Manufacturer Obligations (Medical Device Regulation Art. 10)
Manufacturers are obliged to ensure, when placing their devices on the market or putting them into service, that the device has been designed and manufactured in accordance with the requirements of the Regulation. Article 10 of the Medical Device Regulation defines the whole area of the manufacturer's responsibility in sixteen paragraphs around this basic obligation. The article builds a chain of linked obligations, from risk management to technical documentation, from the quality management system to post-market surveillance and financial coverage. The obligation is not a one-off: it continues as long as the device is on the market.
Design, risk and clinical evaluation
The first paragraphs of the article determine which processes the device must have gone through before it reaches the market.
• Manufacturers establish, document, implement and maintain a risk management system as defined in point 3 of Annex I (Article 10, second paragraph).
• Clinical evaluation, including post-market clinical follow-up, is carried out in accordance with Article 61 and Annex XIV (third paragraph).
• For devices other than custom-made devices, technical documentation containing the points specified in Annex II and Annex III is prepared and kept up to date (fourth paragraph).
• Manufacturers of custom-made devices prepare documentation under point 2 of Annex XIII and keep it to present to the Authority on request (fifth paragraph).
Declaration of conformity, CE marking and registration
When the conformity assessment procedure is completed, the manufacturer draws up the EU declaration of conformity under Article 19 and affixes the CE conformity marking under Article 20 (sixth paragraph). The seventh paragraph also requires compliance with the UDI obligations in Article 27 and the registration obligations in Articles 29 and 31. These three references form the legal basis of all procedures on the ÜTS and UDI side.
The documentation retention periods are clear in the eighth paragraph:
• Copies of the technical documentation, the EU declaration of conformity and the certificates issued under Article 56 are kept for at least 10 years after the last device covered by the declaration has been placed on the market.
• For implantable devices, this period is at least 15 years.
• If the registered place of business is outside Türkiye or the EU member states, the manufacturer ensures that its authorised representative keeps the necessary documentation permanently available.
Quality management system and post-market processes
The ninth paragraph, with the exception of investigational devices, makes it mandatory to establish and continuously improve a quality management system proportionate to the risk class and type of the device. Among the minimum elements of the system are the regulatory compliance strategy, management responsibility, selection of suppliers and contractors, risk management, clinical evaluation, verification of UDI assignment and post-market surveillance. The tenth paragraph requires the post-market surveillance system to be implemented under Article 81, and the thirteenth paragraph requires a system for recording and reporting adverse events and field safety corrective actions under Articles 85 and 86. The twelfth paragraph regulates the obligation to bring the device into conformity, withdraw it from the market or recall it when a suspicion of nonconformity arises, and to inform distributors and importers.
Contract manufacturing, financial coverage and supply interruption
According to the fifteenth paragraph, if the manufacturer has its device designed or manufactured by another natural or legal person, the information on the identity of this person becomes part of the information to be submitted under the fourth paragraph of Article 29. The sixteenth paragraph provides for sufficient financial coverage proportionate to the risk class, device type and size of the business. In addition, Article 10/A, added by the amendment of 17.08.2024, requires that, when a supply interruption or discontinuation that may create a risk of serious harm to patients or public health is foreseen, the Authority and the supply chain be informed, and, except in exceptional circumstances, that this information be given at least six months before the interruption.
Effect on the ÜTS record
The ÜTS record is only a part of the obligations in Article 10. After opening its firm record in ÜTS, the domestic manufacturer makes a document record of the type EU Declaration of Conformity (MDR/IVDR); this application is sent with the “Application for Documents with e-signature for Manufacturer Firms” type and with the electronic signature of the person registered as Responsible Manager or Signatory on behalf of the firm. In the product record, the UDI Issuing Entity, UDI-DI Number and Basic UDI-DI fields are filled in. For manufacturers who have their products made by contract manufacturing, the title and address of the contract manufacturer are expected to appear in the EU declaration of conformity or the EU certificate; if this information is not in the documents and on the label, the Authority may request that technical file pages be uploaded. After the registration is completed, the obligations of risk management, keeping the technical documentation up to date, post-market surveillance and vigilance continue.
Common mistakes
• Thinking the ÜTS record is the final goal; not setting up post-registration surveillance and vigilance processes.
• Leaving the technical documentation at its first version; not reflecting changes in design, standards or common specifications in the documentation.
• Confusing the retention periods; applying 10 years to implantable devices.
• In contract manufacturing, the title and address of the contract manufacturer not appearing in the declaration or the certificate.
Who is affected: firms that manufacture medical devices in Türkiye and place them on the market under their own name; firms that have products made by contract manufacturing and sell them under their own brand are also considered manufacturers; and authorised representatives that keep documentation on behalf of a manufacturer abroad.
Common mistake: Overlooking that the risk management, technical documentation and post-market surveillance obligations continue after the ÜTS record is completed.
Basis: Medical Device Regulation (OG 02.06.2021 / 31499 bis), Article 10 and Article 10/A; Annex I, Annex II, Annex III, Annex XIII, Annex XIV; referenced Articles 19, 20, 27, 29, 31, 56, 61, 81, 85 and 86.