top of page

What Transition Periods End on 31 December 2027?

31.12.2027 is the date on which the transition period ends for the highest-risk device groups. On the MDR side, all Class III devices and, apart from certain exceptions, Class IIb…

ÜTS frequently asked questions: What Transition Periods End on 31 December 2027?

31.12.2027 is the date on which the transition period ends for the highest-risk device groups. On the MDR side, all Class III devices and Class IIb implantable devices (other than certain exceptions), and on the IVDR side, devices in Class D, may be placed on the market or put into service until this date. After this date, these devices cannot be placed on the market without completing conformity assessment under the new legislation.

Scope on the MDR side

Under sub-paragraph (1) of paragraph (b) of the fifth paragraph of PROVISIONAL ARTICLE 1 of the Medical Device Regulation, the devices that may be placed on the market or put into service until 31.12.2027 are as follows:

  • All Class III devices.
  • Class IIb implantable devices, except sutures, staples, dental fillings, dental braces, tooth crowns, screws, wedges, plates, wires, pins, clips and connectors.

For the exempted products listed in this sub-paragraph, other Class IIb devices, Class IIa devices, and sterile or measuring Class I devices, the date is 31.12.2028.

Scope on the IVDR side

The same date also applies to in vitro diagnostic devices:

  • Devices placed on the market as Annex 2 List A, Annex 2 List B and self-testing devices under the repealed legislation may be placed on the market or put into service until 31.12.2027.
  • The date is also 31.12.2027 for devices that did not require a notified body under the old legislation, for which a declaration of conformity was drawn up before 26.05.2022, and that will be classified as Class D under the new legislation.

What happens after the date

  • After this date, these devices cannot be placed on the market or put into service; the period during which they remain registered in ÜTS also depends on this date.
  • On the other hand, under the sixth paragraph of PROVISIONAL ARTICLE 1, devices legally placed on the market under the Regulations repealed before 26.05.2021, and devices legally placed on the market from 26.05.2021 under the transitional provisions, may continue to be made available on the market or put into service.
  • This is why the date draws a line between the fate of products already in stock and in the supply chain and new supply.

What must be completed by the date

For a device to be placed on the market after 31.12.2027, the conformity assessment procedure appropriate to its risk class under the new Regulation must be completed with a notified body, the certificate must be issued, the declaration of conformity must be updated, and the document and device registration must be updated in ÜTS in Türkiye. To benefit from the transition period, a formal application to a notified body by 26.05.2024 and a written agreement by 26.09.2024 are also required.

Common mistakes

  • Assuming 31.12.2027 is a single final date valid for all devices; there are also 31.12.2028 by class and 31.12.2029 on the IVD side.
  • Overlooking the list of exempted products in the sub-paragraph for Class IIb implantable devices.
  • Treating placing on the market and making available on the market as the same thing.
  • Leaving the new certificate process until the transition date approaches.

Who it concerns: Manufacturers of Class III and Class IIb implantable devices, IVD companies, importers and the supply units of healthcare institutions. Sınıf III ve Sınıf IIb implante edilebilir cihaz imalatçıları, IVD firmaları, ithalatçılar ve sağlık kuruluşlarının tedarik birimleri.

Common mistake: Applying the date as a single final date for the entire product portfolio and not distinguishing by class. Tarihi tüm ürün portföyü için tek bir son tarih olarak uygulamak ve sınıf bazlı ayrımı yapmamak.

Basis: Medical Device Regulation (OG 02.06.2021-31499 bis) PROVISIONAL ARTICLE 1; TİTCK ÜTS Registration Guide (TCKKD-KLVZ-03, Rev.00) ARTICLE 45 and ARTICLE 48. Tıbbi Cihaz Yönetmeliği (RG 02.06.2021-31499 mükerrer) GEÇİCİ MADDE 1; TİTCK ÜTS Kayıt Kılavuzu (TCKKD-KLVZ-03, Rev.00) MADDE 45 ve MADDE 48.

MDR and IVDR Transition

Medical Device ÜTS Registration Consultancy

Request an ÜTS registration consultancy quote

Ehlibeyt Mahallesi, Tekstilciler Caddesi No: 35/7 06520 Çankaya / Ankara - TÜRKİYE

Cevizli Mahallesi, Zuhal Caddesi Ritim İstanbul B Blok, No:44 Maltepe / İstanbul - TÜRKİYE

info@medexmedikal.com
+90 505 580 4622
+90 312 389 1354

ÜTS Guides · Legislation Calendar · Glossary · ÜTS Guide Index · Frequently Asked Questions

Services: ÜTS Consultancy · Who Carries Out the ÜTS Registration? · ÜTS Document Registration Consultancy · ÜTS Product Registration Consultancy · Importer ÜTS Registration Consultancy · Sales Centre Authorisation Certificate · EUDAMED Registration Consultancy · MDR/IVDR Declaration of Conformity · Technical File Preliminary Review · Label and Instructions for Use Check · PRRC Service · PMS and PSUR · Clinical Evaluation · EMDN and GMDN Code Determination · Certificate of Free Sale · EUDAMED XML Generator · ÜTS Registration Training · Correction of a Rejected ÜTS Application

KVKK and Privacy Policy | Cookie Policy | Accessibility Statement