Exporting Medical Devices to Finland: ÜTS, EUDAMED, CE and Label Language
Finland's language rule contains a distinction based on two official languages and a risk assessment: information necessary for safe use must be in Finnish and Swedish, and the remaining information in Finnish, Swedish or English. For a manufacturer established in Türkiye that will export medical devices to Finland, this page brings together the relevant sections of Medical Devices Act No 719/2021, Fimea's registration page, MDCG 2022-16 and the ÜTS rules.
Valtuutettu edustaja: Not required for a Turkish manufacturer
In Finnish documents the authorised representative is called valtuutettu edustaja. Fimea's registration page counts the authorised representative among the operators that notify Fimea; since a manufacturer established in Türkiye has no authorised representative, this item does not apply.
"'Union market' means the territories of the European Union Member States, extended to Norway, Liechtenstein and Iceland through the European Economic Area (EEA), and to Türkiye through the Customs Union Agreement."
Source: Turkish translation, MDCG 2022-16 Guidance on Authorised Representatives, footnote 1.
MDR Article 11(1):
"Where the manufacturer of a device is not established in a Member State, the device may only be placed on the Union market if the manufacturer designates a sole authorised representative."
Source: MDR, Article 11, first paragraph.
The definition in the Medical Device Regulation:
"Authorised representative: a natural or legal person established in Türkiye or an EU Member State who has received and accepted a written mandate from a manufacturer established outside Türkiye or the EU Member States to act on the manufacturer's behalf in relation to specified tasks with regard to the latter's obligations under this Regulation"
Source: Medical Device Regulation (OG 31499), Article 4, first paragraph, point (lll).
The EUDAMED manufacturer registration and the Customs Union basis are shown to the Finnish buyer. Authorised representation is for manufacturers outside Türkiye and the EU.
Label and instructions for use language: Finnish and Swedish, risk-based distinction
"Manufacturers shall ensure that the device is accompanied by the information set out in Section 23 of Annex I in an official Union language(s) determined by the Member State in which the device is made available to the user or patient. The particulars on the label shall be indelible, easily legible and clearly comprehensible to the intended user or patient."
Source: MDR, Article 10(11) (manufacturer obligations).
In the unofficial English translation on Finlex of Finland's Medical Devices Act No 719/2021 (Laki lääkinnällisistä laitteista, 15 July 2021), § 5 sets this rule: the information and documents under MDR Article 10(11) are in Finnish, Swedish or English, unless given in internationally recognised symbols; information necessary for safe use must be available in Finnish and Swedish, and the manufacturer determines which information is necessary for safe use on the basis of a risk assessment; if the device is intended for patient or consumer use, the instructions for use and safe-use information are in Finnish and Swedish; the instructions for use and label of single-use devices may be in Finnish, Swedish or both, according to user need. The implant card (MDR Article 18(1)) is in Finnish, Swedish and English; the declaration of conformity and certificates may be drawn up in Finnish, Swedish or English. Fimea may require certain information to be provided in Finnish or Swedish where safety requires.
Planning for a manufacturer in Türkiye: English IFU is possible for a device for professional use, but the warnings that the risk assessment deems necessary for safe use are added in Finnish and Swedish; for a consumer device, a full Finnish and Swedish IFU is prepared. The Swedish content prepared for Sweden is reused here. For the basis of the Turkish requirement, see the article on Turkish labels and instructions for use.
Fimea, CERE registration and distributor notification
Finland's competent authority is the Lääkealan turvallisuus- ja kehittämiskeskus Fimea (Finnish Medicines Agency, Helsinki); it appears in the Commission's May 2026 national authority list for the MDR and IVDR. According to Fimea's page on device and operator registers, the registration obligation is based on § 49 of Act No 719/2021; an operator obliged to notify reports its activity and devices to the national CERE register kept by Fimea or to EUDAMED; importers link their device to the manufacturer they import it from.
According to the Finlex translation of § 49 of the Act, a manufacturer established in Finland, a custom-made device manufacturer, a system or procedure pack producer, a sterilisation service provider, an authorised representative and an importer notify Fimea of their activity and device before placing the device on the market or making it available. The third paragraph of the same section also provides for notification by an operator that distributes devices to retailers, health and social care organisations and other professional users and makes them available in Finland; this notification includes the operator's name, place of business and the list of devices to be made available in Finland. According to Fimea's announcement of 2 June 2023, these distributors are not required to submit a device notification until a further announcement; it is sufficient for them to enter their operator information in Fimea's electronic service.
A manufacturer in Türkiye is not established in Finland and does not fall under the first paragraph of § 49; its registration is made in EUDAMED. A distributor in Finland makes its own notification under the third paragraph.
"Member States may maintain or introduce national provisions on registration of distributors of devices which have been made available on their territory."
Source: Regulation (EU) 2017/745 (MDR), Article 30(2).
EUDAMED registration for Finland
Fimea's page defines registration in two stages: first operator registration (Actor ID/SRN in EUDAMED or CERE reference number), then device registration in the same system. For a manufacturer established in Türkiye, the system is EUDAMED.
"Before placing a device, other than a custom-made device, on the market, manufacturers, authorised representatives and importers shall, in order to register, submit to the electronic system referred to in Article 30 the information referred to in Section 1 of Part A of Annex VI, provided that they have not already registered in accordance with this Article."
Source: Regulation (EU) 2017/745 (MDR), Article 31(1).
"Before placing a device, other than a custom-made device, on the market, the manufacturer shall, in accordance with the rules of the issuing entity referred to in Article 27(2), assign a Basic UDI-DI as defined in Part C of Annex VI to the device and shall provide it to the UDI database together with the other core data elements referred to in Part B of Annex VI related to that device."
Source: Regulation (EU) 2017/745 MDR, Article 29, first paragraph.
The provision from which the mandatory use date is calculated:
"the obligations and requirements that relate to any of the electronic systems referred to in Article 33(2) shall apply from the date corresponding to 6 months from the date of publication of the notice referred to in Article 34(3), informing that the relevant electronic system is functional and meets the functional specifications drawn up pursuant to Article 34(1)."
Source: MDR Article 123(3)(d), introductory sentence, as amended by Article 1(3)(a) of Regulation (EU) 2024/1860.
According to the Commission's EUDAMED page, the publication date is 27 November 2025 (Decision (EU) 2025/2371); the four modules have been mandatory since 28 May 2026. The steps are in the articles on EUDAMED actor registration, EUDAMED device registration and legacy device registration; for scope, see EUDAMED registration consultancy.
ÜTS export notification and free sale certificate for Finland
Shipments to Finland go by sea and via the Baltic route; the ÜTS notification is made with the export declaration in Türkiye.
"An 'Export Notification' is made for individual products exported by the company."
Source: TİTCK Guide on Individual Movement Notifications, Article 17(1).
"When making an Export Notification in ÜTS, after querying by product number, lot and/or serial number, it is mandatory to enter data in the Customs Declaration Number and Quantity fields."
Source: TİTCK Guide on Individual Movement Notifications, Article 17(2).
"If an Export Notification was made incorrectly, it is cancelled with an 'Export Cancellation Notification'."
Source: TİTCK Guide on Individual Movement Notifications, Article 17, third paragraph.
The details of the notification are in the articles on ÜTS individual movement notifications and on ÜTS records for exports, free zones and bonded warehouses; and on the individual notification procedures page for the monthly service.
A free sale certificate is not a condition for Finland; if the buyer requests it, it is requested through ÜTS.
"For export purposes and upon the request of a manufacturer or an authorised representative, the Agency issues a free sale certificate stating that the manufacturer's or, where applicable, the authorised representative's place of business is registered in Türkiye and that the device in question, bearing the CE marking in accordance with this Regulation, may be marketed in Türkiye and in EU Member States."
Source: Medical Device Regulation (OG 31499), Article 60, first paragraph.
The application route is explained in the article on the free sale certificate (FSC) and ÜTS. Nordic markets: Sweden and Denmark.
Finland summary table
TopicStatusCompetent authorityFimea (Lääkealan turvallisuus- ja kehittämiskeskus, Finnish Medicines Agency), HelsinkiLabel / IFU languageFinnish, Swedish or English; information necessary for safe use in Finnish and Swedish; IFU for consumer devices in Finnish and Swedish (719/2021 § 5)National additional registration / notificationManufacturers, ARs, importers and distributors established in Finland notify Fimea (CERE or EUDAMED) (§ 49); a manufacturer in Türkiye registers in EUDAMEDIs an AR requiredNo (manufacturer established in Türkiye, Customs Union; MDCG 2022-16 footnote 1)
Medical Device ÜTS Registration Consultancy for exports to Finland
We handle the ÜTS product registration, export notifications and EUDAMED actor and device registration of a manufacturer exporting to Finland; we check that the Finnish and Swedish safe-use information on the label and IFU is consistent with the registration data. Get a price quote; for questions, +90 505 580 46 22 and info@medexmedikal.com
Frequently Asked Questions
Is English IFU sufficient in Finland for a hospital device?
According to Act 719/2021 § 5, the information covered by MDR Article 10(11) may be in Finnish, Swedish or English; however, the information necessary for safe use must be available in Finnish and Swedish, and the manufacturer determines this through risk assessment. For a patient or consumer device, the instructions for use are in Finnish and Swedish.
Do I need to register with Fimea in CERE?
No. The first paragraph of § 49 of the Act covers manufacturers, authorised representatives and importers established in Finland; a manufacturer established in Türkiye registers in EUDAMED. A distributor in Finland makes its own notification under the third paragraph of § 49.
Must a distributor in Finland notify a device list?
According to Fimea's announcement of 2 June 2023, distributors that supply retailers and professional users are not required to submit a device notification until a further announcement; it is sufficient for them to enter their operator information in Fimea's electronic service.
In which languages must the implant card be?
According to Act 719/2021 § 5, the implant card information under MDR Article 18(1) must be in Finnish, Swedish and English.
What happens if an export notification is made incorrectly in ÜTS?
Under Article 17(3) of the Guide on Individual Movement Notifications, an incorrect Export Notification is cancelled with an Export Cancellation Notification and made again with the correct customs declaration number and quantity.