How to Make an ÜTS Use Notification: Scope, Mandatory Fields, Cancellation and Consumption Notification

12 hours ago
7 min read
ÜTS kullanım bildirimi nasıl yapılır
  • SUMMARY

    An ÜTS use notification is the notification a healthcare service provider makes to the Product Tracking System (ÜTS), with information identifying the patient, when it uses a medical device that it has implanted into the body or that must be directly associated with the patient. The use of products for which no use notification is made is reported by a consumption notification.

    In this article we explain the two notifications through literal quotations (unofficial translations) from Articles 12 and 13 of the TİTCK Guideline on Single Movement Notifications of the Product Tracking System for Medical Devices. For all types of notifications, see our article on ÜTS single movement notifications.

    For your product registration and single notification transactions: Medical Device ÜTS Registration Consultancy

    ÜTS Login: How to Log In to ÜTS for Notifications

    Notifications are made in the Login area at utsuygulama.saglik.gov.tr/UTS. The description of the area on the ÜTS login page:

    “You can log in to the system with your user registered in ÜTS.”

    Two ways of logging in are offered:

    “Log in using your electronic signature”

    “Log in via the e-Devlet gateway”

    For login methods, user types and error messages: ÜTS login guide.

    Contents

    ÜTS Login1. What Is an ÜTS Use Notification and Who Makes It?2. Which Devices Require a Use Notification?3. Mandatory Fields in a Use Notification4. How Is Patient Information Entered?5. Cancelling an Erroneous Use Notification6. Quantity and Limited Use Rules7. Difference From a Consumption Notification8. Situations Where a Use Notification Cannot Be Made9. Frequently Asked Questions

    1. What Is an ÜTS Use Notification and Who Makes It?

    “Article 12- (1) When a medical device that is implanted into the body by the healthcare service provider and/or is within the scope of reimbursement and must be directly associated with the patient is used, a "Use Notification" is made for that product through patient information that identifies the person.”

    The healthcare service provider makes the notification. The Regulation on Medical Device Sales, Advertising and Promotion defines the healthcare service provider as follows:

    “k) (Amended: OG-26/5/2023-32202) Healthcare service provider: Natural persons and public and private law legal entities, and their branches without legal personality, that provide healthcare services, produce, and/or, provided that ownership remains with them, temporarily supply medical devices to patients,”

    For the ÜTS registration of a healthcare service provider, see our article on the healthcare service provider.

    2. Which Devices Require a Use Notification?

    Whether a use notification can be made is determined by the answers given in the device's ÜTS product registration. Under the second paragraph of Article 12, one of the following three situations must exist in the product registration:

    “The answer "yes" has been given to the question "Can it be implanted into the body?"”

    “The answer "yes" has been given to one of the questions "Is it single-use?" or "Does the product have a limited number of uses?" and the option "SUT match exists" has been selected for "SUT Match Status"”

    “The answer "yes" has been given to the question "Can it be used for a single patient?"”

    The meaning of the field "Can it be used for a single patient?" is explained in Article 27 of ÜTS Registration Guideline TCKKD-KLVZ-03:

    “2) This field is selected as "Yes" if the device in question can be used multiple times on a single patient (single patient multiple use), and as "No" if it is a device not intended to be used multiple times on a single patient.”

    These fields are entered by the company registering the product in ÜTS. For the fields of product registration, see our article on ÜTS product registration. For a device not defined in the product registration, the guideline opens a request route:

    “(10) If a request is made in ÜTS for medical devices for which the company or the healthcare service provider thinks that a use notification should be made, the Authority may define a use notification.”

    3. Mandatory Fields in a Use Notification

    Under the third paragraph of Article 12, after a query is made with the product number and the lot and/or serial number, the following fields are mandatory:

    • Quantity Used

    • Date of Use

    • Patient/Consumer Name-Surname Information

    • Identity Information Type

    “(4) The Turkish Republic Identity Number, Foreign Identity Number, Passport Number, YUPASS Number, and information on a newborn can be entered in the identity information type field.”

    4. How Is Patient Information Entered?

    The fifth paragraph of Article 12 lists the information to be entered according to the patient's status:

    “If the patient is a citizen of the Republic of Türkiye; the T.R. Identity Number, Patient Name and Patient Surname are entered.”

    “If the patient is a foreign national and has a foreign identity number; the Foreign Identity Number, Patient Name and Patient Surname are entered.”

    “If the patient is a foreign national and has no foreign identity number; the Passport Number, Patient Name and Patient Surname are entered.”

    “If there is a YUPASS number; the YUPASS number, Patient Name and Patient Surname are entered.”

    “If the patient is a newborn (and does not yet have a T.R. identity number); the Patient Name and Patient Surname are entered. By selecting "yes" for the "Is it a newborn?" question, the relevant identity number is entered in the "person identifier number" field through the mother's identity type.”

    “If the patient's identity is unknown, "unknown" is entered in the name-surname part, "identity information type" is selected as "identity unknown", and any number, such as the file number opened for that patient by the health facility, may be entered in the "person identifier number" field.”

    5. Cancelling an Erroneous Use Notification

    “(6) If a Use Notification was made erroneously, it is cancelled with a "Use Cancellation Notification". However, if the finalisation process was carried out in MEDULA after the use notification was made, the finalisation in MEDULA must first be cancelled, and then the use notification in ÜTS must be cancelled.”

    In other words, for a use finalised in MEDULA, the order is first MEDULA, then ÜTS.

    6. Quantity and Limited Use Rules

    “(7) For products invoiced to MEDULA with surface area information (cm2, etc.), the use notification in ÜTS is made on a per-piece basis.”

    “(8) For products whose "Number of Products in the Content" information entered in the product registration is more than "1", as many use notifications can be made as the number of products in the content.”

    “(9) For products with limited use, as many use notifications can be made as the "Limited Use Count" entered in the product registration.”

    For implantable and single-use devices, the limited use count is fixed by ÜTS (TCKKD-KLVZ-03, Article 27):

    “(The limited use count of devices that can be implanted into the body and/or are single-use is 1.)”

    7. Difference From a Consumption Notification

    “Article 13- (1) When products for which a use notification will not be made by the healthcare service provider are used, a "Consumption Notification" is made for these products.”

    Under the second paragraph of Article 13, for a consumption notification the answer "no" must have been given to the question "Can it be implanted into the body?" in the product registration; in addition, the answer "yes" must have been given to one of the questions "Is it single-use?" or "Does the product have a limited number of uses?" and "no SUT match exists" must have been selected in "SUT Match Status".

    “(3) When a Consumption Notification is made, after the query made with the product number and the lot and/or serial number;”

    “Quantity Consumed”

    “Date of Consumption”

    “it is mandatory to enter data in these fields.”

    The patient information field is not counted among the mandatory fields in a consumption notification. As with the use notification, a definition may be requested from the Authority for a consumption notification (Article 13, fourth paragraph).

    8. Situations Where a Use Notification Cannot Be Made

    The final provisions of the guideline close the use notification for some products:

    • Products past their expiry date: under Article 27, only recall giving, recall receiving and destruction/disposal notifications can be made.

    • Products subject to a recall decision: the list of permitted transactions in Article 26 does not include a use notification.

    • Market placement barcode: under Article 28, only production, import, export, giving and receiving notifications can be made through this barcode.

    The transfer of a product from a company to the stock of a healthcare service provider takes place through giving and receiving notifications; the stock transfer does not take place unless a receiving notification is made (Article 9). For this flow, see our article on ÜTS giving and receiving notification.

    9. Frequently Asked Questions

    Who makes an ÜTS use notification?

    The healthcare service provider makes it. The notification is made with patient information when a device that is implanted into the body, or that is within the scope of reimbursement and must be directly associated with the patient, is used.

    What is done if a use notification cannot be made for a product?

    A use notification depends on the answers in the product registration. For a device for which a use notification is thought to be necessary, the company or the healthcare service provider can open a request in ÜTS; the Authority may define a use notification.

    What is the difference between a use notification and a consumption notification?

    A use notification is made with patient information. A consumption notification is made when products for which no use notification will be made are used, and its mandatory fields are the quantity consumed and the date of consumption.

    Can a use notification finalised in MEDULA be cancelled?

    Yes, but the finalisation in MEDULA is cancelled first. Then a Use Cancellation Notification is made in ÜTS.

    How many use notifications are made if there are several products in a box?

    If the "Number of Products in the Content" in the product registration is more than 1, as many use notifications can be made as the number of products in the content.

    For product registration and single notifications suitable for use notifications: Medical Device ÜTS Registration Consultancy

    Sources

    • TİTCK: Guideline on Single Movement Notifications of the Product Tracking System for Medical Devices, Articles 9, 12, 13, 26, 27 and 28

    • TİTCK: ÜTS Registration Guideline TCKKD-KLVZ-03, Article 27

    • Legislation: Regulation on Medical Device Sales, Advertising and Promotion, Article 4

    • ÜTS application

 
 

Medical Device ÜTS Registration Consultancy

Request an ÜTS registration consultancy quote

Ehlibeyt Mahallesi, Tekstilciler Caddesi No: 35/7 06520 Çankaya / Ankara - TÜRKİYE

Cevizli Mahallesi, Zuhal Caddesi Ritim İstanbul B Blok, No:44 Maltepe / İstanbul - TÜRKİYE

info@medexmedikal.com
+90 505 580 4622
+90 312 389 1354

ÜTS Guides · Legislation Calendar · Glossary · ÜTS Guide Index · Frequently Asked Questions

Services: ÜTS Consultancy · Who Carries Out the ÜTS Registration? · ÜTS Document Registration Consultancy · ÜTS Product Registration Consultancy · Importer ÜTS Registration Consultancy · Sales Centre Authorisation Certificate · EUDAMED Registration Consultancy · MDR/IVDR Declaration of Conformity · Technical File Preliminary Review · Label and Instructions for Use Check · PRRC Service · PMS and PSUR · Clinical Evaluation · EMDN and GMDN Code Determination · Certificate of Free Sale · EUDAMED XML Generator · ÜTS Registration Training · Correction of a Rejected ÜTS Application

KVKK and Privacy Policy | Cookie Policy | Accessibility Statement

© 2026 Medex Kurumsal Danışmanlık Ltd. Şti.

bottom of page
WhatsApp Hemen Ara