Products Excluded in the MDR Transition and ÜTS: Sutures, Screws, Plates and Others

SUMMARY
Regulation (EU) 2023/607 provides the date 31.12.2027 for Class IIb implantable devices, but leaves a list of products outside this date. These products are sutures, staples, dental fillings, dental braces, tooth crowns, screws, wedges, plates, wires, pins, clips and connectors.
The most common mistake is tying the products on the list to the date 31.12.2027. The Regulation and the TİTCK guideline tie these products to 31.12.2028, together with the other Class IIb devices. Below we give the relevant provisions literally (unofficial translations of the Turkish guideline text).
Let us determine together which transition date your product is subject to: Medical Device ÜTS Registration Consultancy
Contents
1. List of Excluded Products2. Which Date Do the Excluded Products Fall Under?3. The Regulation's Own Wording4. The Situation for Class III Products5. Conditions for Benefiting from the Date6. How Does the Extension in Türkiye Reflect in ÜTS?7. Frequently Asked Questions
1. List of Excluded Products
The list is in the fourth paragraph of Article 45 of TİTCK ÜTS Registration Guideline TCKKD-KLVZ-03:
“(4) Provided that the conditions specified in the Regulation mentioned in the first paragraph of this article are fulfilled, Class-III devices under the MDD or AIMDD and Class-IIb implantable devices, except sutures, staples, dental fillings, dental braces, tooth crowns, screws, wedges, plates, wires, pins, clips and connectors, may be placed on the market or put into service until 31.12.2027.”
Sutures
Staples
Dental fillings
Dental braces
Tooth crowns
Screws
Wedges
Plates
Wires
Pins
Clips
Connectors
2. Which Date Do the Excluded Products Fall Under?
Since the products on the list remain outside the scope of the fourth paragraph, they fall under the fifth paragraph:
“(5) Provided that the conditions specified in the Regulation mentioned in the first paragraph of this article are fulfilled, Class-IIb devices under the MDD, other than those covered by the fourth paragraph of this article, and Class-IIa devices, Class-Im devices and Class-Is devices, may be placed on the market or put into service until 31.12.2028.”
In other words, when the conditions are met, the excluded products may be placed on the market until 31.12.2028, instead of the 31.12.2027 foreseen for Class IIb implantable devices.
3. The Regulation's Own Wording
The same distinction is also in paragraph 3a of Article 120 of the MDR as amended by Regulation (EU) 2023/607:
“(a) 31 December 2027, for all class III devices, and for class IIb implantable devices except sutures, staples, dental fillings, dental braces, tooth crowns, screws, wedges, plates, wires, pins, clips and connectors;”
“(b) 31 December 2028, for class IIb devices other than those covered by point (a) of this paragraph, for class IIa devices, and for class I devices placed on the market in sterile condition or having a measuring function.”
4. The Situation for Class III Products
The list is an exception only for Class IIb implantable devices. The fourth paragraph of the guideline foresees the date 31.12.2027 for "Class-III devices under the MDD or AIMDD"; if a product is Class III, it is subject to this date even if it is named in the list. To determine the class of your product, see our article on MDR risk classes.
5. Conditions for Benefiting from the Date
The dates do not operate automatically. The guideline uses the expression "provided that the conditions specified in the Regulation are fulfilled" in every paragraph. The conditions are in the manufacturer's declaration; for example:
“b) A statement that there has been no significant change in the design and intended purpose of the relevant device or devices,”
“ç) A statement that a quality management system (QMS) will be/has been put in place in accordance with MDR Article 10(9), no later than 26.05.2024,”
For all the conditions and the date table by class, see our article on the (EU) 2023/607 MDR transition timeline.
6. How Does the Extension in Türkiye Reflect in ÜTS?
The procedure is the same for the excluded products: first the "Time Extension Application No. 2023/KK-1" in EBS and, if the Agency finds it appropriate, the "Document Application under Announcement No. 2023/KK-1" in ÜTS for the EC certificate. The guideline writes that, in an application found appropriate, the validity date of the certificate will be extended "until 31.12.2027 or 31.12.2028, according to the risk class of the device". For the steps, see our article on the MDR time extension application.
7. Frequently Asked Questions
Are sutures and screws subject to the date 31.12.2027?
No. These products are left outside the date 31.12.2027 foreseen for Class IIb implantable devices. When the conditions are met, they may be placed on the market until 31.12.2028.
Why have the products on the list been kept separate?
The Regulation and the guideline do not state a reason; they only list the products. Therefore, which date a product is subject to is decided by looking at the list and its class.
What is the date of a Class IIb implantable device that is not on the list?
Class IIb implantable devices that are not on the list are subject to 31.12.2027.
What if the excluded product is Class III?
The exception is only for Class IIb implantable devices. Class III devices are subject to 31.12.2027.
For the preparation of your EBS and ÜTS applications: Medical Device ÜTS Registration Consultancy
Sources
TİTCK: ÜTS Registration Guideline TCKKD-KLVZ-03, Article 45
EUR-Lex: Regulation (EU) 2023/607