Is a Different Barcode of the Same Product Registered?
As a rule, no. Device registration in ÜTS is made with the single barcode assigned to the smallest package placed on the market (the unit package or single package).

As a rule, no. Device registration in ÜTS is made with the single barcode assigned to the smallest package placed on the market (the unit package or single package). No separate record is opened for different package sizes of the same device; after the device registration is completed, these barcodes are added with their content quantities to the "Market Supply Information" field. A new barcode is assigned only in situations that change the identification of the device.
How multiple packages are notified
The barcodes of sales packages containing multiple products, formed by combining several devices registered with the barcode assigned to the smallest package, are added through the "Add Market Supply Information" screen in the "Market Supply Information" field on the relevant device's screen. The same can be done in bulk through the "Bulk Add Market Supply Information" screen in the "Bulk Operations" field of the "List My Medical Devices" screen. For this, the device must first have reached "Registered" status.
When a new UDI-DI is required
When there is a change that could lead to the device being wrongly identified and/or to ambiguity in its traceability, a new UDI-DI number must be assigned to the resulting new device. The changes listed in the Guide are:
- A change of the device name or trade name.
- A change of the device version/model.
- A change relating to whether it is single-use.
- A change relating to whether it is packed sterile.
- A change relating to whether it needs to be sterilised before use.
- A change in the quantity of devices in the package/box.
- Changes relating to critical warnings or contraindications.
In addition, any change in trade names resulting from translation into the languages of the EU countries where the device is placed on the market also leads to the affected device being placed on the market with a new UDI-DI.
The same barcode at different companies
Different companies may register with the same barcode. In the "Other Products with the Same Barcode" screen opened from the "Products with the Same Barcode" link on the registration screen, the product records of other registered companies are listed with the UDI-DI number entered. If there are data differences between these records, the relevant data fields must be made identical and consistent in ÜTS before the device application. For this, an imported product data change request is created; the request is submitted for opinion to the other companies with a record with the same barcode, and the period for giving an opinion is 5 days.
As regards the brand, the rule is strict: devices with the same UDI-DI number cannot be registered in the system under different brand names. If it is found that the same device has been registered by different companies under different brand names, action is taken on the unsuitable records as a result of the Agency's evaluation.
Different barcode in the transition from the old regulation
If a device is registered with a different barcode under the MDD or AIMDD and will be registered with a new barcode under the MDR, the field "Did it transition to the MDR with a different product number?" on the registration screen is selected as "Yes" and the primary product number of the previous record is selected in the field that opens. Only one barcode can be selected in this field.
Common mistakes
- Opening three separate device records for the single, 10-pack and carton packages of the same device.
- Continuing the registration with the old barcode even though the quantity of devices in the package has changed.
- Making a device application without resolving the data differences in other companies' records with the same barcode.
- Registering a device with the same barcode under your own brand.
Who it concerns: Manufacturers that supply the same device in different packages and multiple companies importing the same product. Aynı cihazı farklı ambalajlarda arz eden imalatçılar ve aynı ürünü ithal eden birden fazla firma.
Common mistake: Thinking that a change in package size is a change that requires a new device record. Ambalaj büyüklüğü değişikliğini yeni cihaz kaydı gerektiren bir değişiklik sanmak.
Basis: Guide on Registration Processes of Medical Devices in the Product Tracking System (TCKKD-KLVZ-03, Rev.00, effective 13.05.2026) ARTICLE 27, ARTICLE 36, ARTICLE 43 and ARTICLE 44. Tıbbi Cihazların Ürün Takip Sistemine Kayıt Süreçlerine İlişkin Kılavuz (TCKKD-KLVZ-03, Rev.00, yürürlük 13.05.2026) MADDE 27, MADDE 36, MADDE 43 ve MADDE 44.