Exporting Medical Devices to Austria: ÜTS, EUDAMED, CE and Label Language

Austria is the second gateway of the German-speaking market after Germany, and the German label set prepared for Germany is used here too; however, the basis of the language requirement is a different law: the Medizinproduktegesetz 2021. This page explains, with official sources, the authorised representative, language, BASG, EUDAMED and ÜTS topics for a manufacturer established in Türkiye exporting medical devices to Austria.

Bevollmächtigter: is it needed for Austria

In Austrian documents the authorised representative appears as Bevollmächtigter. When hospital purchasing units and distributors in Vienna ask the Turkish manufacturer for this information, the Customs Union basis is shown.

"'Union market' refers to the territories of the Member States of the European Union and, owing to the European Economic Area (EEA), is extended to Norway, Liechtenstein and Iceland, and through the Customs Union Agreement to Türkiye."

Source: MDCG 2022-16 (Guidance on Authorised Representatives), Turkish translation, footnote 1.

MDR Article 11(1) ties the authorised representative requirement to a manufacturer not established in a Member State.

"Where the manufacturer of a device is not established in a Member State, the device may only be placed on the Union market if the manufacturer designates a sole authorised representative."

Source: Regulation (EU) 2017/745 (MDR), Article 11(1).

The definition in the Medical Device Regulation also counts Türkiye and the EU together.

"Authorised representative: a natural or legal person established in Türkiye or an EU member state who has been appointed in writing by the manufacturer to fulfil certain obligations under this Regulation on its behalf, where the manufacturer is not established in Türkiye or an EU member state"

Source: Medical Device Regulation, Article 4/1-lll.

The set of documents to give the Austrian buyer: the EUDAMED manufacturer registration (SRN), the EU declaration of conformity, the notified body certificate if any, and the trade registry record showing that the manufacturer is established in Türkiye. An authorised representative contract is not part of this set. Authorised representation is a need of manufacturers established outside Türkiye and the EU.

 

Label and instructions for use language: German (MPG 2021 § 7)

"Manufacturers shall ensure that the device is accompanied by the information set out in Section 23 of Annex I in an official Union language(s) determined by the Member State in which the device is made available to the user or patient. The particulars on the label shall be indelible, easily legible and clearly comprehensible to the intended user or patient."

Source: Regulation (EU) 2017/745 (MDR), Article 10, paragraph eleven.

For Austria, the Commission language table refers to the Medizinproduktegesetz 2021 of 30 June 2021 (Bundesgesetz betreffend Medizinprodukte): label and instructions for use for patients or lay users in German (§ 7 paragraph 1); for devices intended for professional users, German or English (§ 7 paragraph 1); implant card in German (§ 7 paragraph 4); declaration of conformity in German (§ 7 paragraph 2); field safety notice in German (§ 7 paragraph 6); conformity assessment documents in German or English (§ 7 paragraph 7 No. 1). By marking the Austria row with an asterisk, the Commission notes that up-to-date information has not been received from the country; the basis is the consolidated MPG 2021 text in the RIS.

Difference from Germany: in the Austria table only German is written for the declaration of conformity, whereas in Germany German or English is accepted. The German label and IFU prepared for Germany can also be used in Austria; however, the German version of the declaration of conformity must be prepared separately for Austria. For preparing the declaration of conformity, see the MDR and IVDR declaration of conformity page, and for the Turkish label rule, the article on Turkish labels and instructions for use.

 

BASG and the Ministry of Health

The Commission's May 2026 national authority contact list names two bodies for Austria: the Bundesministerium für Soziales, Gesundheit, Pflege und Konsumentenschutz (medicines and medical devices department, Vienna) and the Bundesamt für Sicherheit im Gesundheitswesen (BASG, Vienna). Since the BASG website was unreachable due to scheduled maintenance on the date this article was prepared, the rules on national operator or device registration in Austria could not be verified and have not been written on this page. Your Austrian distributor's national registration obligation should be confirmed directly with BASG once its site is accessible again.

"Member States may maintain or introduce national provisions on registration of distributors of devices which have been made available on their territory."

Source: Regulation (EU) 2017/745 on Medical Devices (MDR), Article 30, second paragraph.

 

EUDAMED actor and device registration for Austria

The documentary backbone of selling to Austria is EUDAMED: the manufacturer SRN and the device's Basic UDI-DI registration.

"Before placing a device, other than a custom-made device, on the market, manufacturers, authorised representatives and importers shall, in order to register, submit to the electronic system referred to in Article 30 the information referred to in Section 1 of Part A of Annex VI, provided that they have not already registered in accordance with this Article."

Source: (EU) 2017/745 MDR, Article 31, first paragraph.

"Not later than one year after submission of the information in accordance with paragraph 1, and every second year thereafter, the economic operator shall confirm the accuracy of the data."

Source: (EU) 2017/745 MDR, Article 31(5).

"Before placing a device, other than a custom-made device, on the market, the manufacturer shall, in accordance with the rules of the issuing entity referred to in Article 27(2), assign a Basic UDI-DI as defined in Part C of Annex VI to the device and shall provide it to the UDI database together with the other core data elements referred to in Part B of Annex VI related to that device."

Source: Regulation (EU) 2017/745 on Medical Devices (MDR), Article 29(1).

The provision that is the source of the transition timeline:

"no later than 12 months from the date of publication of the notice referred to in Article 34(3) in respect of the electronic system referred to in Article 33(2), points (a) and (b), manufacturers shall ensure that the information to be entered in Eudamed in accordance with Article 29 is entered in that electronic system, including regarding the following devices, provided that those devices are also placed on the market from 6 months from the date of publication of that notice:"

Source: MDR Article 123(3)(e) [(EU) 2024/1860, Article 1(3)(b)].

According to the Commission's EUDAMED page, Decision (EU) 2025/2371 was published on 27 November 2025 and the actor registration, UDI/device registration, notified bodies and certificates, and market surveillance modules have been mandatory since 28 May 2026; the 12-month period for device data ends at the end of November 2026. The steps are in the articles on EUDAMED actor registration, EUDAMED device registration and legacy device registration; the service content is EUDAMED registration consultancy.

 

ÜTS notification and free sale certificate for exports to Austria

Shipments to Austria are mostly made by road via Germany or Slovenia; what determines the ÜTS notification is not the route but the export declaration in Türkiye.

"An 'Export Notification' is made for individually tracked products exported by the company."

Source: TİTCK Guide on Individual Movement Notifications, Article 17, first paragraph.

"When making an Export Notification in ÜTS, after querying with the product number, lot and/or serial number, entering data in the Customs Declaration number and Quantity fields is mandatory."

Source: TİTCK Guide on Individual Movement Notifications, Article 17, second paragraph.

"If an Export Notification was made incorrectly, it is cancelled with an 'Export Cancellation Notification'."

Source: Guide on Individual Movement Notifications of the Product Tracking System for Medical Devices (TİTCK), Article 17, third paragraph.

The notification procedure is in the articles on ÜTS individual movement notifications and on export, free zone and bonded warehouse ÜTS registration; and on the monthly service individual notification operations page.

A free sale certificate is not a precondition for Austria; it is obtained from ÜTS if the buyer asks.

"Free Sale Certificate: the certificate issued by the Authority, upon the request of the domestic manufacturer or the Türkiye-established EU authorised representative of a foreign manufacturer, in order to export devices within the scope of the Medical Device Regulations, declaring that the relevant devices bearing the CE marking may be marketed in Türkiye and EU member states"

Source: ÜTS Registration Guide (TCKKD-KLVZ-03, 13.05.2026), Article 4, Definitions, item (zz).

The steps are in the free sale certificate article. The other gateway of the German-speaking market: Germany; neighbouring market: Italy.

 

Austria summary table

 

TopicStatusCompetent authorityBASG (Bundesamt für Sicherheit im Gesundheitswesen) and the Ministry of Health (Bundesministerium für Soziales, Gesundheit, Pflege und Konsumentenschutz), ViennaLabel / IFU languagePatient and lay user: German (MPG 2021 § 7/1); professional user: German or English (§ 7/1); declaration of conformity German (§ 7/2)National additional registration / notificationThe BASG site is under maintenance; the national registration rule could not be verified and should be confirmed with BASG. Manufacturer registration is in EUDAMEDIs an AR requiredNo (manufacturer established in Türkiye, Customs Union; MDCG 2022-16 footnote 1)

 

 

Medical Device ÜTS Registration Consultancy for exports to Austria

We handle the ÜTS product registration, export notifications, preparation of the German declaration of conformity and EUDAMED actor and device registration of manufacturers exporting to Austria. Send your quote request through the request a quote form; phone +90 505 580 46 22, e-mail info@medexmedikal.com.

Frequently Asked Questions

Is the German label I prepared for Germany valid in Austria?

For the label and instructions for use, yes: according to the Commission language table, MPG 2021 § 7 paragraph 1 provides German for patients and German or English for professional users. For the declaration of conformity, however, only German is written in the Austria row (§ 7 paragraph 2); a German declaration of conformity is prepared separately.

Do I need to appoint a Bevollmächtigter in Austria?

No. MDCG 2022-16 footnote 1 states that the Union market covers Türkiye through the Customs Union; MDR Article 11(1) provides for an authorised representative for a manufacturer not established in a Member State. A manufacturer established in Türkiye registers directly in EUDAMED as the manufacturer.

Should I make a national registration with BASG?

Since the BASG site was under maintenance on the date this article was prepared, the national registration rule in Austria could not be verified. MDR Article 30(2) gives Member States the authority for national rules only for distributor registration; as manufacturer, your EUDAMED registration is essential. The Austrian distributor should confirm its own obligation with BASG.

How often must I verify the SRN data?

MDR Article 31(5): verification of accuracy is made at the latest one year after the information is submitted and every two years thereafter; if no verification is made within six months, the Member State may take corrective measures in its territory.

How is the ÜTS notification made for a consignment going by road via Slovenia?

The route does not matter. Under Article 17 of the Individual Movement Notifications Guide, the Export Notification is made by entering the export declaration number in Türkiye and the quantity; the Export Cancellation Notification is used in case of error.

Ehlibeyt Mahallesi, Tekstilciler Caddesi No: 35/7 06520 Çankaya / Ankara - TÜRKİYE

Cevizli Mahallesi, Zuhal Caddesi Ritim İstanbul B Blok, No:44 Maltepe / İstanbul - TÜRKİYE

info@medexmedikal.com
+90 505 580 4622
+90 312 389 1354

ÜTS Guides · Legislation Calendar · Glossary · ÜTS Guide Index · Frequently Asked Questions

Services: ÜTS Consultancy · Who Carries Out the ÜTS Registration? · ÜTS Document Registration Consultancy · ÜTS Product Registration Consultancy · Importer ÜTS Registration Consultancy · Sales Centre Authorisation Certificate · EUDAMED Registration Consultancy · MDR/IVDR Declaration of Conformity · Technical File Preliminary Review · Label and Instructions for Use Check · PRRC Service · PMS and PSUR · Clinical Evaluation · EMDN and GMDN Code Determination · Certificate of Free Sale · EUDAMED XML Generator · ÜTS Registration Training · Correction of a Rejected ÜTS Application

KVKK and Privacy Policy | Cookie Policy | Accessibility Statement

© 2026 Medex Kurumsal Danışmanlık Ltd. Şti.

bottom of page
WhatsApp Hemen Ara