Is CE Required for a Custom-Made Device?
No. Article 20 of the Medical Device Regulation exempts custom-made devices from the obligation to bear the CE conformity marking.

No. Article 20 of the Medical Device Regulation exempts custom-made devices from the obligation to bear the CE conformity marking. Instead, manufacturers of custom-made devices follow the procedure set out in Annex XIII and, before placing the device on the market, draw up the statement referred to in point 1 of Annex XIII. For Class III custom-made implantable devices, an additional conformity assessment is required.
What is a custom-made device
Article 3 of the Regulation defines it as a device specifically made in accordance with a written prescription issued by a person authorized under Law No. 1219, who bears responsibility for it, stating specific design characteristics, and intended for the sole use of a particular patient exclusively to meet their individual conditions and needs; mass-produced devices that need to be adapted to meet the specific requirements of any professional user, and devices that are mass-produced by industrial manufacturing processes in accordance with the written prescription of an authorized person, are excluded.
Content of the Annex XIII statement
For custom-made devices, the manufacturer or authorized representative draws up a statement containing, as a minimum, the following information:
- The name and address of the manufacturer and of all manufacturing sites.
- Where applicable, the name and address of the authorized representative.
- Data allowing identification of the device.
- A statement that the device is intended for exclusive use by a particular patient or user, identified by name, an acronym or a numerical code.
- The name of the authorized person who made out the prescription and, where applicable, the name of the healthcare institution concerned.
- The specific characteristics of the product as indicated in the prescription.
- A statement that the device in question conforms to the general safety and performance requirements set out in Annex I and, where applicable, indicating which general safety and performance requirements have not been fully met, with the grounds.
- Where applicable, an indication that the device contains or incorporates a medicinal substance, including human blood or plasma derivatives, or tissues or cells of human or animal origin.
The statement accompanies the device. The manufacturer of the custom-made device submits the list of such devices made available on the market to the Agency upon request.
Retention and the Class III exception
The statement is kept for at least 10 years after the device has been placed on the market; for implantable devices, at least 15 years. Manufacturers of Class III custom-made implantable devices are subject, in addition to the Annex XIII procedure, to the conformity assessment specified in Chapter I of Annex IX; alternatively, the conformity assessment in Part A of Annex XI may be applied. Under the transitional provisions, these devices may be placed on the market or put into service without a notified body certificate until 26.05.2026, provided that a formal application was lodged with a notified body by 26.05.2024 and a written agreement was signed by 26.09.2024.
ÜTS side
In ÜTS, an application for a custom-made medical device under the MDR is made by centres whose field of activity is "Custom-Made Orthosis-Prosthesis Application Centre" or by healthcare institutions that have this manufacturing service in-house, for custom-made orthopaedic prostheses and orthoses; for other custom-made devices, it is made by companies that manufacture or import the device concerned and whose field of activity is "Manufacturer/Importer/Dealer/Exporter".
Common mistakes
- Treating every product measured to the person as a custom-made device; devices that are mass-produced and adapted do not fall under this definition.
- Not preparing the Annex XIII statement or not having it accompany the device.
- Skipping the additional conformity assessment for Class III custom-made implantable devices.
- Ignoring the retention periods of the statement.
Who it concerns: Custom-made prosthesis and orthosis centres, dental prosthesis producers, companies producing patient-specific implants and those who import these products. Ismarlama protez ve ortez merkezleri, diş protez üretimi yapan yerler, kişiye özel implant üreten firmalar ve bu ürünleri ithal edenler.
Common mistake: Treating devices that are mass-produced and adapted to the patient as custom-made devices and considering them exempt from the CE obligation. Seri üretilip hastaya uyarlanan cihazları ısmarlama cihaz sayıp CE yükümlülüğünden muaf kabul etmek.
Basis: Medical Device Regulation (OG 02.06.2021-31499 bis) ARTICLE 3, ARTICLE 20, ARTICLE 21, ARTICLE 52 and PROVISIONAL ARTICLE 1 and Annex XIII; the Agency's ÜTS Registration Guide (TCKKD-KLVZ-03, Rev.00) ARTICLE 29 and ARTICLE 45. Tıbbi Cihaz Yönetmeliği (RG 02.06.2021-31499 mükerrer) MADDE 3, MADDE 20, MADDE 21, MADDE 52 ve GEÇİCİ MADDE 1 ile Ek XIII; TİTCK ÜTS Kayıt Kılavuzu (TCKKD-KLVZ-03, Rev.00) MADDE 29 ve MADDE 45.