What Is Basic UDI-DI? Definition, Documents It Appears In and ÜTS Selection

SUMMARY
The Basic UDI-DI is the primary identifier of a device model. It appears in the EU declaration of conformity and in the relevant EU certificates; in EUDAMED and ÜTS it links devices of the same manufacturer that have the same intended purpose, risk class and essential design characteristics. It does not appear on the device label or the sales packaging.
This article covers only the Basic UDI-DI. For the UDI system as a whole, including the UDI-DI, UDI-PI and UDI carrier, see our article on the UDI system.
For the Basic UDI-DI in your declaration of conformity and ÜTS product registration: Medical Device ÜTS Registration Consultancy
Contents
1. What Is the Basic UDI-DI?2. In Which Documents Does the Basic UDI-DI Appear?3. Who Assigns the Basic UDI-DI and When?4. Issuing Entities and Code Structure5. Basic UDI-DI in EUDAMED6. How Is the Basic UDI-DI Selected in ÜTS?7. Same Application and Free Sale Certificate Rule8. Frequently Asked Questions
1. What Is the Basic UDI-DI?
Definition in Annex VI Part C of the Medical Device Regulation:
“The Basic UDI-DI is the primary identifier of a device model. It is the DI assigned at the level of the unit of use of the device. It is the main key for records in the UDI database and is referenced in relevant certificates and EU declarations of conformity.”
Definition in Article 4 of ÜTS Registration Guideline TCKKD-KLVZ-03:
“ggg) Basic UDI-DI: The identifier code assigned at the level of the unit of use of the device, as the main key in the database and in the relevant documentation (e.g. EU certificates, EU declaration of conformity, technical documentation and summary of safety and clinical performance), in order to establish a link with devices that have the same intended purpose, risk class and essential design and manufacturing characteristics.”
In short, the Basic UDI-DI groups the devices forming the same family under a single key in documents and databases. It must not be confused with the UDI-DI carried on the label.
2. In Which Documents Does the Basic UDI-DI Appear?
EU Declaration of Conformity
“(6) The Basic UDI-DI of the device, as defined in Annex VI Part C, shall be included in the EU declaration of conformity referred to in Article 19.”
Annex IV also lists the Basic UDI-DI among the content of the EU declaration of conformity:
“3. The Basic UDI-DI as referred to in Annex VI Part C,”
“Information allowing identification and traceability, excluding the product name or trade name, may be provided by the Basic UDI-DI referred to in item 3 of this Annex.”
For IVD devices, the same rule also exists in the Regulation on In Vitro Diagnostic Medical Devices:
“(6) The Basic UDI-DI of the device, as defined in Annex VI Part C, shall be included in the EU declaration of conformity referred to in Article 18.”
For the preparation of the declaration of conformity, see our article on the manufacturer's declaration of conformity.
EU Certificates
“a) EU technical documentation assessment certificates, EU type-examination certificates and EU product verification certificates shall contain a clear identification of the device or devices, including name, model and type, as included by the manufacturer in the instructions for use, the intended purpose, the risk class, and the Basic UDI-DI as referred to in the sixth paragraph of Article 27, relating to the conformity assessment procedure to which the device is subject.”
The ÜTS Registration Guideline gives this distinction by ÜTS document types (Article 27, fifth paragraph, subparagraph (c)):
“5) As also mentioned in subparagraph (h) of the second paragraph of Article 7 of this Guideline, the Basic UDI-DI information of the device is mandatory in EU certificates of the types "EU Technical Documentation Certificate", "EU Type-Examination Certificate" and "EU Product Verification Certificate". The Basic UDI-DI information is not mandatory in EU certificates of the types "EU Quality Management System Certificate" and "EU Production Quality Assurance Certificate".”
Summary of Safety and Clinical Performance and Free Sale Certificate
“a) Information identifying the device and the manufacturer, including the Basic UDI-DI and, if already issued, the single registration number.”
This subparagraph is the minimum content of the summary of safety and clinical performance in Article 32 of the Medical Device Regulation. The Basic UDI-DI is also included in the free sale certificate (Article 60):
“In the free sale certificate, the Basic UDI-DI of the device entered into the UDI database under Article 29 is stated.”
Label and Sales Packaging
“7) The Basic UDI-DI is not included on the market placement label and/or sales packaging of the device.”
The code carried on the label is the UDI, made up of the UDI-DI and the UDI-PI. For the carrier and barcode format, see our article on the UDI carrier and barcode format.
3. Who Assigns the Basic UDI-DI and When?
Assignment is the manufacturer's obligation (Article 29 of the Medical Device Regulation):
“ARTICLE 29- (1) Before placing a device other than a custom-made device on the market, the manufacturer shall assign to the device, in accordance with the rules of the issuing entities referred to in the second paragraph of Article 27, a Basic UDI-DI as defined in Annex VI Part C, and shall enter it in the UDI database together with the other core data elements referred to in Annex VI Part B relating to that device.”
For devices going through a notified body assessment, assignment is made before the application:
“(3) For devices subject to a conformity assessment referred to in the third paragraph of Article 52 and in subparagraphs (b) and (c) of the fourth paragraph of Article 52, the Basic UDI-DI assignment referred to in the first paragraph of this Article shall be carried out before the manufacturer applies to a notified body for that assessment.”
For systems and procedure packs (SPP), assignment belongs to the responsible person who places the pack on the market (second paragraph of Article 29). The ÜTS Registration Guideline explains the purpose for SPPs as follows:
“2) For SPPs, the Basic UDI-DI is important for establishing a link between SPPs of the same SPP manufacturer that have the same group of components and the same intended purpose, independently of the original component manufacturers.”
4. Issuing Entities and Code Structure
“8) The Basic UDI-DI issuing entities and standards registered in ÜTS under the MDR and authorised for medical devices are GS1, HIBCC, ICCBBA and IFA.”
“10) According to the Basic UDI-DI assignment rules, the Basic UDI-DI code value that the issuing entities may assign consists of at most 25 characters.”
“11) Based on the algorithm determined by the issuing entities, a check digit/character must be a part of the Basic UDI-DI. This algorithm is provided to the Commission and to manufacturers by the issuing entities.”
For the differences between the entities, see our article on UDI issuing entities: GS1, HIBCC, ICCBBA and IFA.
5. Basic UDI-DI in EUDAMED
Annex VI Part A of the Medical Device Regulation starts the device information to be submitted at device registration with the Basic UDI-DI ("2.1. Basic UDI-DI"). Among the core data elements to be entered together with the UDI-DI in Annex VI Part B is the following item:
“2. The Basic UDI-DI and any additional UDI-DIs, as referred to in Article 29,”
For the registration steps in EUDAMED, see our article on EUDAMED device registration.
6. How Is the Basic UDI-DI Selected in ÜTS?
In a product registration under the MDR, the Basic UDI-DI is under the "Product Documents" heading (TCKKD-KLVZ-03, Article 27, fifth paragraph, subparagraph (c)):
“1) This field is a mandatory field to be selected in ÜTS.”
“3) Before making a selection from the "Select Basic UDI-DI" screen in this field, it is mandatory to select, from the field mentioned in subparagraph (a) of this paragraph, the documents (EU Declaration of Conformity, EU Certificate) related to the relevant device and to the Basic UDI-DI of the device, in accordance with the document business rule mentioned in Article 66 of this Guideline.”
“4) Through the "Select Basic UDI-DI" screen in this field, the Basic UDI-DI identifying the relevant device is selected only from the Basic UDI-DI numbers associated with the EU Declaration of Conformity selected from the field mentioned in subparagraph (a) of this paragraph. More than one selection cannot be made in this field.”
The order is as follows:
In the product registration, the device's EU declaration of conformity registered in ÜTS and, if any, its EU certificate are selected from the "Documents" field.
On the "Select Basic UDI-DI" screen, only the Basic UDI-DIs associated with the selected declaration are listed.
The single Basic UDI-DI that identifies the device is selected.
The same field also exists in product registration under the IVDR (Article 30). For the other fields of product registration, see our article on ÜTS product registration.
7. Same Application and Free Sale Certificate Rule
In an ÜTS device application under the new regulations, all devices must be linked to the same Basic UDI-DI (TCKKD-KLVZ-03, Article 40):
“(8) In device applications under the new Medical Device Regulations (excluding ITCs), applications are permitted only for devices linked to the same Basic UDI-DI code.”
For an example of a rejection arising from this rule, see our page on devices linked to different Basic UDI-DIs in the same application. The same rule exists in a free sale certificate application (Article 42):
“A separate free sale certificate application is made for devices with different Basic UDI-DIs.”
The Master UDI-DI used for products such as spectacle frames and lenses is a separate structure. For details, see our article on the Master UDI-DI.
8. Frequently Asked Questions
Does the Basic UDI-DI appear on the label?
No. According to the ÜTS Registration Guideline, the Basic UDI-DI is not included on the device's market placement label and sales packaging. The label carries the UDI made up of the UDI-DI and the UDI-PI.
In which documents is the Basic UDI-DI mandatory?
It is mandatory in the EU declaration of conformity and in EU technical documentation, EU type-examination and EU product verification certificates. It is not mandatory in EU quality management system and EU production quality assurance certificates.
Who assigns the Basic UDI-DI?
The manufacturer assigns it, according to the rules of the issuing entity, before placing the device on the market. For devices going through a notified body assessment, assignment is made before applying to the notified body.
Can more than one Basic UDI-DI be selected in a product registration?
No. The "Select Basic UDI-DI" screen lists only the Basic UDI-DIs associated with the selected EU declaration of conformity, and more than one selection cannot be made.
Can devices with different Basic UDI-DIs be put in the same ÜTS application?
No. In device applications under the new regulations (excluding ITCs), applications can be made only for devices linked to the same Basic UDI-DI code.
For a declaration of conformity and ÜTS product registration consistent with the Basic UDI-DI: Medical Device ÜTS Registration Consultancy
Sources
Legislation: Medical Device Regulation, Articles 27, 29, 32 and 60, Annex IV, Annex VI and Annex XII
Legislation: Regulation on In Vitro Diagnostic Medical Devices, Article 25
TİTCK: ÜTS Registration Guideline TCKKD-KLVZ-03, Articles 4, 27, 28, 30, 40 and 42