Master UDI-DI and ÜTS: MDCG 2025-8 for Spectacle Frames, Lenses and Ready-to-Wear Reading Spectacles

12 hours ago
4 min read
Master UDI-DI gözlük çerçevesi cam ve okuma gözlüğü MDCG 2025-8
  • SUMMARY

    Master UDI-DI is a type of UDI-DI defined for devices that vary greatly from person to person, such as spectacle frames, spectacle lenses and ready-to-wear reading spectacles. Its legal basis is Commission Delegated Regulation (EU) 2025/1920; its application is explained in the MDCG 2025-8 guidance.

    In this article we give the assignment rules and dates of MDCG 2025-8 Rev.1 (March 2026). The EU texts are quoted in their English original, with an explanation added below. For the ÜTS registration of spectacle frames, see our spectacle frame ÜTS registration page.

    For the EUDAMED and ÜTS registrations of your spectacle products: Medical Device ÜTS Registration Consultancy

    Contents

    1. What Is a Master UDI-DI?2. Basic UDI-DI, Master UDI-DI and UDI-PI3. Spectacle Frames4. Spectacle Lenses5. Ready-to-Wear Reading Spectacles6. Dates: When Does It Become Mandatory?7. The Situation in ÜTS8. Frequently Asked Questions

    1. What Is a Master UDI-DI?

    “Master UDI-DI: Unique identifier used for grouping certain devices presenting a high level of individualisation ('highly individualised devices', HID) as for spectacle frames, spectacle lenses and ready-to-wear reading spectacles as defined in Commission Delegated Regulation (EU) 2025/1920. It is acting as UDI-DI as defined in Part C of Annex VI to the MDR.”

    In other words: the Master UDI-DI is the unique identifier used to group devices with a high level of individualisation, and it performs the UDI-DI function in Part C of MDR Annex VI.

    2. Basic UDI-DI, Master UDI-DI and UDI-PI

    • Basic UDI-DI: appears in the EU declaration of conformity, does not appear on the device label, and is registered in EUDAMED.

    • Master UDI-DI: appears on the device label and is registered in EUDAMED.

    • UDI-PI: appears on the label (such as lot number and manufacturing date).

    “1 Master UDI-DI can be linked to only 1 Basic UDI-DI but 1 Basic UDI-DI can be linked to multiple Master UDI-DIs”

    A Master UDI-DI is linked to only one Basic UDI-DI; more than one Master UDI-DI can be linked to one Basic UDI-DI.

    3. Spectacle Frames

    For frames, the Basic UDI-DI is determined by two parameters: the frame structure (full rim, semi-rimless, rimless, other) and the frame material (metal, plastic, combination, other). The additional parameter for the Master UDI-DI is the lens size:

    “A UDI-DI shall be assigned to spectacle frames that have the same combination of design parameters, including at least the horizontal boxed lens size ('Master UDI-DI').”

    “Master UDI-DI design parameter is the horizontal boxed lens size with discrete values in a typical range between 40 mm and 60 mm (assignment triggers).”

    4. Spectacle Lenses

    For lenses, the Basic UDI-DI is determined by two parameters: the commercial index and lens material (organic or mineral), and the design (single vision, multifocal, progressive, other). The Master UDI-DI, on the other hand, is assigned according to groups of vision impairment:

    “A UDI-DI shall be assigned to spectacle lenses that have the same combination of design parameters, including at least groups of mean spheres (spherical equivalent power), groups of addition power and groups of similar vision impairments ('Master UDI-DI').”

    The spherical equivalent power is calculated with the following formula:

    “M = Spherical power in dioptres + (cylinder in dioptres/2)”

    According to the result, the lens falls into one of six vision impairment groups: M less than zero (myopia), zero (emmetropia) or greater than zero (hyperopia), each with or without cylinder. For presbyopia, the addition values are divided into three groups:

    “Mild/Low presbyopia with a typical nominal prescribed addition value in range (assignment trigger) between > 0.00 D and < 1.25 D”

    “Moderate presbyopia with a typical nominal prescribed addition value in range (assignment trigger) between ≥ 1.25 D and < 2.00 D”

    “Advanced/High presbyopia with a typical nominal prescribed addition value in range (assignment trigger) ≥ 2.00 D”

    5. Ready-to-Wear Reading Spectacles

    The Basic UDI-DI is determined by three parameters: frame structure, frame material and lens material (mineral, or organic including polycarbonate). The Master UDI-DI parameters:

    “A UDI-DI shall be assigned to ready-to-wear reading spectacles that have the same combination of design parameters, including at least the horizontal boxed lens size and lens spherical power ('Master UDI-DI').”

    “Lens spherical power with discrete values in the range: +1.00 D to +3.50 D (assignment triggers)”

    6. Dates: When Does It Become Mandatory?

    “Commission Delegated Regulation (EU) 2025/1920 on Master UDI-DI for spectacle frames, spectacle lenses and ready-to-wear reading spectacles was published in the Official Journal of the European Union (OJEU) on 23/09/2025 and entered into force on 13/10/2025. It applies from 01/11/2028, providing three-years transitional period as from its entry into force before the assignment of Master UDI-DI to spectacle frames, spectacle lenses and ready-to-wear reading spectacles labelling becomes mandatory.”

    “Spectacle frames, spectacle lens, ready to wear reading spectacles produced prior to 1st of November 2028 are not required to have a Master UDI-DI on the label.”

    In other words: the Regulation was published on 23/09/2025, entered into force on 13/10/2025 and applies from 01/11/2028. Products manufactured before this date are not required to carry a Master UDI-DI on their label.

    7. The Situation in ÜTS

    The expression Master UDI-DI does not appear in the text of TİTCK ÜTS Registration Guideline TCKKD-KLVZ-03. In ÜTS, device registration under the MDR is today made with the "UDI-DI Number" field in Article 27 of the guideline, and the barcode must be entered as it appears on the label and in the documents. For barcode rules, see our article on UDI issuing entities.

    8. Frequently Asked Questions

    When will the Master UDI-DI become mandatory?

    Regulation (EU) 2025/1920 applies from 01/11/2028. According to MDCG 2025-8, spectacle frames, lenses and ready-to-wear reading spectacles produced before this date are not required to carry a Master UDI-DI on their label.

    Is a new Master UDI-DI needed when the brand or model name changes?

    The Master UDI-DI is assigned according to the design parameters listed in MDCG 2025-8: for frames, structure, material and lens size; for lenses, index, material, design and vision impairment group.

    How many Master UDI-DIs can be linked to one Basic UDI-DI?

    According to MDCG 2025-8, a Basic UDI-DI can be linked to more than one Master UDI-DI; a Master UDI-DI is linked to only one Basic UDI-DI.

    What power range is defined for ready-to-wear reading spectacles?

    MDCG 2025-8 defines discrete values between +1.00 D and +3.50 D for the lens spherical power as the assignment trigger.

    For the EUDAMED and ÜTS registrations of your spectacle products: Medical Device ÜTS Registration Consultancy

    Sources

    • European Commission: MDCG 2025-8 Rev.1 (March 2026)

    • EUR-Lex: Commission Delegated Regulation (EU) 2025/1920

    • EUR-Lex: Regulation (EU) 2017/745 (MDR)

 
 

Medical Device ÜTS Registration Consultancy

Request an ÜTS registration consultancy quote

Ehlibeyt Mahallesi, Tekstilciler Caddesi No: 35/7 06520 Çankaya / Ankara - TÜRKİYE

Cevizli Mahallesi, Zuhal Caddesi Ritim İstanbul B Blok, No:44 Maltepe / İstanbul - TÜRKİYE

info@medexmedikal.com
+90 505 580 4622
+90 312 389 1354

ÜTS Guides · Legislation Calendar · Glossary · ÜTS Guide Index · Frequently Asked Questions

Services: ÜTS Consultancy · Who Carries Out the ÜTS Registration? · ÜTS Document Registration Consultancy · ÜTS Product Registration Consultancy · Importer ÜTS Registration Consultancy · Sales Centre Authorisation Certificate · EUDAMED Registration Consultancy · MDR/IVDR Declaration of Conformity · Technical File Preliminary Review · Label and Instructions for Use Check · PRRC Service · PMS and PSUR · Clinical Evaluation · EMDN and GMDN Code Determination · Certificate of Free Sale · EUDAMED XML Generator · ÜTS Registration Training · Correction of a Rejected ÜTS Application

KVKK and Privacy Policy | Cookie Policy | Accessibility Statement

© 2026 Medex Kurumsal Danışmanlık Ltd. Şti.

bottom of page
WhatsApp Hemen Ara