ÜTS Product Image and Label Rules

A device application can end negatively because of a blurry photograph or an incomplete label, even when there is no problem at all in the technical file

ÜTS Product Image and Label Rules

A device application can end negatively because of a blurry photograph or an incomplete label, even when there is no problem at all in the technical file. The product image and label fields in ÜTS are mandatory, and their content is defined item by item in the Guide on the Registration Processes of Medical Devices to the Product Tracking System. Below, based on the source, we explain how many images are uploaded, which image is not accepted and what information is looked for on the label. If you would like your label and instructions for use to be checked before registration, see our label and instructions for use check service.

Product Images Are a Mandatory Field

Uploading images is not an optional job. In paragraph 6 of Article 25 of the Guide, which regulates registration under the old regulations, Article 27, which regulates MDR registration, and Article 30, which regulates IVDR registration, the "Product Images" field is defined as a mandatory field to be selected in all three frameworks.

Photographs and/or technical drawings of the relevant device that clearly show the device are uploaded to this field. The rule is the same in all three articles: only images of the relevant device are placed in this field. Filling in the image block incompletely or wrongly is a deficiency sufficient on its own for the application to end negatively.

How Many Images, in Which Format?

The rules on quantity, format and clarity are the same in all three frameworks and are in paragraph 6 of Articles 25, 27 and 30 of the Guide.

  • Quantity: A maximum of 10 product images can be uploaded to this field.

  • Format: The product image is uploaded as a jpeg, jpg, pdf or png file, without exceeding the maximum file size specified in ÜTS.

  • Clarity: If images that are not clear and cannot be read are uploaded, the application ends negatively.

The clarity condition is the most overlooked rule in practice. A photograph in which the barcode on the label, the expiry date symbol or the lot field cannot be read makes the label-to-record comparison impossible during review. It is worth opening the images at full size before uploading and checking that the text on them can be read.

Why Is a Photograph of the Packaging Alone Not Enough?

Paragraph 6 of Article 25 of the Guide directly prohibits a frequent mistake: only the package or packaging of the device is not uploaded as the product image. Images that clearly show a photograph or technical drawing of the device must also be uploaded to this field together with the package/packaging images. The same provision is repeated in Articles 27 and 30 for records under the MDR and IVDR.

  • If there is an image of the device on the packaging: If the package or packaging of the device bears the device's own image, this package/packaging image can be uploaded to this field.

  • Exception: If opening the device package or packaging would impair properties such as the sterility and reliability of the device and/or if the device is in liquid form, only the package/packaging image can be uploaded to this field.

The exception is interpreted narrowly. If you cannot photograph the inner product without opening the package of a sterile single-use device, the packaging image is sufficient; the same justification is not accepted for a non-sterile, reusable device.

Images of Other Devices and Catalogue Images Are Not Uploaded

There are two clear prohibitions in the image block, and both are the same in all three frameworks.

  • Images of other devices: Under paragraph 6 of Article 25 of the Guide, images of other devices are strictly not uploaded to this field, even if they belong to similar devices.

  • Catalogue and multi-device images: The catalogue itself, or an image containing more than one device that makes it difficult to understand the device, is not uploaded as a product image.

Together, these two rules rule out a frequently used shortcut: uploading the product family's collective catalogue page unchanged to every record. Because the registration is made with a single barcode, the image must also show the device corresponding to that barcode.

Image File Naming Rule

The Guide also controls the name of the file you upload. According to paragraph 6 of Articles 25, 27 and 30 of the Guide, the image file names must be named so as to describe the device and the image; OuterBox.png, inner_box.jpeg and product-image.pdf are given as examples.

The same subparagraph also states that commercial extensions and expressions such as "WhatsApp" must not appear in image names; names in the form whatsapp123abcde.jpg are given as examples. The most practical way of complying with this article is to rename images coming from a phone according to the device and image content rather than uploading them directly.

Images for Software Devices

For devices without a physical product the Guide gives a separate solution. According to paragraph 6 of Article 25 of the Guide, if the device is stand-alone software, a screenshot taken from the software interface can be uploaded to this field as the image.

In records under the MDR the same subparagraph asks for one more item: according to paragraph 6 of Article 27 of the Guide, if the device is stand-alone software, a screenshot and the software identity can be uploaded to this field. The set of images to be prepared for software medical devices on the MDR side is therefore one item larger.

Sağlık Market / SGK Image Standards

If your device will be sold within Sağlık Market under the Government Supplies Office (Devlet Malzeme Ofisi) or will be reimbursed by SGK, additional standards come into play for the image block. Paragraph 7 of Article 25 of the Guide lists the rules that apply when the field "Will It Be Sold in Sağlık Market / Will SGK Reimbursement Be Made?" is selected as "Yes". The same rules are in the "Product Image Standards" subparagraph of Article 27 on the MDR side and Article 30 on the IVDR side.

  • Minimum number of photographs: There must be at least 3 photographs for each device.

  • Package types: For devices with inner and outer packaging, there must be at least 3 photographs of each package type.

  • Components: If there are any, images of the device's spare parts, consumables, accessories and all components must also be provided; as an example the Guide gives the barrel, needle tip and needle cap of a three-piece syringe.

  • Technical specifications: Each photograph must be at least 600x600 pixels, square, at most 2 MB, and in one of the jpeg, jpg or png formats.

  • Shooting angles: The photographs must show all details according to the way the device is used and be taken from different directions.

The square and pixel requirement here is not in the general image rule. Companies that mark the Sağlık Market option "Yes" need to prepare their image set to these dimensions from the start.

The "Is the Original Label in Turkish?" Question and the Fields It Opens

The label block begins with a single question, and the answer to that question determines the rest of the screen. According to paragraph 7 of Articles 25, 27 and 30 of the Guide, the "Is the Original Label in Turkish?" field is a field that must be selected in ÜTS.

  • Yes: If the original market placement label and sales package prepared by the manufacturer are prepared in Turkish or in several languages including Turkish, this field is selected as "Yes" and the current, original Turkish market placement label of the device is uploaded together with the inner/outer sales package if necessary.

  • No: If the original label is prepared in a foreign language, this field is selected as "No" and the current, original market placement label of the device is uploaded together with the inner/outer sales package if necessary.

  • Difference in field name: In a record under the old regulations the name of the field is "Original Label". In MDR and IVDR records, according to Articles 27 and 30 of the Guide, the name of the field is "Original Label and Sales Package".

For notifications concerning domestic manufacturers, on the MDR and IVDR side the question is answered directly "Yes" and the current and original Turkish label is uploaded; for applications concerning foreign manufacturers, the language of the label determines the answer.

ÜTS Generates the Turkish Label; the Company Does Not Upload It

This is one of the most commonly misunderstood subjects. According to paragraph 7 of Article 25 of the Guide, when the "Is the Original Label in Turkish?" field is selected as "No", after the device registration/update application is made, the "Turkish Label" field is generated automatically by ÜTS on the registration screen.

  • How it is generated: This field is generated by automatically processing the label information of the device through the information entered into ÜTS.

  • The company does not upload it: Companies therefore do not upload a separate Turkish label. This is why entering the product information correctly into ÜTS is decisive.

  • Consistency requirement: The information on the label uploaded to the "Original Label" field and in the content of the inner/outer packaging, if any, must be consistent with the information entered into ÜTS.

  • If "Yes": When the question is selected as "Yes", this field does not appear on the ÜTS screen and is not generated automatically.

The same mechanism is repeated in paragraph 7 of Article 27 of the Guide for records under the MDR and in paragraph 7 of Article 30 for records under the IVDR.

The Template Generated by ÜTS Does Not Replace the Original Label

The "Turkish Label" subparagraph in paragraph 7 of Articles 25, 27 and 30 of the Guide ends with a one-sentence warning: the template generated in this field does not replace the original label.

Its practical meaning is this: printing the Turkish template generated by ÜTS and sticking it on the product does not meet the original label obligation. The template is a registration output; the market placement label of the device is the original label prepared by the manufacturer.

Information Required on the MDR Label

According to paragraph 7 of Article 27 of the Guide, in the assessment of the device registration application, the common information required to be on the label and/or sales package, as set out in "Chapter III of Annex I, General Safety and Performance Requirements" of the MDR, is checked.

  • Identifying information: The product name or trade name of the device; details necessary for the user to identify the device, the contents of the package and, where this is not obvious to the user, the intended purpose.

  • Economic operators: The name, registered trade name or registered trade mark and the address of the place of business of the manufacturer; if the manufacturer is established outside the EU and Türkiye, the name and address of the EU authorised representative; for imported devices, the name and address of the importer if they are not in the instructions for use or on the sales package.

  • Indications of special content: If the device contains a medicinal substance, including a human blood or plasma derivative, tissues or cells of human origin, or tissues or cells of animal origin, indications to this effect.

  • CMR and endocrine-disrupting substances: If the device contains CMR and/or endocrine-disrupting substances and parts of the device or the materials used contain these substances at a concentration above 0.1% by weight, information indicating the presence of these substances and the list of the substances.

  • UDI carrier: The UDI carrier, taking into account the dates of application in point (c) of paragraph 1 of Article 111 of the MDR.

  • Time information: Where relevant, a clear indication, at minimum in years and months, of the time limit for safely using or implanting the device; if there is no such indication, the date of manufacture.

  • Conditions and status indications: Any specific storage and/or handling conditions; if the device is supplied sterile, an indication of its sterile status and the sterilisation method; if single-use, an indication to that effect.

  • Warnings: Warnings or precautions that need to be brought to the immediate attention of the user of the device and of any other person. If there is a reference to the instructions for use on the label or packaging and the information is explained in detail in the instructions, this information may be kept to a minimum.

  • CE marking and device indication: The CE marking and, if a notified body has been involved in the conformity assessment procedure, the four-digit notified body number; an indication that the device is a medical device and other information required on the label under the MDR.

Information Required on the IVDR Label

Paragraph 7 of Article 30 of the Guide sets up the same structure for the IVDR: in the assessment, the common information required to be on the label and/or sales package, as set out in "Chapter III of Annex I, General Safety and Performance Requirements" of the IVDR, is checked. The list largely overlaps with that of the MDR, but includes items specific to in vitro diagnostic devices.

  • Net content quantity: Where relevant, an indication of the net content quantity expressed by weight or volume, by numerical count or by other units that accurately reflect the content of the package; this includes the number of products in the contents and, if any, the number of limited uses.

  • Special operating instructions: Special operating instructions, where applicable.

  • Device type indication: An indication that the device is an in vitro diagnostic device, or, if the device is intended for a performance study, an indication to that effect.

  • Self-testing and near-patient testing indication: If the device is intended for self-testing or near-patient testing, an indication to that effect; a clear statement of the exclusion if rapid tests are not intended for self-testing or near-patient testing.

  • Kit contents: If device kits contain individual reagents and substances that may be made available as separate devices, each of these devices complies with the labelling requirements of Annex I of the IVDR.

  • Hazardous substance labelling: For devices containing substances or mixtures that may be considered hazardous, the hazard pictograms and labelling requirements of the Regulation on the Classification, Labelling and Packaging of Substances and Mixtures apply.

The field "Is it a device for self-testing or near-patient testing?" in the same paragraph is linked to this indication: if the device is a self-testing or near-patient testing device, it is mandatory for this information to be on the label and/or sales package and in the instructions for use and for the field to be selected as "Yes".

Label Information Under the Old Regulations

Paragraph 7 of Article 25 of the Guide lists, as a separate list, the common information required on the label and/or sales package for devices under the old regulations; this list applies together with the additional information in the "Annex I Essential Requirements" section of the relevant regulation.

  • Economic operators: The name or trade name and address of the manufacturer; for imported devices, the name and address of the importer if not in the instructions for use; if the manufacturer is established outside the EU and Türkiye, the name and address of the EU authorised representative.

  • Product description: The product name and/or trade name and the details, particularly for the user, that describe the contents of the package and the device.

  • Sterility: If the device is supplied sterile, the word "STERILE" and, if the sterilisation method is known, the sterilisation method used in the packaging.

  • Time: If the device has a shelf life, the expiry date in months and years, or the shelf life directly.

  • Single use and storage: If the device is single-use, the indication "Single Use" and any specific storage conditions.

  • CE marking: The CE marking and, if involved in the conformity assessment procedure, the four-digit notified body number.

  • Warnings and other information: Warnings and/or precautions concerning the device and its use, and other information required on the label under the regulation to which it belongs.

Can Information That Does Not Fit on the Label Be Written on the Inner/Outer Packaging?

The Guide gives a conditional yes to this question. According to paragraph 7 of Articles 25, 27 and 30 of the Guide, some information required to be on the label may be placed on the inner and/or outer sales package, for technical reasons such as label size or at the manufacturer's preference.

The ÜTS counterpart is an upload obligation: if the original label does not contain all of the information specified in the relevant regulation, the inner/outer sales package of the device must also be uploaded to this field. The sales package at whichever packaging level the information is found is uploaded. According to the same subparagraphs, the information needed for the safe use of the device must be present on the device itself and/or on the packaging of each part, or, where appropriate, on the commercial packaging.

Label Template Exception for Contact Lenses and Optical Products

For this product group the label is not printed so as to carry the specific values of each product individually. Taking this into account, the Guide has introduced a special provision.

According to paragraph 7 of Article 25 of the Guide, only for optical products of the "contact lens" type, provided that specific information such as primary product number, version/model, reference/catalogue number, product name and/or dioptre, or other information that must be on the label but is not in the template, is entered in the product description field and/or is stated in the declaration of conformity or added under the uploaded label, the official label and sales package template prepared by the manufacturer may be uploaded to this field in place of the market placement label and sales package. Article 27 of the Guide establishes the same exception for optical products under the MDR through the UDI-DI number.

Label–Record Consistency: Barcode, Shelf Life, Storage Condition

In the ÜTS review the label is not assessed alone but by comparison with the record. The four headings below turn the consistency between the label and the ÜTS record directly into a rule.

  • Barcode: According to paragraph 2 of Articles 25 and 27 of the Guide, the allocated barcode of the device must be entered into the system exactly as it is, without any change. An application for a device record whose barcode is not entered as it appears in its technical documents (such as the label/sales package, the declaration of conformity and the certificate) ends negatively.

  • Shelf life and expiry date: According to paragraph 7 of Article 25 of the Guide, if the device has a shelf life, it is mandatory for the original label or sales package to carry an expiry date. This information may not be directly on the label; it may be shown indirectly by the expiry and manufacturing dates. The shelf life information reflected on the label and in the product images must be consistent. The same provision is in Article 27 on the MDR side.

  • Storage and handling condition: According to paragraph 9 of Articles 25 and 27 of the Guide, the storage condition entered into the system must be in the instructions for use, whether or not it has specific values. If the condition has specific values, it must also be on the label and/or sales package; if not, it is not looked for on the label. For the handling condition there is an obligation to be in the instructions, but not on the label.

  • Latex and phthalate/DEHP: According to paragraph 7 of Articles 25 and 27 of the Guide, if the device or its components contain latex, a warning about this must be on the label and/or sales package of the device and in the instructions for use. The same rule applies to phthalate/DEHP; if the instructions for use give a more detailed explanation, this information may be kept to a minimum on the label or packaging.

Label Display for CMR/ED Substances

In records under the MDR and IVDR there is a separate block titled "CMR and/or Endocrine-Disrupting Substance Information". According to paragraph 10 of Article 27 of the Guide, if the device contains CMR and/or endocrine-disrupting substances as set out in point 10.4.1 of Annex I of the MDR, the field is selected as "Yes" and it becomes mandatory to add the substances by selecting them from the list. On the IVDR side the same structure is set up in Article 30, with reference to point 10.3 of Annex I of the IVDR.

  • Source of the list: In the list that appears in the field, some CMR and/or ED substances obtained from the website of the European Chemicals Agency (ECHA) are registered as options.

  • Missing substance: If the substance that needs to be selected is not in the list, a request can be made for the substance to be added by selecting the "CMR / ED Substance Addition" request field on the "Submit Request" screen in ÜTS. If the request is considered appropriate, the substance is added as an option.

  • Instructions and label: Information on the selected substances and the precautions to be taken must be in the instructions for use. If the instructions for use contain more detailed information, this information may be kept to a minimum on the label, taking the target users into account.

Affixing the Importer Label: The Manufacturer's Label Cannot Be Covered

How importer information is added to the product for imported devices is regulated not in the Guide but in the Medical Device Regulation. According to paragraph 3 of Article 13 of the Regulation, importers indicate their name, registered trade name or registered trade mark and the registered place of business at which they can be contacted, on the device or on its packaging or in a document accompanying the device.

The same paragraph also sets the limit: importers ensure that an additional label does not obscure any information on the label provided by the manufacturer. Affixing an importer label so that it covers the barcode or the lot field is contrary to this provision.

  • Operations not considered a modification: According to subparagraph (a) of paragraph 2 of Article 16 of the Regulation, in relation to a device already placed on the market, providing the information supplied by the manufacturer in accordance with point 23 of Annex I, including its translation, and other information necessary to market the device, is not considered a modification that could affect conformity.

  • Quality management system requirement: According to paragraph 4 of the same article, distributors and importers guarantee that they have a quality management system containing procedures ensuring that the translation of the information is accurate and up to date, that these activities are carried out so as to preserve the original condition of the device, and that the packaging of the repackaged device is not defective, of poor quality or untidy.

  • Information obligation: According to paragraph 3, the distributor or importer carrying out these activities indicates, on the device, its packaging or a document accompanying it, together with the activity carried out, its name, registered trade name or trade mark and the address of its place of business.

Turkish Requirement for the Label and Instructions for Use

According to paragraph 11 of Article 10 of the Medical Device Regulation, manufacturers ensure that the information specified in point 23 of Annex I accompanies the device in accordance with subparagraph (ğ) of paragraph 1 of Article 7 of Law No. 7223. The same paragraph also brings a form requirement: the details on the label must be indelible, easily legible and clearly comprehensible to the intended user or patient.

There is a similar language requirement on the document side. According to paragraph 1 of Article 19 of the Regulation, the EU declaration of conformity contains at minimum the information in Annex IV, and if the declaration is drawn up in other languages for devices placed on the market in Türkiye, its Turkish translation is attached.

Implant Card Information

For implantable devices a third information carrier is added alongside the label and the instructions. According to Article 18 of the Medical Device Regulation, the manufacturer of an implantable device provides with the device the name, address and website of the manufacturer; information allowing identification of the device, including the device name, serial number, lot number, UDI and device model; warnings and precautions to be taken with respect to foreseeable external influences and medical examinations; information on the expected lifetime of the device and the necessary follow-up; and other information to ensure safe use of the device by the patient.

According to paragraph 2 of the same article, this information is presented in Turkish and, where necessary, in English, in a way that allows quick access for the patient in whom the device is implanted, and is written so as to be easily understood by a lay person. Manufacturers also provide the identification information on an implant card supplied with the device. Paragraph 4 exempts sutures, staples, dental fillings, dental braces, tooth crowns, screws, wedges, plates, wires, pins, clips and connectors from these obligations.

Image and Label Preparation Checklist

The list below makes the rules on this page verifiable one by one before the application.

  • Number and format of images: Has the limit of 10 been exceeded? Are the files jpeg, jpg, pdf or png? Are the sizes under the ÜTS limit?

  • Clarity: Can the text on the label, the barcode and the date fields be read in the image?

  • Packaging and device image: Was only the packaging uploaded? Was a photograph or technical drawing of the device itself added? Does the exception reason (sterility or liquid form) really exist?

  • Image content: Has an image of another device, a catalogue page or an image containing more than one device got mixed in?

  • File names: Do the names describe the device and the image? Have commercial extensions and expressions such as "WhatsApp" been cleaned up?

  • Sağlık Market standards: If the field is "Yes", are the requirements of at least 3 photographs for each device and each package type, component images, 600x600 pixels, square and at most 2 MB met?

  • Label language question: Was "Is the Original Label in Turkish?" answered correctly? If "No", is it known that ÜTS will generate the Turkish label?

  • Label content: Is the information in the MDR, IVDR or old regulation list, according to the framework of the device, on the label and/or sales package? For information that does not fit, has the inner/outer package been uploaded?

  • Label–record consistency: Was the barcode entered as written on the label? Are the shelf life, storage condition, sterility, latex and phthalate information consistent between the label, the instructions and the record?

  • Importer label: Does the additional label cover no information on the manufacturer's label?

All of these checks are carried out through the company record in ÜTS. Companies that do not want to run the process with their own team can open the "Medical Device Registration Officer" authorisation for a consultant from the company page in ÜTS and have the image and label blocks prepared with this authorisation. The e-signature stays with the company's signatory; the company is responsible for the accuracy of the data.

If you also place your device on the EU market, you can look at our English guide on medical device symbols.

ÜTS Registration

Medical Device ÜTS Registration Consultancy

Request an ÜTS registration consultancy quote

Ehlibeyt Mahallesi, Tekstilciler Caddesi No: 35/7 06520 Çankaya / Ankara - TÜRKİYE

Cevizli Mahallesi, Zuhal Caddesi Ritim İstanbul B Blok, No:44 Maltepe / İstanbul - TÜRKİYE

info@medexmedikal.com
+90 505 580 4622
+90 312 389 1354

ÜTS Guides · Legislation Calendar · Glossary · ÜTS Guide Index · Frequently Asked Questions

Services: ÜTS Consultancy · Who Carries Out the ÜTS Registration? · ÜTS Document Registration Consultancy · ÜTS Product Registration Consultancy · Importer ÜTS Registration Consultancy · Sales Centre Authorisation Certificate · EUDAMED Registration Consultancy · MDR/IVDR Declaration of Conformity · Technical File Preliminary Review · Label and Instructions for Use Check · PRRC Service · PMS and PSUR · Clinical Evaluation · EMDN and GMDN Code Determination · Certificate of Free Sale · EUDAMED XML Generator · ÜTS Registration Training · Correction of a Rejected ÜTS Application

KVKK and Privacy Policy | Cookie Policy | Accessibility Statement

© 2026 Medex Kurumsal Danışmanlık Ltd. Şti.

bottom of page
WhatsApp Hemen Ara