Class I Medical Device: Class I Other, Is, Im and Ir Differences and ÜTS Registration

SUMMARY
A Class I medical device is a device that falls into the lowest risk class under the rules of Annex VIII of the Medical Device Regulation. In this class the manufacturer prepares the technical documentation, draws up the EU declaration of conformity and declares conformity itself. If the device is sterile, has a measuring function or is a reusable surgical instrument, the notified body assesses only the relevant aspect.
In this article we give, through literal quotations (unofficial translations) from the Medical Device Regulation and TİTCK ÜTS Registration Guideline TCKKD-KLVZ-03, the basis of Class I, the difference between Is, Im and Ir, the rules on the CE marking and the declaration of conformity, and the choice of class in ÜTS. For the general map of all risk classes, see our article on the MDR risk class.
For the declaration of conformity and ÜTS registration of your Class I devices: Medical Device ÜTS Registration Consultancy
Contents
1. What Is a Class I Medical Device?2. Rules That Direct a Device to Class I3. Difference Between Class I Other, Is, Im and Ir4. Conformity Assessment: Role of the Notified Body5. CE Marking and Notified Body Number6. Class I in the EU Declaration of Conformity7. Registering a Class I Device in ÜTS8. Class I Devices Under the MDD and Transition9. Frequently Asked Questions
1. What Is a Class I Medical Device?
“(1) Devices are classified as class I, IIa, IIb or III, taking into account the risks inherent in the intended purpose and nature of the devices. Classification is carried out in accordance with Annex VIII.”
This provision is in Article 51 of the Medical Device Regulation. The class is found according to the intended purpose determined by the manufacturer. The application rule in Annex VIII:
“3.1. The application of the classification rules shall be governed by the intended purpose of the devices.”
“3.5. If several rules, or several sub-rules within the same rule, apply to the same device based on its intended purpose, the strictest rule and sub-rule resulting in the higher classification shall apply.”
In other words, if a device is Class I under one rule and falls in a higher class under another rule, the higher class applies.
2. Rules That Direct a Device to Class I
The provisions in Annex VIII that result in Class I are as follows:
“All non-invasive devices are classified as class I, unless one of the rules set out below applies.”
“In all other cases, such devices are classified as class I.”
“If they are intended to be used as a mechanical barrier, for compression or for absorption of exudates, they are classified as class I,”
“If they are intended for transient use, they are classified as class I,”
“If they are reusable surgical instruments, they are classified as class I,”
“All other software is classified as class I.”
“All other active devices are classified as class I.”
In order: Rule 1 (non-invasive devices), Rule 2 (other cases for devices for channelling or storing blood, body liquids or gases), Rule 4 (mechanical barriers in contact with injured skin), Rule 5 (invasive devices with respect to body orifices, for transient use), Rule 6 (reusable surgical instruments), Rule 11 (other software) and Rule 13 (other active devices). Rule 5 also classifies as Class I devices used in short-term use in the oral cavity (as far as the pharynx), in the ear canal (up to the ear drum) and in the nasal cavity. Rule 10 places in Class I diagnostic devices intended to illuminate the patient's body in the visible spectrum.
The durations in the rules are defined in Chapter I of Annex VIII:
“1.1. Transient: Normally intended for continuous use for less than 60 minutes.”
“1.2. Short term: Normally intended for continuous use for between 60 minutes and 30 days.”
For software, the details are in our article on medical device software and Rule 11.
3. Difference Between Class I Other, Is, Im and Ir
The MDR defines Class I as a single class. In ÜTS, however, there are sub-options according to the characteristics of the device. Article 56 of the Guideline:
“(Additional classifications have been made in ÜTS according to characteristic product features, such as whether the device is packaged sterile or has a measuring function. For example, both a device that is "Class-Ir" in ÜTS and a device that is "Class-I Other" in ÜTS are classified as Class-I under the MDR.)”
Class I Other: a Class I device that is non-sterile, has no measuring function and is not a reusable surgical instrument.
Class Is: a Class I device placed on the market in a sterile condition.
Class Im: a Class I device with a measuring function.
Class Ir: a reusable surgical instrument.
The definition of a reusable surgical instrument is in Annex VIII:
“2.3. Reusable surgical instrument: An instrument intended for surgical use in cutting, drilling, sawing, scratching, scraping, clamping, retracting, clipping or similar procedures, without a connection to an active device, and which is intended by the manufacturer to be reused after appropriate procedures such as cleaning, disinfection and sterilisation have been carried out.”
4. Conformity Assessment: Role of the Notified Body
Sixth paragraph of Article 52 of the Medical Device Regulation:
“(6) Manufacturers of class I devices, other than custom-made or investigational devices, shall declare the conformity of their products by drawing up the EU declaration of conformity referred to in Article 19, after preparing the technical documentation set out in Annex II and Annex III. If these devices are placed on the market in a sterile condition, have a measuring function, or are reusable surgical instruments, the manufacturer shall apply the procedures set out in Chapters I and III of Annex IX or in Part A of Annex XI. However, the involvement of the notified body in those procedures shall be limited:”
“a) in the case of devices placed on the market in a sterile condition, to the aspects relating to establishing, securing and maintaining sterile conditions,”
“b) in the case of devices with a measuring function, to the aspects relating to the conformity of the devices with the metrological requirements,”
“c) in the case of reusable surgical instruments, to the aspects relating to the reuse of the device, in particular cleaning, disinfection, sterilisation, maintenance and functional testing, and the related instructions for use,”
“to those aspects.”
Result: in Class I Other there is no notified body. In Is, Im and Ir there is a notified body, but it assesses only the sterility, metrology or reuse aspect. For a comparison of conformity assessment routes, see our article on choosing the conformity assessment procedure.
5. CE Marking and Notified Body Number
“(5) Where a notified body is responsible for the conformity assessment procedures referred to in Article 52, the CE marking shall be followed by the identification number of the notified body. The identification number shall also be indicated in any promotional material which mentions that a device fulfils the requirements for CE marking.”
This provision is in Article 20 of the Medical Device Regulation. On Is, Im and Ir devices, the four-digit number of the notified body is placed next to the CE marking. Article 17 of the ÜTS Registration Guideline writes the following exception for the CE marking in the instructions for use:
“For Class-I Other, Class A and IVD Other products, only the CE marking is sufficient.”
6. Class I in the EU Declaration of Conformity
In Class I, conformity is declared by the manufacturer's EU declaration of conformity. The minimum content of the declaration is in Annex IV of the Medical Device Regulation; the fields checked in ÜTS are listed in Article 11 of the Guideline. For notified body information, the Guideline makes the following distinction:
“ı) For EU declarations of conformity other than those covering only Class-I Other or Class A devices: the name and identification number of the notified body, a description of the conformity assessment procedure performed, and the identification of the EU certificate or EU certificates issued (certificate number or numbers),”
Accordingly, a declaration covering an Is, Im or Ir device contains the notified body name, number and certificate number. For the content of the declaration, see our article on the manufacturer's EU declaration of conformity.
7. Registering a Class I Device in ÜTS
In a device registration under the MDR, the class is selected from the mandatory "Class" field under the "Classification Information" heading (Guideline Article 27, fourth paragraph). The options for Class I are "Class-I Other (not sterile and without a measuring function)", "Class-Ir (Reusable surgical instruments)", "Class-Im (With a measuring function)" and "Class-Is (With a sterile function)".
The documents to be associated with the device are determined under Article 66 of the Guideline:
“(3) For devices to be registered in ÜTS under the MDR as "Class-I Other", "Class-Is", "Class-Im" or "Class-Ir", the appropriate one of the ÜTS document business rules shown in Annex-88 is carried out.”
Rule on uploading instructions for use (Guideline Article 17, third paragraph):
“(3) For other risk groups of device groups (for Class-I Other, Class-Im, Class-Is, Class-Ir, Class A, Class A Sterile and IVD Other devices), uploading the instructions for use is optional.”
Under the same paragraph, the Authority may request that the instructions for use of these devices be uploaded to ÜTS where it deems necessary. For all registration fields, see our article on ÜTS product registration; for registering documents, see our article on ÜTS document registration.
8. Class I Devices Under the MDD and Transition
Article 45 of the ÜTS Registration Guideline gives the following dates for the Class I sub-groups, provided that the conditions in Regulation (EU) 2023/607 are met:
“(5) Provided that the conditions specified in the Regulation referred to in the first paragraph of this Article are fulfilled, Class-IIb devices, other than those covered in the fourth paragraph of this Article, and Class-IIa devices, Class-Im devices and Class-Is devices under the MDD may be placed on the market or put into service until 31.12.2028.”
“(6) Provided that the conditions specified in the Regulation referred to in the first paragraph of this Article are fulfilled, devices for which the conformity assessment procedure under the MDD did not require the involvement of a notified body, for which a declaration of conformity was drawn up before 26 May 2021, and for which the conformity assessment procedure under the MDR requires the involvement of a notified body (devices that were Class-I Other under the MDD but whose class is upgraded under the MDR) may be placed on the market or put into service until 31.12.2028.”
For these devices, the application is made first through EBS and then in ÜTS (Guideline Article 45, seventh paragraph). For the process, see our article on the MDR extension application.
For the registration steps of Class I devices, see our ÜTS transactions guide; for the code requested at registration, our article on the GMDN code; and for having the registrations carried out on your behalf, our ÜTS consultancy page.
9. Frequently Asked Questions
Is a notified body required for a Class I device?
It is not required for Class I Other; the manufacturer prepares the technical documentation and draws up the EU declaration of conformity. In Is, Im and Ir, the notified body assesses only the sterility, metrology or reuse aspects.
Is there a number in the CE marking of Is, Im and Ir devices?
Yes. Where a notified body is responsible for the conformity assessment, the CE marking is followed by the identification number of the notified body. For Class I Other, the CE marking alone is sufficient.
Which class is a reusable surgical instrument?
Under Rule 6 it is Class I and is selected as "Class-Ir" in ÜTS. Instruments that are not connected to an active device, are used in surgical procedures such as cutting, drilling and scraping, and are reused after cleaning and sterilisation fall within this definition.
Is it mandatory to upload the instructions for use to ÜTS for a Class I device?
No, it is optional. However, the Authority may request that the instructions for use be uploaded where it deems necessary.
Does the declaration of conformity of a Class I device contain notified body information?
It is not required only in a declaration covering Class I Other devices. In a declaration covering an Is, Im or Ir device, the name and number of the notified body, the description of the procedure and the certificate number are included.
For the declaration of conformity and ÜTS registration of your Class I devices: Medical Device ÜTS Registration Consultancy
Sources
Legislation: Medical Device Regulation (OG 02.06.2021, 31499 bis), Articles 20, 51 and 52, Annex VIII
TİTCK: ÜTS Registration Guideline TCKKD-KLVZ-03, Articles 11, 17, 27, 45, 56 and 66