Authorized Representative: Medical Device Regulation Article 11, Mandate, Responsibility and ÜTS Registration

SUMMARY
The authorized representative is the natural or legal person established in Türkiye or the EU whom a manufacturer established outside Türkiye or the EU member states must appoint in order to place its devices on the market. The appointment is made by a mandate.
In this article we give, through literal quotations (unofficial translations) from TCKKD-KLVZ-03 and the Medical Device Regulation, Articles 11 and 12 and the way the mandate is registered in ÜTS. For the details of the service, see our TR/EU authorized representation service page.
For the ÜTS registration of your foreign manufacturer's products: Medical Device ÜTS Registration Consultancy
Contents
1. What Is an Authorized Representative, When Is It Mandatory?2. When Does the Appointment Become Valid?3. Minimum Tasks in the Mandate4. Non-Transferable Obligations and Joint and Several Liability5. Termination of the Mandate6. Change of Authorized Representative7. Verification Duty of the Importer and the Expertise Requirement8. Registration of the Mandate in ÜTS9. Frequently Asked Questions
1. What Is an Authorized Representative, When Is It Mandatory?
The definition in Article 3 of the Medical Device Regulation:
“lll) Authorized representative: The natural or legal person established in Türkiye or the EU member states who has received and accepted a written mandate from a manufacturer, established outside Türkiye or the EU member states, to act on the manufacturer's behalf in relation to specified tasks with regard to the latter's obligations under this Regulation”
“ARTICLE 11 -(1) Where the manufacturer of a device is not established in Türkiye or the EU member states, the device may be placed on the market only if the manufacturer designates an authorized representative.”
A manufacturer established in Türkiye or the EU does not appoint an authorized representative. TCKKD-KLVZ-03:
“d) Manufacturers established in an EU member state or in Türkiye cannot appoint an EU authorized representative. The EU authorized representative requirement exists only for manufacturers not established in an EU member state or in Türkiye.”
2. When Does the Appointment Become Valid?
“(2) The designation of the authorized representative becomes valid when a mandate is issued and accepted in writing by the authorized representative, and covers, as a minimum, all devices of the same generic device group.”
There are two conditions: the issuance of the mandate and the written acceptance of the representative. The scope is at least all devices in the same generic device group.
3. Minimum Tasks in the Mandate
“(3) The authorized representative performs the tasks specified in the mandate agreed between the authorized representative and the manufacturer. The authorized representative provides a copy of the mandate to the Agency upon request. The mandate requires the authorized representative to perform, as a minimum, the following tasks in relation to the devices it covers, and the manufacturer enables the authorized representative to perform these tasks:”
“a) Verify that the EU declaration of conformity and the technical documentation have been drawn up by the manufacturer and, where applicable, that an appropriate conformity assessment procedure has been carried out.”
“b) Keep a copy of the technical documentation, the EU declaration of conformity and, if applicable, a copy of the relevant certificate issued in accordance with Article 56, including any amendments and supplements, available for the Agency for the period referred to in the eighth paragraph of Article 10.”
“c) Comply with the registration obligations laid down in Article 31 and verify that the manufacturer has complied with the registration obligations laid down in Articles 27 and 29.”
“ç) In response to a request from the Agency, provide the Agency with all the information and documentation necessary to demonstrate the conformity of a device, in Turkish and/or in an official EU language determined by the Agency.”
“d) Forward to the manufacturer any request by the Agency for samples or access to a device and verify that the Agency receives the samples or is given access to the device.”
“e) Cooperate with the competent authorities on any preventive or corrective action taken to eliminate or, if that is not possible, mitigate the risks posed by devices.”
“f) Inform the manufacturer immediately about complaints and reports from healthcare professionals, patients and users about suspected incidents related to a device for which they have been designated.”
“g) Terminate the mandate if the manufacturer acts contrary to its obligations under this Regulation.”
Article 31 in subparagraph (c) is the registration of the manufacturer, authorized representative and importer in the electronic system (EUDAMED); after registration, the single registration number (SRN) is provided (Article 31, second paragraph). For the EUDAMED actor registration, see our EUDAMED registration article.
4. Non-Transferable Obligations and Joint and Several Liability
“(4) The manufacturer cannot delegate the obligations laid down in the first, second, third, fourth, sixth, seventh, ninth, tenth, eleventh and twelfth paragraphs of Article 10 within the scope of the mandate referred to in the third paragraph.”
“(5) Without prejudice to the fourth paragraph, where the manufacturer does not comply with the obligations laid down in Article 10, the authorized representative is legally liable for defective devices on the same basis as, and jointly and severally with, the manufacturer.”
The general obligations of the manufacturer are in Article 10; for the details, see our article on the general obligations of manufacturers.
5. Termination of the Mandate
“(6) An authorized representative who terminates the mandate on the grounds referred to in subparagraph (g) of the third paragraph immediately informs the Agency and, where applicable, the notified body that was involved in the conformity assessment of the device, of the termination of the mandate and the reasons for it.”
6. Change of Authorized Representative
“(1) The detailed arrangements for a change of authorized representative are clearly defined in an agreement between the manufacturer, the incoming authorized representative and, where practicable, the outgoing authorized representative. This agreement addresses at least the following points:”
“a) The date of termination of the mandate of the outgoing authorized representative and the date of commencement of the mandate of the incoming authorized representative.”
“b) The date until which the outgoing authorized representative may be indicated in the information supplied by the manufacturer, including any promotional material.”
“c) The transfer of documents, including confidentiality aspects and property rights.”
“ç) The obligation of the outgoing authorized representative, after the end of the mandate, to forward to the manufacturer or the incoming authorized representative any complaints or reports from healthcare professionals, patients or users about suspected incidents related to a device for which it had been designated as authorized representative.”
7. Verification Duty of the Importer and the Expertise Requirement
According to the second paragraph of Article 13, in order to place a device on the market, the importer also verifies the following:
“2) That the manufacturer has been identified and that an authorized representative has been designated by the manufacturer in accordance with Article 11,”
The person responsible for regulatory compliance requirement for the authorized representative is in Article 15:
“(6) Authorized representatives have permanently and continuously at their disposal at least one person responsible for regulatory compliance who possesses the requisite expertise in the field of regulatory requirements for medical devices.”
For the qualifications of this person, see our article on the person responsible for regulatory compliance, and for the service, our regulatory compliance responsible service page.
8. Registration of the Mandate in ÜTS
In ÜTS, the mandate is not a separate document type; it is uploaded when the foreign manufacturer's EU declaration of conformity is registered. The seventh paragraph of Article 12 of TCKKD-KLVZ-03:
“a) In the document registration notifications of the "EU Declaration of Conformity (MDR/IVDR)" type belonging to manufacturers not established in an EU member state or in Türkiye and mentioned in this article, it is mandatory to make an entry in the "EU Authorized Representative" field on the ÜTS registration screen.”
“c) The original of the said document and its Turkish translation by a sworn translator are uploaded together in the "Mandate" field. No apostille or embassy/consulate approval is required on the original of the mandate.”
Instead of the mandate, a printout of the manufacturer actor screen in EUDAMED in which the authorized representative information is clearly visible can also be uploaded (subparagraph (b) of the same paragraph). The name or trade name and address of the EU authorized representative on the EU certificate and the EU declaration of conformity are also among the fields checked in ÜTS (Articles 7 and 11, second paragraph, subparagraph (b)). For the translation, see our sworn translation and document preparation service page, and for the document registration, our ÜTS document registration article.
It also appears on the label. Medical Device Regulation Annex I, 23.2:
“ç) Where applicable, the name and registered place of business of the authorized representative,”
9. Frequently Asked Questions
Does a Turkish manufacturer appoint an authorized representative?
No. According to TCKKD-KLVZ-03, manufacturers established in an EU member state or in Türkiye cannot appoint an EU authorized representative; the requirement applies only to manufacturers established outside these countries.
When does the appointment of an authorized representative become valid?
When the mandate is issued and accepted in writing by the authorized representative. The appointment covers, as a minimum, all devices in the same generic device group.
Does the authorized representative bear the manufacturer's responsibility?
The manufacturer cannot delegate its core obligations listed in Article 10. However, if the manufacturer does not comply with Article 10, the authorized representative is jointly and severally liable with the manufacturer for defective devices on the same basis.
Is an apostille required for the mandate uploaded to ÜTS?
No. According to Article 12 of TCKKD-KLVZ-03, no apostille or embassy/consulate approval is required on the original of the mandate; the original and its Turkish translation by a sworn translator are uploaded together.
What is done when the authorized representative changes?
According to Article 12, an agreement is made between the manufacturer, the incoming and, where practicable, the outgoing representative. The agreement covers the termination and commencement dates, the last date the outgoing representative may be indicated in the information, the transfer of documents and the obligation to forward complaints.
For your mandate, declaration of conformity and product registration: Medical Device ÜTS Registration Consultancy
Sources
Mevzuat.gov.tr: Medical Device Regulation, Articles 3, 11, 12, 13, 15 and 31, Annex I
TİTCK: ÜTS Registration Guideline TCKKD-KLVZ-03, Articles 7, 11 and 12
EUR-Lex: Regulation (EU) 2017/745 (MDR), Article 11