How Is the Risk Class of a Device Determined?
The risk class of a medical device is determined on the basis of the risks that the device carries by reason of its intended purpose and design.

The risk class of a medical device is determined on the basis of the risks that the device carries by reason of its intended purpose and design. Under Article 51 of the Medical Device Regulation, devices are divided into four classes, Class I, Class IIa, Class IIb and Class III, and classification is carried out according to the rules in Annex VIII of the Regulation. The manufacturer does the classification; the class determined also determines the conformity assessment procedure to be applied to the device.
Structure of the rule set
- Devices under the MDR are classified according to a total of 22 rules under the headings "Non-invasive Devices", "Invasive Devices", "Active Devices" and "Special Rules".
- Rules 1 to 4 apply to non-invasive devices, Rules 5 to 8 to invasive devices, Rules 9 to 13 to active devices, and Rules 14 to 22 to special cases.
- For devices under the repealed Medical Device Directive (MDD), the number of rules is 18. For devices on the market with an old document, attention must be paid to which rule set is used.
Basic principles in applying the rules
- Application of the rules is based on the intended purpose of the device.
- If the device is intended to be used in combination with another device, the rules apply to each device separately. Accessories are classified in their own right, separately from the device with which they are used.
- Software that drives a device or influences the use of a device falls within the same class as the device. If the software is independent of any device, it is classified in its own right.
- If the device is not intended to be used solely or principally in a specific part of the body, it must be considered and classified on the basis of the most critical specified use.
- If more than one rule, or more than one sub-rule within the same rule, can be applied to the same device, the strictest rule resulting in the higher classification applies.
Definitions of duration of use
- Transient: normally intended for continuous use for less than 60 minutes.
- Short term: normally intended for continuous use for between 60 minutes and 30 days.
- Long term: normally intended for continuous use for more than 30 days.
- In calculating continuous use, temporary interruptions of use during a procedure, or temporary removal of the device for purposes such as cleaning or disinfection, are not taken into account; for devices intended by the manufacturer to be replaced promptly by another device of the same type, total use is the basis.
Reflection of the class in ÜTS and disputes
- Under the MDR in ÜTS there are the options "Class-I Other", "Class-Ir", "Class-Im", "Class-Is", "Class-IIa", "Class-IIb" and "Class-III". All of these sub-distinctions, made according to sterile packaging, measuring function and reusable surgical instrument features, are Class I for the purposes of the Regulation.
- Disputes between notified bodies and a manufacturer established in Türkiye arising from the application of Annex VIII are referred to the Agency for a decision.
Common mistakes
- Determining the class by looking at the trade name of the device or the class of similar products on the market; the rules apply to the device's own intended purpose.
- Preferring the rule that gives the lower class when more than one rule can be applied.
- Counting the accessory within the class of the main device.
- Calculating duration of use from the duration of a single application rather than total continuous use; taking the sub-distinction in ÜTS to be a separate risk class.
Who it concerns: Medical device manufacturers, importers, authorised representatives and companies registering devices in ÜTS. Tıbbi cihaz imalatçıları, ithalatçılar, yetkili temsilciler ve ÜTS'de cihaz kaydı yapan firmalar.
Common mistake: Classifying by product type, taking the class of similar products as the basis instead of the device's declared intended purpose. Sınıflandırmayı ürün tipine göre yapmak, cihazın beyan edilen kullanım amacı yerine benzer ürünlerin sınıfını esas almak.
Basis: Medical Device Regulation (OG 02.06.2021-31499 bis) ARTICLE 51 and Annex VIII; TİTCK Guide on Registration Processes of Medical Devices in the Product Tracking System (TCKKD-KLVZ-03, Rev.00, effective 13.05.2026) ARTICLE 51, ARTICLE 52, ARTICLE 53 and ARTICLE 56. Tıbbi Cihaz Yönetmeliği (RG 02.06.2021-31499 mükerrer) MADDE 51 ve Ek VIII; TİTCK Tıbbi Cihazların Ürün Takip Sistemine Kayıt Süreçlerine İlişkin Kılavuz (TCKKD-KLVZ-03, Rev.00, yürürlük 13.05.2026) MADDE 51, MADDE 52, MADDE 53 ve MADDE 56.