Class IIa and IIb Medical Devices: Annex VIII Rules, Conformity Route and ÜTS Registration

SUMMARY
Class IIa and Class IIb are the two risk classes between Class I and Class III under the Medical Device Regulation. In both, a notified body takes part in the conformity assessment. The difference lies in the rules in Annex VIII that direct a device to these classes, in the scope in which the technical documentation is assessed, and in the ÜTS document business rules.
In this article we compare the two classes through literal quotations (unofficial translations) from the Medical Device Regulation and the TİTCK ÜTS Registration Guideline TCKKD-KLVZ-03. For the general map of all classes, see our MDR risk class article.
For the ÜTS document and product registration of your Class IIa and IIb devices: Medical Device ÜTS Registration Consultancy
Contents
1. How Are Class IIa and IIb Determined?2. Rules That Direct a Device to Class IIa3. Rules That Direct a Device to Class IIb4. Conformity Assessment Route5. CE Mark and EU Declaration of Conformity6. Document Business Rules in ÜTS7. Instructions for Use Upload Rule8. Class IIa and IIb Devices Under the MDD9. Frequently Asked Questions
1. How Are Class IIa and IIb Determined?
“(1) Devices are classified as class I, IIa, IIb or III, taking into account the risks they carry due to their intended purpose and structure. The classification is carried out in accordance with Annex VIII.”
The class derives not from the technical characteristic of the device but from its intended purpose. If more than one rule applies, the rule giving the highest class is taken as the basis:
“3.5. If several rules, or several sub-rules within the same rule, apply to the same device based on the intended purpose of the device, the strictest rule and sub-rule resulting in the higher classification apply.”
In the rules, the distinction between short-term and long-term use is decisive:
“1.2. Short term: Means intended for continuous use for between 60 minutes and 30 days under normal conditions.”
“1.3. Long term: Means intended for continuous use for more than 30 days under normal conditions.”
2. Rules That Direct a Device to Class IIa
The main provisions in Annex VIII that result in Class IIa:
“If intended for short-term use, they are classified as class IIa; however, those used in the oral cavity as far as the pharynx, in an ear canal up to the ear drum, or in a nasal cavity are classified as class I, and”
“All surgically invasive devices intended for transient use are classified as class IIa.”
“All surgically invasive devices intended for short-term use are classified as class IIa.”
“If intended to be placed in the teeth, they are classified as class IIa,”
“Software intended to provide information which is used to take decisions with diagnosis or therapeutic purposes is classified as class IIa.”
“Devices specifically intended for recording of diagnostic images generated by X-ray radiation are classified as class IIa.”
These are, respectively, the provisions of Rule 5 (invasive devices in relation to body orifices), Rules 6 and 7 (the basic rule for surgically invasive devices), Rule 8 (implantable devices placed in the teeth), Rule 11 (software) and Rule 17. Rules 2, 3, 4, 9, 10, 12, 16, 19, 20 and 21 also result in Class IIa in certain cases. Most of these rules move the device to a higher class with exceptions in the same sentence, so the whole rule must be read.
3. Rules That Direct a Device to Class IIb
The main provisions in Annex VIII that result in Class IIb:
“All implantable devices and long-term surgically invasive devices are classified as class IIb.”
“If intended for long-term use, they are classified as class IIb; however, those used in the oral cavity as far as the pharynx, in an ear canal up to the ear drum, or in a nasal cavity and not liable to be absorbed by the mucous membrane are classified as class IIa.”
“If they are intended to supply energy in the form of ionizing radiation, they are classified as class IIb,”
“All active devices intended for therapeutic purposes that emit ionizing radiation, including devices which control or monitor such devices or which directly influence their performance, are classified as class IIb.”
“All devices used for contraception or the prevention of the transmission of sexually transmitted diseases are classified as class IIb. However, if they are implantable or long-term invasive devices, they are classified as class III.”
“All devices intended specifically to be used for disinfecting, cleaning, rinsing or, where appropriate, hydrating contact lenses are classified as class IIb.”
The continuation of Rule 8 moves some of the implantable devices to Class III (such as contact with the heart or the central circulatory system, being totally absorbed, breast implants, joint replacements). Rules 2 (blood bags), 3, 4, 6, 7, 9, 10, 11, 12, 16, 19, 20 and 21 also result in Class IIb in certain cases. For Class III, see our article on Class III medical devices and ÜTS registration documents.
4. Conformity Assessment Route
Article 52 of the Medical Device Regulation, for Class IIa:
“(5) Manufacturers of class IIa devices, other than custom-made or investigational devices:”
“a) are subject to a conformity assessment as specified in Chapters I and III of Annex IX, including an assessment of the technical documentation as specified in item 4 of Annex IX of at least one representative device for each device category,”
“b) Alternatively, the manufacturer may choose to draw up the technical documentation specified in Annexes II and III accompanied by the conformity assessment specified in item 10 or item 18 of Annex XI. The assessment of the technical documentation is applied for at least one representative device from each device category.”
For Class IIb:
“a) Manufacturers of class IIb devices, other than custom-made or investigational devices, are subject to a conformity assessment as specified in Chapters I and III of Annex IX, including an assessment of the technical documentation as specified in item 4 of Annex IX of at least one representative device for each generic device group.”
“b) However, in the case of class IIb implantable devices, other than sutures, staples, dental fillings, dental braces, tooth crowns, screws, wedges, plates, wires, pins, clips and connectors, the assessment of the technical documentation as specified in item 4 of Annex IX is applied for every device.”
“c) Alternatively, the manufacturer may choose to apply a conformity assessment based on type-examination specified in Annex X accompanied by a conformity assessment based on product conformity verification specified in Annex XI.”
The difference is this: in Class IIa, the technical documentation is assessed for at least one representative device from each device category; in Class IIb, for at least one representative device from each generic device group. In Class IIb implantable devices other than the listed devices such as sutures, screws and plates (called "WET devices" in the Guideline), the assessment is made for every device. For the comparison of the procedures, see our article on the selection of the conformity assessment procedure.
5. CE Mark and EU Declaration of Conformity
Since a notified body takes part in the conformity assessment in both classes, the fifth paragraph of Article 20 of the Medical Device Regulation applies:
“(5) Where there is a notified body responsible for the conformity assessment procedures referred to in Article 52, the CE mark is followed by the identification number of the notified body. The identification number is also indicated in any promotional material which mentions that a device fulfils the requirements for CE marking.”
Notified body information is also sought in the EU declaration of conformity. Article 11 of the ÜTS Registration Guideline:
“ı) For EU declarations of conformity other than those whose scope covers only Class-I Other or Class A devices; the name and identification number of the notified body, a description of the conformity assessment procedure carried out and the identification of the EU certificate or EU certificates issued (certificate number or numbers),”
For the whole content of the declaration, see our article on the EU declaration of conformity and the CE mark.
6. Document Business Rules in ÜTS
In the device registration in ÜTS, "Class-IIa" or "Class-IIb" is selected. The documents to be associated with the device are shown in different annexes according to the class and the type of device (Guideline Article 66):
“(4) For devices to be registered in ÜTS as "Class-IIa" under the MDR, the appropriate one of the ÜTS document business rules shown in Annex-89 is carried out.”
“(5) For devices to be registered in ÜTS as "Class-IIb" under the MDR, other than devices implantable in the body or devices within the scope of Rule-12 of MDR Annex VIII, the appropriate one of the ÜTS document business rules shown in Annex-90 is carried out.”
“(6) For devices implantable in the body, other than sutures, staples, dental fillings, dental braces, tooth crowns, screws, wedges, plates, wires, pins, clips and connectors, to be registered in ÜTS as "Class-IIb" under the MDR, the appropriate one of the ÜTS document business rules shown in Annex-91 is carried out.”
“(7) For devices within the scope of Rule-12 of MDR Annex VIII to be registered in ÜTS as "Class-IIb" under the MDR, the appropriate one of the ÜTS document business rules shown in Annex-92 is carried out.”
That is, Class IIb devices are divided in ÜTS into three separate document business rules: general IIb, implantable IIb other than WET, and IIb within the scope of Rule 12. For the steps of document registration, see our ÜTS document registration article; for the product registration fields, see our ÜTS product registration articles.
7. Instructions for Use Upload Rule
According to Article 17 of the ÜTS Registration Guideline, in devices certified under the MDR and MDD, uploading to ÜTS is mandatory for:
“1) For Class-III and Class-IIb devices, the instructions for use drawn up by the manufacturer,”
“2) For Class-IIa devices, the instructions for use if the instructions for use have been drawn up by the manufacturer; as an exception, only for Class-IIa devices under the MDD, if the instructions for use have not been drawn up by the manufacturer, the intended use form/declaration mentioned in Article 19 of this Guideline, signed by the responsible manager of the manufacturer or importer company,”
In Class IIa devices that are accessories of another device, if the "Is it a component/accessory of a device?" field is selected as "Yes" in the registration, there is no obligation to upload instructions for use (Article 17, fourth paragraph).
8. Class IIa and IIb Devices Under the MDD
Article 45 of the ÜTS Registration Guideline writes the following dates, provided that the conditions in Regulation (EU) 2023/607 are met:
“(4) Provided that the conditions specified in the Regulation mentioned in the first paragraph of this article are fulfilled, Class-III devices under the MDD or AIMDD and Class-IIb implantable devices other than sutures, staples, dental fillings, dental braces, tooth crowns, screws, wedges, plates, wires, pins, clips and connectors may be placed on the market or put into service until 31.12.2027.”
“(5) Provided that the conditions specified in the Regulation mentioned in the first paragraph of this article are fulfilled, Class-IIb devices under the MDD other than those covered by the fourth paragraph of this article, as well as Class-IIa devices, Class-Im devices and Class-Is devices, may be placed on the market or put into service until 31.12.2028.”
The date 31.12.2027 is for Class IIb implantable devices other than the listed exceptions only; devices such as sutures, screws and plates are in the 31.12.2028 group. For these devices, the application is made first to EBS and then to ÜTS. For details, see our article on the (EU) 2023/607 MDR transition timetable.
9. Frequently Asked Questions
What is the main difference between Class IIa and IIb?
The class derives from the intended purpose according to the Annex VIII rules. In the conformity assessment, the technical documentation is assessed for at least one representative device from each device category in Class IIa and from each generic device group in Class IIb; in IIb implants other than WET, it is assessed for every device.
Is the same document rule applied in ÜTS for Class IIa and IIb?
No. For Class IIa, the document business rules in Annex-89 are applied; for Class IIb, those in Annex-90, Annex-91 or Annex-92, according to the type of device.
Is it mandatory to upload instructions for use to ÜTS for a Class IIa device?
It is uploaded if the manufacturer has drawn up instructions for use. In Class IIb devices, uploading the manufacturer's instructions for use is mandatory.
What is the MDD transition date for Class IIb implant devices?
For Class IIb implantable devices other than the listed devices such as sutures, staples, dental fillings, screws and plates, it is 31.12.2027. For the listed devices and other Class IIb devices, it is 31.12.2028; in both cases the conditions in the Regulation must be met.
Does the CE mark of a Class IIa device bear the notified body number?
Yes. If there is a notified body responsible for the conformity assessment, the CE mark is followed by the identification number of that body. This applies to Class IIa and IIb.
For the ÜTS document and product registration of your Class IIa and IIb devices: Medical Device ÜTS Registration Consultancy
Sources
Legislation: Medical Device Regulation (OG 02.06.2021, 31499 bis), Articles 20, 51 and 52, Annex VIII
TİTCK: ÜTS Registration Guideline TCKKD-KLVZ-03, Articles 11, 17, 45 and 66