device are classified in their own right, separately from the device with which they are used.
  • Calibrators intended to be used with a device are in the same class as the device.
  • Control materials with quantitative or qualitative assigned values for a specific analyte or multiple analytes are in the same class as the device.
  • Each of the classification rules is applied to primary-level analyses, confirmatory analyses and supplemental analyses.
  • Where more than one rule can be applied, the strictest rule that gives the highest classification applies.

Sub-distinctions in ÜTS

In ÜTS, additional distinctions are made according to features such as whether the device is packed sterile or is a companion diagnostic. For example, a device registered as "Near-Patient Test Device (Class C)" is classified as Class C under the IVDR. The options "Class A Sterile", "Self-Test Device" and "Companion Diagnostic Device" are also among these distinctions.

Common mistakes

  • Treating Rule 6 as the default and not assessing Rule 1 to Rule 5 first.
  • Treating a self-test device and a near-patient test device as the same; near-patient test devices are classified in their own right.
  • Determining the class of calibrators and control materials independently of the main device.
  • Taking the sub-distinction in ÜTS to be a separate risk class.

Who it concerns: IVD device manufacturers, importers, suppliers of laboratory products and ÜTS registration officers. IVD cihaz imalatçıları, ithalatçılar, laboratuvar ürünleri tedarikçileri ve ÜTS kayıt sorumluları.

Common mistake: Classifying by the technology of the test kit, rather than on the target analyte and intended purpose. Sınıflandırmayı test kitinin teknolojisine göre yapmak, hedef analit ve kullanım amacını esas almamak.

Basis: In Vitro Diagnostic Medical Device Regulation (OG 02.06.2021-31499 bis); TİTCK ÜTS Registration Guide (TCKKD-KLVZ-03, Rev.00) ARTICLE 52 and ARTICLE 57. İn Vitro Tanı Amaçlı Tıbbi Cihaz Yönetmeliği (RG 02.06.2021-31499 mükerrer); TİTCK ÜTS Kayıt Kılavuzu (TCKKD-KLVZ-03, Rev.00) MADDE 52 ve MADDE 57.

Classification and CE

Medical Device ÜTS Registration Consultancy

Request an ÜTS registration consultancy quote

Ehlibeyt Mahallesi, Tekstilciler Caddesi No: 35/7 06520 Çankaya / Ankara - TÜRKİYE

Cevizli Mahallesi, Zuhal Caddesi Ritim İstanbul B Blok, No:44 Maltepe / İstanbul - TÜRKİYE

info@medexmedikal.com
+90 505 580 4622
+90 312 389 1354

ÜTS Guides · Legislation Calendar · Glossary · ÜTS Guide Index · Frequently Asked Questions

Services: ÜTS Consultancy · Who Carries Out the ÜTS Registration? · ÜTS Document Registration Consultancy · ÜTS Product Registration Consultancy · Importer ÜTS Registration Consultancy · Sales Centre Authorisation Certificate · EUDAMED Registration Consultancy · MDR/IVDR Declaration of Conformity · Technical File Preliminary Review · Label and Instructions for Use Check · PRRC Service · PMS and PSUR · Clinical Evaluation · EMDN and GMDN Code Determination · Certificate of Free Sale · EUDAMED XML Generator · ÜTS Registration Training · Correction of a Rejected ÜTS Application

KVKK and Privacy Policy | Cookie Policy | Accessibility Statement

© 2026 Medex Kurumsal Danışmanlık Ltd. Şti.

bottom of page
WhatsApp Hemen Ara