MDR Registration with a Different Barcode in ÜTS: How Does a Device with an MDD Record Transition?

SUMMARY
A device registered in ÜTS under the MDD or AIMDD may receive a new barcode (UDI-DI) when moving to the MDR. In this case, the barcode of the old record is specified separately in the MDR record. TİTCK ÜTS Registration Guideline TCKKD-KLVZ-03 defines two fields for this.
In this article we explain these fields and the changes that require a new barcode through literal quotations (unofficial translations) from the guideline.
So that your MDR records are opened correctly in ÜTS: Medical Device ÜTS Registration Consultancy
Contents
1. The "Did It Transition to the MDR with a Different Product Number?" Field2. How Is the Old Barcode Selected?3. What Is Selected If the Barcode Stayed the Same?4. Which Changes Require a New UDI-DI?5. The New Barcode Must Match the Documents6. Check of Products with the Same Barcode7. Frequently Asked Questions
1. The "Did It Transition to the MDR with a Different Product Number?" Field
This field is mandatory in the "Identifying Information" section of a device record under the MDR:
“c) Did it transition to the MDR with a different product number?;”
“2) This field is selected as "Yes" if the relevant device requested to be registered in ÜTS under the MDR was previously registered in ÜTS under the MDD or AIMDD with a different barcode, and as "No" if it was not previously registered in ÜTS under the MDD or AIMDD with a different barcode.”
2. How Is the Old Barcode Selected?
If the field is selected as "Yes", the area for selecting the old barcode opens on the screen:
“ç) The Product Number under 93/42/EEC or 90/385/EEC;”
“2) Using the list on the "Select Medical Device" screen in this field, the primary product number (barcode) of the relevant device is selected from among the device records previously registered in ÜTS under the MDD or AIMDD. To make the relevant barcode selection, a query can be made with the "Primary Product Number", "Brand Name", "Product Name" and "Version/Model" fields on the relevant screen.”
“3) Only one barcode can be selected in this field. More than one selection cannot be made.”
3. What Is Selected If the Barcode Stayed the Same?
If the device transitions to the MDR with the same barcode, or if it was not previously registered in ÜTS under the MDD or AIMDD, the field is selected as "No". In this case, the old barcode field does not open.
4. Which Changes Require a New UDI-DI?
The guideline explicitly lists the situations that require a new barcode:
“7) When there is a change that may lead to the incorrect identification of the device and/or to uncertainty in its traceability, a new UDI-DI Number must be allocated for the resulting new device.”
A change in the device name or trade name
A change in the device version or model
A change related to whether it is single-use or not
A change related to whether it is packaged sterile or not
A change related to whether it must be sterilized before use or not
A change in the quantity of devices in the package or box
A change related to critical warnings or contraindications (such as latex, DEHP, CMR or endocrine-disrupting substances)
5. The New Barcode Must Match the Documents
The UDI-DI entered in the MDR record must be the same as it appears in the label, the EU declaration of conformity and the EU certificate:
“4) The device application made under Article 40 of this Guideline for a device record whose barcode has not been entered into ÜTS as it appears in the technical documents (label/sales packaging, EU declaration of conformity, EU certificate, etc.) is concluded negatively.”
For barcode issuing entities and format rules, see our article on UDI issuing entities.
6. Check of Products with the Same Barcode
On the MDR registration screen, there is a link to see the records of other companies with the same UDI-DI:
“1) In this field, the list of records of devices previously registered in ÜTS by other companies with the same UDI-DI Number (barcode) can be viewed.”
If there are data differences between records with the same barcode, these differences must be resolved before the first device application. For the process, see our article on the imported product data change request.
7. Frequently Asked Questions
When is the "Did it transition to the MDR with a different product number?" field selected as "Yes"?
It is selected as "Yes" if the device was previously registered in ÜTS under the MDD or AIMDD with a different barcode.
Can more than one barcode be entered in the old barcode field?
No. According to the guideline, only one barcode can be selected in this field.
Is a new barcode needed if the name of the device changes?
Yes. The guideline lists a change in the device name or trade name among the situations that require a new UDI-DI.
What happens if the barcode in the MDR record differs from the documents?
The application made for a device record whose barcode has not been entered as it appears in the technical documents is concluded negatively.
For the preparation and application of your MDR records: Medical Device ÜTS Registration Consultancy
Sources
TİTCK: ÜTS Registration Guideline TCKKD-KLVZ-03, Article 27