What Is a Notified Body and What Does It Do?

Article 3 of the Medical Device Regulation defines a notified body as a conformity assessment body designated under the Regulation.

O

Article 3 of the Medical Device Regulation defines a notified body as a conformity assessment body designated under the Regulation. The same article defines a conformity assessment body as a body that performs third-party conformity assessment activities including calibration, testing, certification and inspection. In short, a notified body is a third-party assessment body that has been officially designated under the relevant regulation. Under Article 35, the authority responsible for planning or designating a body as a notified body to carry out conformity assessment activities is the Authority. The Authority is also responsible for monitoring notified bodies, including their subcontractors, branches and representative offices.

Why it matters

Whether a notified body will be involved in the process before a device can bear the CE marking is determined by the risk class of the device. Under the sixth paragraph of Article 52, manufacturers of Class I devices declare the conformity of their products themselves by drawing up the EU declaration of conformity after preparing the technical documentation. However, if these devices are placed on the market in a sterile condition, have a measuring function or are reusable surgical instruments, a notified body is involved; its involvement is limited, respectively, to the aspects relating to establishing sterile conditions, conformity with metrological requirements and the aspects relating to the reuse of the device. Class IIa, IIb and III devices are subject to notified body assessment under the procedures in the third, fourth and fifth paragraphs of the same article.

On the ÜTS side, notified body information is reflected directly in document registration. Under Article 6 of the TİTCK Registration Guide, the name, identification number and address information of the notified body are among the fields checked during document review in ÜTS at the time of certificate registration, and the information entered into ÜTS must be the same as the information on the certificate. For certificates issued under the new medical device regulations, the document notification is made under the type “EU Certificate (MDR/IVDR)”; the type “EC Certificate” is used only for certificates issued under the old medical device regulations.

Designation and identification number

Under Article 38, conformity assessment bodies established in Türkiye apply to the Authority to be designated. Under Article 43, an identification number is assigned by the Commission to each notified body whose notification becomes valid; a single identification number is given even if the body is notified under more than one legislation. Bodies notified under the repealed Medical Device Regulation and the Active Implantable Medical Devices Regulation keep their previous identification numbers when they are designated under this Regulation. NANDO, mentioned in the definitions article of the Regulation, is the European Union’s New Approach Notified and Designated Organisations Information System.

Obligations and supervision

Under Article 36, notified bodies carry out the tasks for which they are designated, meet the organisational and general requirements and the quality management, resource and process requirements, and in particular comply with Annex VII. Under the same article, they permanently have available, in sufficient numbers, administrative, technical and scientific personnel and personnel with relevant clinical knowledge; the personnel referred to in points 3.2.3 and 3.2.7 of Annex VII cannot be external services or contractors. Under Article 37, they may subcontract certain of their tasks or use a branch or representative office; in that case they verify that the relevant requirements are met, inform the Authority and take full responsibility for the work carried out. The list of branches and representative offices is kept publicly available. Under Article 44, they notify the Authority without delay, and at the latest within 15 days, of all changes that may affect their compliance with the requirements of Annex VII or their ability to carry out conformity assessment activities. Under the second paragraph of Article 36, they keep all relevant documentation, including the manufacturer’s documentation, and submit it to the Authority upon request.

Often confused

• Notified body and competent authority: the competent authority in Türkiye is the Authority; it is the party that designates and monitors the notified body and takes measures where necessary. The notified body is the third party that carries out conformity assessment.

• Who affixes the CE marking: the notified body does not affix the CE marking; under the sixth paragraph of Article 10, it is the manufacturer that draws up the EU declaration of conformity and affixes the CE marking.

• The assumption that it is needed for every device: for some Class I devices, notified body involvement is not required; the sterile, measuring function and reusable surgical instrument cases are exceptions.

• Certificate type: certificates under the MDR are notified to ÜTS under the type “EU Certificate (MDR/IVDR)”; the type “EC Certificate” is for old regulation certificates.

Who is affected: manufacturers, authorised representatives, importers and firms that make certificate notifications in ÜTS.

Often confused with: the competent authority, the accreditation body and the test laboratory.

Basis: For medical device ÜTS registration processes, see our Medical Device ÜTS Registration Consultancy service.

Actors

Medical Device ÜTS Registration Consultancy

Request an ÜTS registration consultancy quote

Ehlibeyt Mahallesi, Tekstilciler Caddesi No: 35/7 06520 Çankaya / Ankara - TÜRKİYE

Cevizli Mahallesi, Zuhal Caddesi Ritim İstanbul B Blok, No:44 Maltepe / İstanbul - TÜRKİYE

info@medexmedikal.com
+90 505 580 4622
+90 312 389 1354

ÜTS Guides · Legislation Calendar · Glossary · ÜTS Guide Index · Frequently Asked Questions

Services: ÜTS Consultancy · Who Carries Out the ÜTS Registration? · ÜTS Document Registration Consultancy · ÜTS Product Registration Consultancy · Importer ÜTS Registration Consultancy · Sales Centre Authorisation Certificate · EUDAMED Registration Consultancy · MDR/IVDR Declaration of Conformity · Technical File Preliminary Review · Label and Instructions for Use Check · PRRC Service · PMS and PSUR · Clinical Evaluation · EMDN and GMDN Code Determination · Certificate of Free Sale · EUDAMED XML Generator · ÜTS Registration Training · Correction of a Rejected ÜTS Application

KVKK and Privacy Policy | Cookie Policy | Accessibility Statement

© 2026 Medex Kurumsal Danışmanlık Ltd. Şti.

bottom of page
WhatsApp Hemen Ara