How Is a Recall Carried Out and Who Notifies It?

Recall is carried out and reported mainly by the manufacturer; importers, distributors and sales centres cooperate with the manufacturer, the authorised representative and the Agency to ensure that this corrective action is carried out…

ÜTS frequently asked questions: How Is a Recall Carried Out and Who Notifies It?

Recall is carried out and reported mainly by the manufacturer; importers, distributors and sales centres cooperate with the manufacturer, the authorised representative and the Agency to ensure that this corrective action is carried out. Recall is any measure aimed at achieving the return of a device that has already been made available to the end user to the economic operator. As a result of its evaluation, the Agency may also request corrective actions, including recall, from economic operators.

Difference between recall and withdrawal

In the Medical Device Regulation, recall is a measure aimed at achieving the return to the economic operator of a device that has been made available to the end user. Withdrawal is a measure aimed at preventing a device in the supply chain from being made available on the market. A field safety corrective action is a corrective action taken by the manufacturer for technical or medical reasons to prevent or reduce the risk of a serious adverse event in relation to a device made available on the market. A field safety notice is the communication sent by the manufacturer to users or customers in relation to this action.

Who needs to do what

  • Manufacturer: If it considers or has reason to believe that a device it has placed on the market or put into service is not in conformity with the Regulation, it promptly takes the appropriate corrective action, such as bringing the device into conformity, withdrawing it or recalling it, and informs its distributors and, where applicable, its authorised representative and importers. If the device presents a serious risk, it promptly informs the Agency, the relevant competent authorities and, if applicable, the notified body that issued the certificate
  • Importer: In case of non-conformity, promptly informs the manufacturer and the authorised representative; cooperates with the manufacturer, the authorised representative and the Agency to ensure that corrective action is carried out. If the device presents a significant risk, promptly informs the Agency and the notified body, giving details. Keeps a register of complaints, non-conforming devices, recalls and withdrawals
  • Distributor: Is subject to the same information and cooperation obligations; in case of a serious risk also promptly informs the Agency and keeps a register of recalls and withdrawals
  • Sales centre: Must immediately stop selling devices found to be unsafe or in breach of technical regulations within the scope of market surveillance and inspection and the warning system and declared by the Agency, return these devices to their manufacturer or importer, and share the relevant information and documents with the Agency on request

Reporting and announcement to users

Manufacturers report any field safety corrective action concerning devices lawfully made available on the market in Türkiye or EU member states through the electronic system. Except in urgent cases where action must be taken promptly, this is reported without delay before the action is carried out. The manufacturer ensures that information about the action is brought to the attention of the users of the device without delay by means of a field safety notice. The notice is drawn up in Turkish and the relevant official EU languages; except in urgent cases, the draft content is submitted to the Agency to allow comments. The notice includes the relevant UDIs so that the device and the manufacturer carrying out the action are correctly identified; it sets out clearly the reasons for the action and the level of risk without understating it, and specifies all actions to be taken by users. The manufacturer enters the notice in the electronic system and the notice is made accessible to the public.

The Agency's power to request and take measures

If, in its evaluation, the Agency finds that a device presents an unacceptable risk, it requires economic operators, within a clearly defined and notified reasonable period, to bring the device into conformity, to restrict its making available on the market in proportion to the nature of the risk, to withdraw it or to recall it. If the economic operator does not take adequate corrective action within this period, the Agency takes all appropriate measures, including withdrawing or recalling the device. Even in non-conformities that do not present an unacceptable risk, if the non-conformity is not ended within the given period, the Agency may take measures to ensure recall or withdrawal.

Common mistakes

  • Sending the field safety notice draft directly to users without submitting it to the Agency
  • Giving incomplete information on the affected lot, serial and UDI in the notice
  • Not stopping the sale of a device declared at the sales centre and keeping it in stock
  • Treating recall and withdrawal as the same and collecting only the products in the warehouse

Who it concerns: Manufacturers, authorised representatives, importers, distributors, sales centres, healthcare institutions and users. İmalatçılar, yetkili temsilciler, ithalatçılar, dağıtıcılar, satış merkezleri, sağlık kuruluşları ve kullanıcılar.

Common mistake: Leaving the devices held by end users out of scope and collecting only the stock in the supply chain. Son kullanıcıdaki cihazları kapsam dışı bırakıp yalnızca tedarik zincirindeki stoku toplamak.

Basis: Medical Device Regulation (OG 02.06.2021-31499 bis) ARTICLE 3, ARTICLE 10, ARTICLE 13, ARTICLE 14, ARTICLE 85, ARTICLE 87, ARTICLE 92, ARTICLE 94, ARTICLE 95; Medical Device Sales, Advertising and Promotion Regulation (OG 15.05.2014-29001) ARTICLE 31. Tıbbi Cihaz Yönetmeliği (RG 02.06.2021-31499 mükerrer) MADDE 3, MADDE 10, MADDE 13, MADDE 14, MADDE 85, MADDE 87, MADDE 92, MADDE 94, MADDE 95; Tıbbi Cihaz Satış, Reklam ve Tanıtım Yönetmeliği (RG 15.05.2014-29001) MADDE 31.

Notification and Follow-up

Medical Device ÜTS Registration Consultancy

Request an ÜTS registration consultancy quote

Ehlibeyt Mahallesi, Tekstilciler Caddesi No: 35/7 06520 Çankaya / Ankara - TÜRKİYE

Cevizli Mahallesi, Zuhal Caddesi Ritim İstanbul B Blok, No:44 Maltepe / İstanbul - TÜRKİYE

info@medexmedikal.com
+90 505 580 4622
+90 312 389 1354

ÜTS Guides · Legislation Calendar · Glossary · ÜTS Guide Index · Frequently Asked Questions

Services: ÜTS Consultancy · Who Carries Out the ÜTS Registration? · ÜTS Document Registration Consultancy · ÜTS Product Registration Consultancy · Importer ÜTS Registration Consultancy · Sales Centre Authorisation Certificate · EUDAMED Registration Consultancy · MDR/IVDR Declaration of Conformity · Technical File Preliminary Review · Label and Instructions for Use Check · PRRC Service · PMS and PSUR · Clinical Evaluation · EMDN and GMDN Code Determination · Certificate of Free Sale · EUDAMED XML Generator · ÜTS Registration Training · Correction of a Rejected ÜTS Application

KVKK and Privacy Policy | Cookie Policy | Accessibility Statement

© 2026 Medex Kurumsal Danışmanlık Ltd. Şti.

bottom of page
WhatsApp Hemen Ara