What Is IVD Device Classification?

IVD device classification is the division of in vitro diagnostic devices into four classes, Class A, B, C and D, taking into account the risks inherent in the intended purpose and structure of the device.

I

IVD device classification is the division of in vitro diagnostic devices into four classes, Class A, Class B, Class C and Class D, taking into account the risks inherent in the intended purpose and structure of the device. This structure rests on the In Vitro Diagnostic Medical Device Regulation published in the Official Gazette No. 31499 bis of 02.06.2021. Article 57 of the TİTCK Guide on the Registration Processes of Medical Devices in the Product Tracking System (TCKKD-KLVZ-03, Rev.00, in force from 13.05.2026) lists these four classes and a total of 7 classification rules. IVD devices are subject to a regulation separate from that for medical devices; the Class I, IIa, IIb, III structure of medical devices does not apply to them.

Why it matters

The class determines whether a notified body will be involved in the conformity assessment. The class determines the option to be selected in ÜTS and the set of documents to be associated with the device. For Class D devices, the EU reference laboratory and expert panel mechanisms may come into play.

What the seven rules say

The rules in Article 57 of the Guide are, in summary, as follows:

• Rule 1 (Class D): detection of the presence of, or exposure to, a transmissible agent in blood, blood components, cells, tissues or organs for assessing their suitability for transfusion, transplantation or cell administration; detection of a transmissible agent that causes a life-threatening disease with a high or suspected high risk of propagation; determining the infectious load of a life-threatening disease where monitoring is critical in the process of patient management.

• Rule 2 (Class C, Class D for certain markers): blood grouping, determination of foetomaternal blood group incompatibility or tissue typing. Class D if intended for determining markers of the ABO, Rhesus, Kell, Kidd and Duffy systems.

• Rule 3 (Class C): the cases listed in the rule, such as detection of sexually transmitted agents, use as a companion diagnostic, cancer screening and staging, and human genetic testing.

• Rule 4 (Class C, Class B for some tests): devices for self-testing. Devices intended for the detection of pregnancy, fertility testing, determining cholesterol level, and the detection of glucose, erythrocytes, leucocytes and bacteria in urine are Class B. Near-patient testing devices are classified in their own right.

• Rule 5 (Class A): general culture media and histological stains, products for general laboratory use, accessories without a critical characteristic, buffer solutions, washing solutions, instruments intended specifically for in vitro diagnostic procedures, and specimen receptacles.

• Rule 6 (Class B): all other in vitro diagnostic devices.

• Rule 7 (Class B): controls without a quantitative or qualitative assigned value.

The general principles in Article 52 of the Guide also apply: accessories of an IVD device are classified in their own right; calibrators intended to be used with a device are in the same class as the device; control materials with an assigned value are in the same class as the device; and each of the rules applies to first-line assays, confirmatory assays and supplemental assays.

Is a notified body involved

Article 62 of the Guide describes the route by class. For Class A devices, the manufacturer draws up the declaration of conformity after preparing the technical documentation specified in Annexes II and III. If the device is placed on the market in a sterile condition (Class A Sterile), Chapters I and III of Annex IX or one of the procedures in Annex XI are also carried out; however, the involvement of the notified body is limited to the aspects relating to establishing, securing and maintaining sterile conditions. For Class B, Class C and Class D devices, the notified body is involved. For Class D devices, the notified body requests one of the EU reference laboratories to verify the performance by laboratory testing.

Effect on ÜTS registration

Selecting the “Class” field in ÜTS is mandatory, and the options that can be selected under the IVDR are Class A Sterile, Class A, Class B, Class C, Class D, Self-Test Device (Class B/C/D), Near-Patient Test Device (Class B/C/D) and Companion Diagnostic Device (Class C/D). Article 67 of the Guide provides different document business rules according to these options.

Often confused

• The IVD classes A, B, C, D cannot be matched to the I, IIa, IIb, III structure of medical devices.

• Class A does not mean exemption. For Class A Sterile devices, the notified body is involved, limited to sterility.

• A self-test device and a near-patient test device are not the same thing. A self-test device is designed to be used by lay persons; a near-patient test device is designed to be used by a healthcare professional outside a laboratory environment.

• The Annex II List A and List B structure of the repealed IVDD does not match the IVDR classes one-to-one.

• Article 48 of the Guide sets the transitional dates, provided the conditions are met, as 31.12.2027 for Class D, 31.12.2028 for Class C, and 31.12.2029 for Class B and Class A Sterile.

Who is affected: firms that manufacture or import in vitro diagnostic devices, reagent and kit suppliers, and regulatory units with a laboratory products portfolio.

Often confused with: matching IVD classes to medical device classes, and treating a self-test device and a near-patient test device as the same category.

Basis: For medical device ÜTS registration processes, see our Medical Device ÜTS Registration Consultancy service.

Classification

Medical Device ÜTS Registration Consultancy

Request an ÜTS registration consultancy quote

Ehlibeyt Mahallesi, Tekstilciler Caddesi No: 35/7 06520 Çankaya / Ankara - TÜRKİYE

Cevizli Mahallesi, Zuhal Caddesi Ritim İstanbul B Blok, No:44 Maltepe / İstanbul - TÜRKİYE

info@medexmedikal.com
+90 505 580 4622
+90 312 389 1354

ÜTS Guides · Legislation Calendar · Glossary · ÜTS Guide Index · Frequently Asked Questions

Services: ÜTS Consultancy · Who Carries Out the ÜTS Registration? · ÜTS Document Registration Consultancy · ÜTS Product Registration Consultancy · Importer ÜTS Registration Consultancy · Sales Centre Authorisation Certificate · EUDAMED Registration Consultancy · MDR/IVDR Declaration of Conformity · Technical File Preliminary Review · Label and Instructions for Use Check · PRRC Service · PMS and PSUR · Clinical Evaluation · EMDN and GMDN Code Determination · Certificate of Free Sale · EUDAMED XML Generator · ÜTS Registration Training · Correction of a Rejected ÜTS Application

KVKK and Privacy Policy | Cookie Policy | Accessibility Statement

© 2026 Medex Kurumsal Danışmanlık Ltd. Şti.

bottom of page
WhatsApp Hemen Ara