What Is a Single-Use Device? Its Definition in the Legislation
In Article 3 of the Medical Device Regulation, a single-use device is defined as a device intended to be used on one individual during a single procedure.

Article 3 of the Medical Device Regulation defines a single-use device as a device that is intended to be used on one individual during a single procedure. The TİTCK Guide repeats the same definition word for word. Two limits stand together in the definition: a single procedure and a single person. That a device is not used a second time on the same patient, and not used on different patients, are both consequences of this definition. The single-use character does not come from the material of the device but from the manufacturer’s declaration. If the manufacturer has intended the device as single-use, the fact that the device looks physically durable does not change this character.
Why it matters
When a device is registered in ÜTS, the “Single use?” field is mandatory. The Guide states that, if the device concerned is single-use, this information must appear both on the label and/or sales packaging and in the instructions for use, and that this field in ÜTS must also be selected as “Yes”. So the label, the instructions for use and the system record must be consistent with each other. When the field is selected as “Yes”, ÜTS fills in and locks some fields automatically: the limited number of uses field is set to “Yes”, the number of uses to “1”, and the other related fields are selected as “No”. In this case the fields concerned cannot be changed. The same logic works when the implantable device field is selected as “Yes”.
Reprocessing
Article 17 regulates the reprocessing of single-use devices. Under the first paragraph, reprocessing and further use are carried out in accordance with the guides published by the Authority and with this article. The second paragraph sets out the critical consequence: a natural or legal person who reprocesses a single-use device to make it suitable for further use is considered to be the manufacturer of the reprocessed device and assumes the obligations incumbent on manufacturers, including the obligations relating to traceability. The other paragraphs draw the limits:
• Seventh paragraph: single-use devices may only be reprocessed in a way considered safe on the basis of the latest scientific evidence.
• Fifth paragraph: conformity with the common specifications or, in their absence, with the relevant harmonised standards and the relevant legislation is certified by a notified body.
• Eighth paragraph: the name and address of the reprocessor and the other relevant information in point 23 of Annex I are indicated on the label of the reprocessed device and, where applicable, in its instructions for use. The name and address of the manufacturer of the original single-use device no longer appear on the label, but are indicated in the instructions for use of the reprocessed device.
• Ninth paragraph: the Authority may introduce provisions stricter, more restrictive or prohibitive than those in the Regulation on reprocessing and further use.
The third paragraph provides a limited exception for single-use devices that are reprocessed and used within a health institution; however, this exception is subject to the condition that the safety and performance of the reprocessed device are equivalent to those of the original device and that the common specifications detailing the requirements for risk management, process validation, product release and performance testing, the quality management system, adverse event reporting and traceability are complied with.
Relationship with devices with a limited number of uses
The Guide defines the limited number of uses as “the total number of tests or uses that can be performed with the device (a single product in the content)”. For a single-use device this number is necessarily 1. A device with a limited number of uses greater than 1 is not single-use; for such devices the number concerned is entered in ÜTS as indicated on the label and in the instructions for use.
Often confused
• Multiple-use device on a single patient: a device that can be used in more than one procedure on the same patient does not meet the “single procedure” criterion in the definition, so it is not single-use.
• Sterile device: sterility is a separate characteristic. There are non-sterile single-use devices, and also multiple-use devices that can be re-sterilised.
• Reusable surgical instrument: it is defined separately in the Guide and is treated as Class Ir under the MDR.
• Reprocessed device: because the reprocessing party is considered the manufacturer, it is not the same product as the original device.
Who is affected: manufacturers, importers, health institutions, parties carrying out sterilisation and reprocessing, and firms carrying out ÜTS registration.
Often confused with: a multiple-use device on a single patient, a sterile device and a reusable surgical instrument.
Basis: For medical device ÜTS registration processes, see our Medical Device ÜTS Registration Consultancy service.