ÜTS Document Types
In ÜTS the document is registered before the product, and the first thing you choose in the document record is the document type. If you pick the wrong type, the application type, the required attachments and the approval route all change
ÜTS Document Types: 11 Document Types and the Obligation Table
In ÜTS, the document is registered before the product, and the first thing you select in the document registration is the document type. If the wrong type is selected, the application type, the required attachments and the approval route all change. Eleven document types are defined between Articles 6 and 24 of the Manual on Registration Procedures of Medical Devices in the Product Tracking System (Tıbbi Cihazların Ürün Takip Sistemine Kayıt Süreçlerine İlişkin Kılavuz). Below we first give all the official data in two appendix tables, and then go through each type.
[APPENDIX: F] Document Type × Regulation Obligation Matrix
This table lists, one by one, the eleven document types defined in Articles 6 to 24 of the Manual. Each row shows what the document is, the framework in which it is used and whether it is mandatory or optional, what is uploaded to ÜTS, and under which article's procedures and principles its notification is carried out. The obligation is applied according to the document business rules set out by framework and risk class in Articles 63 to 67 of the Manual.
• EC Certificate: Used only for certificates prepared by notified bodies under the old regulations (MDD, AIMDD, IVDD), and mandatory for routes that involve a notified body. It is uploaded completely, with all its annexes and additional pages. The notification procedure is in Article 8 for importers and Article 9 for domestic manufacturers; the application is created under Article 39.
• EU Certificate (MDR/IVDR): Used only for certificates prepared by a notified body under the new regulations (MDR, IVDR), and mandatory for classes that require notified body involvement; it is also required in records of sterile SİP and Class III implantable custom-made devices. It is uploaded completely with its annexes. The procedures and principles are in Articles 8 and 9.
• Declaration of Conformity: Used only for declarations under the old regulations; it is issued by the legal manufacturer firm and is mandatory in every device record under the old regulations. It is uploaded completely with its annexes; if deemed necessary, additional documents may be requested in the "Technical Document Belonging to the Product(s) in the Document" field. The procedures and principles are in Articles 12 and 13.
• EU Declaration of Conformity (MDR/IVDR): Used only for declarations under the new regulations; it is issued by the legal manufacturer firm and is mandatory in every device record under the MDR and IVDR. It is uploaded with its annexes; for manufacturers not established in EU member states or in Turkey, an authorisation document or a EUDAMED output is also added to the "Authorisation Document" field. The procedures and principles are in Articles 12 and 13.
• SİP Declaration (MDR) (Article-22): The declaration of the SİP manufacturer placing a system/procedure pack under the MDR on the market, mandatory in SİP records under the MDR; if the pack is sterilised, the EU Certificate is also selected alongside it. It is uploaded with its annexes, together with its sworn translation if the original is not in Turkish. The procedures and principles are in Article 15 for imported declarations and Article 16 for domestic declarations.
• Instructions for Use: The manual issued by the manufacturer; whether it is mandatory varies by risk class, and the details are in the table in Appendix G. It is uploaded completely to the original document field and, if required, the Turkish document field. The notification procedure is in Article 18.
• Intended Purpose Form/Declaration: Prepared only for Class IIa devices under the MDD for which the manufacturer has not issued instructions for use, and replaces the manual. It must conform to the format in Annex-15; it is uploaded unchanged, wet-signed or e-signed, by selecting the document type "Instructions for Use" in ÜTS. The application procedure is in Article 19.
• Quality Management Certificate – ISO 13485: The certificate issued in accordance with the ISO 13485 standard; under Article 21 of the Manual it is not mandatory to upload. It is uploaded with its annexes; for importer notifications, by one of the routes of apostille, embassy/consulate certification, e-apostille or e-signed original. The procedures and principles are in Article 21.
• Domestic Product Certificate: Issued by a chamber or exchange affiliated with TOBB or TESK under the Domestic Goods Communiqué published pursuant to Article 63 of Law No. 4734. It is not mandatory; even if it is rejected, the linked devices remain registered. Since the original language is Turkish, only the original or a copy certified by a Turkish notary is uploaded, with its annexes, to the "Original Document" field. The procedures and principles are in Article 22.
• Authorised Distributorship Certificate: A document issued by the manufacturer on behalf of importer firms, declaring that the importer is authorised in Turkey; under Article 23 of the Manual it is not mandatory. It is uploaded with its annexes by one of the routes of apostille, embassy/consulate certification, e-apostille or the manufacturer's e-signed original; if it is in a foreign language, its sworn translation is added. The procedures and principles are in Article 23.
• Catalogue (Product Catalogue): A Turkish catalogue containing images introducing the device, brief information and reference/catalogue/version/model information. It is optional; it is mandatory for devices for which a SUT matching application will be made. Under Article 24 of the Manual, uploaded catalogues are saved without being subject to review expert control; no separate approval procedure is run.
[APPENDIX: G] Instructions for Use Obligation Table
Whether uploading the instructions for use is mandatory depends on the framework and risk class of the device. The table below contains all the rows in the second to sixth paragraphs of Article 17 of the Manual.
• MDR and MDD Class III and Class IIb: Uploading the instructions for use issued by the manufacturer is mandatory.
• MDR and MDD Class IIa: If the manufacturer has issued instructions for use, they are uploaded. As an exception, only for Class IIa devices under the MDD, if the manufacturer has not issued instructions for use, an intended purpose form/declaration signed by the responsible manager (Sorumlu Müdür) of the manufacturer or importer firm is uploaded.
• IVDR Class B, Class C and Class D: Instructions for use issued by the manufacturer are mandatory.
• IVDR Self-Test Devices (Class B/C/D), Near-Patient Test Devices (Class B/C/D) and Companion Diagnostic Devices (Class C/D): Instructions for use issued by the manufacturer are mandatory.
• IVDD Annex II List A, Annex II List B and Self-Test Devices: Instructions for use issued by the manufacturer are mandatory.
• Devices under the AIMDD: Instructions for use issued by the manufacturer are mandatory.
• Class I Other, Class Im, Class Is, Class Ir, Class A, Class A Sterile and IVD Other: Uploading instructions for use is optional. However, the Agency may request the instructions for use of these devices where it deems necessary (for example, where doubt arises about product safety, or where the original label refers to the instructions for use and some information needs to be verified from them).
• Class IIa devices that are accessories of another device: If the relevant accessory question in the device record is answered "Yes", there is no obligation to upload instructions for use.
• MDR and MDD Class IIb and Class III devices that are accessories of another device; IVDR Class B, C, D, Self-Test (B/C/D), Near-Patient Test (B/C/D) and Companion Diagnostic (C/D) devices; IVDD Annex II List A, List B and Self-Test Devices: If the manufacturer of the main device is different from the manufacturer of the accessory, the accessory's own instructions are uploaded if the manufacturer of the accessory has created instructions for use, and otherwise the instructions created by the manufacturer of the main device are uploaded, regardless of criteria such as differences in notified body and address.
• SİP under the MDR and MDD: If a separate instructions for use has been prepared by the SİP manufacturer, the SİP's own instructions are uploaded; otherwise all of the instructions for use of the devices within the SİP that fall under the medical device regulations are uploaded.
• SİP whose content consists entirely of the optional risk group: If all the devices in the SİP belong to the risk groups for which Article 17(3) of the Manual does not require uploading instructions for use, there is no obligation to upload instructions for use for that SİP.
EC Certificate: Scope and Fields Checked
According to Article 6(1) of the Manual, a document of the "EC Certificate" type is used only in notifications and applications concerning certificates prepared by notified bodies under the old regulations. For certificates issued under the new regulations, the notification is made under the "EU Certificate (MDR/IVDR)" type.
Paragraph 2 of the same article lists ten fields that must be on the document and that are checked during review in ÜTS: the name or trade name and address of the manufacturer, the name and address of the EU authorised representative if any, the certificate number, the name and identification number of the notified body, the address of the notified body, the regulation it belongs to, the annex under which it was issued, the document start and end dates, the product scope and the legally binding signature of the notified body. The information entered in ÜTS must be the same as that on the certificate.
• Validity period: According to paragraph 3, except for EC Verification and EC Type Verification certificates, EC certificates have a validity period, which cannot exceed 5 years; it may be extended on the basis of re-assessment by additional periods each not exceeding 5 years.
• Document start date and revision: If the document has been revised, the revision date is entered as the start date; if not, the most current signature date. If there is a revision number shown with the certificate number, the document is recorded together with it.
• If the notified body is not defined in ÜTS: Before document registration, a request is opened through the "Notified/Conformity Assessment Body Procedures" field on the "Report Request" screen; if it is established in the NANDO query that the body has been designated, the body is added to ÜTS.
EU Certificate (MDR/IVDR): Fields Checked
According to Article 7(1) of the Manual, this type is used only for certificates prepared by a notified body under the new regulations. Paragraph 2 keeps the fields checked broader than for the EC certificate.
• Manufacturer and representative: The name or trade name and address of the manufacturer and, where applicable, the name and address of the EU authorised representative. Under Article 11 of the MDR and Article 12 of the IVDR, manufacturers established outside the EU member states and Turkey are required to appoint an authorised representative.
• SRN: If it has already been issued, the manufacturer's Single Registration Number (SRN) is entered in the "Single Identification No" ("Münferit Kimlik No") field in ÜTS.
• Notified body, regulation and dates: The name, identification number and address of the notified body; the regulation it belongs to and the annex under which it was issued; the issue and validity dates of the document and the product scope.
• Amendments and additions: If the certificate has been amended, reissued or supplemented, a reference to each previous certificate and its issue dates, together with a description of the changes.
• Examinations, surveillance and results: The examinations and tests carried out; information on notified body surveillance; the results of the conformity assessment; conditions or limitations concerning the validity of the certificate and the binding signature of the notified body.
According to paragraph 3, EU certificates remain valid for the period stated on the certificate, not exceeding 5 years, as long as no amendment or update is made within that period.
Content That Varies by Certificate Type
Point (h) of Article 7(2) of the Manual splits the information sought on the product scope in two according to the certificate type. This split also determines on which certificates the Basic UDI-DI is sought.
• Technical documentation, type-examination and product verification certificates: A clear identification of the device, including name, model and type; the intended purpose as included by the manufacturer in the instructions for use and relevant to the conformity assessment procedure; the risk class and the Basic UDI-DI are sought.
• Quality management and production quality assurance certificates: The identification of the devices or device groups, the risk class and, for Class IIb devices, the intended purpose are sought. Under Article 28 of the Manual, the Basic UDI-DI information is not mandatory for these two certificate types.
Declaration of Conformity: Content and Issuer
According to Article 10(1) of the Manual, the "Declaration of Conformity" type is used only for declarations prepared under the old regulations; paragraph 3 states that the declaration is issued by the legal manufacturer firm. According to paragraph 2, the items that must be on the document and are checked in ÜTS are the following:
• Manufacturer and representative: The name or trade name and address of the manufacturer and, if any, the name and address of the EU authorised representative.
• Device information: Model, reference number, product name and, if any, a detailed description.
• Regulation, class and annex: The regulation to which the declaration belongs, and the class of the devices and/or the annex under which it was carried out.
• Document start date: The issue date of the declaration.
• Document end date: If the document has an unlimited period, "Unlimited Period" ("Sınırsız Süreli") is selected in ÜTS; if there is an end date, it is entered in the relevant field.
To this are added the name, surname, position and signature/stamp of the authorised person; the e-signed version is also valid. Paragraph 5 requires the product scope to be capable of identifying the product; general expressions such as "surgical hand instruments" are not sufficient, and detail such as forceps or retractors is expected.
EU Declaration of Conformity (MDR/IVDR): Fields and Templates
Article 11 of the Manual regulates the declaration under the new regulations; the fields checked expand markedly compared with a declaration under the old regulations.
• Statement of responsibility and Basic UDI-DI: The declaration contains a statement that the manufacturer draws up the EU declaration of conformity under its sole responsibility, and the Basic UDI-DI.
• Identification and traceability: Precise references such as product name and trade name, product code, catalogue number or a photograph, and the intended purpose, in a form that allows identification and traceability of the device. Except for the product name or trade name, this information can also be provided through the Basic UDI-DI.
• SRN, risk class and regulatory compliance: If it has been issued, the manufacturer's Single Registration Number, the risk class of the device, a statement that the device conforms to the regulation it falls under, and references to any common specifications used.
• Date and signature: The document start and end dates, the place and date of issue of the declaration, and the name, surname, position and signature/stamp of the authorised person.
• Templates: Under paragraph 7, EU Declaration of Conformity templates created especially as examples for domestic manufacturer firms are in Annex-7 and Annex-8 of the Manual.
The Class I Other and Class A Exception
Point (ı) of Article 11(2) of the Manual makes the notified body information conditional. In declarations other than EU declarations of conformity whose scope contains only Class I Other or Class A devices, the name and identification number of the notified body, a description of the conformity assessment procedure carried out, and the identification of the EU certificate or certificates issued are required.
In a declaration whose scope consists only of Class I Other or Class A devices, notified body information is not required; since conformity assessment in these classes does not involve a notified body, the declaration does not carry this information either.
SİP Declaration (MDR) (Article-22)
According to Article 14 of the Manual, this document is the declaration prepared by the SİP manufacturer when it combines devices bearing the CE marking under the MDR with other CE-marked devices, CE-marked in vitro diagnostic devices conforming to the IVDR and, where justified, other products, in order to place them on the market as a SİP.
The declaration contains three commitments: that the mutual compatibility of the devices has been verified in accordance with the manufacturers' instructions, that the pack has been packaged and the relevant information, including that supplied by the manufacturers, has been provided to users, and that the combining activity has been subjected to appropriate methods of internal monitoring, verification and validation.
• Imported SİP declaration: Under Article 15 of the Manual, the apostille-certified copy, the embassy/consulate-certified copy, the verifiable e-apostille-certified original or the e-signed original of the document of the SİP manufacturer is uploaded to the "Original Document" field with its annexes; if it is in a foreign language, its sworn translation is added to the "Turkish Document" field.
• Domestic SİP declaration: Under Article 16 of the Manual, "e-Signed Document Application for Manufacturer Firms" is selected as the Document Application Type, and the original of the document is uploaded e-signed by the person registered on behalf of the firm as "Responsible Manager" ("Mesul Müdür") or "Signatory" ("İmza Yetkilisi").
Instructions for Use: Content and Common Information
Article 17(8) of the Manual lists the common information sought in the instructions for use when assessing a device registration application. These apply together with the additional information in the "Annex-I Essential Requirements" section of the relevant regulation.
• Device and manufacturer: The name or trade name of the device; the name, registered trade name or trademark and registered place of business address of the manufacturer; and, where applicable, the name and registered place of business address of the EU authorised representative.
• Sterility: If the device is supplied sterile, an indication of its sterile condition and of the sterilisation method.
• Single use: If the device is single use, the indication "Single Use".
• Sterilisation before use: If the device must be sterilised before use, the sterilisation method and details.
• Reuse: If the device can be reused, information on the processes that allow its reuse, including cleaning, disinfection and packaging, and on the validated method of re-sterilisation.
• Storage and use conditions: Any specific storage and/or use conditions, and other information required by the regulation it belongs to.
According to paragraph 10, information such as single use, sterility and the sterilisation method can be shown with internationally recognised symbols; every symbol or identification colour used must conform to harmonised standards or common specifications. Paragraph 7 also requires the instructions to contain the name and address of the manufacturer, the product scope information, the intended purpose, the print date and, if any, the latest update information.
The Four-Digit Notified Body Number Beside the CE Marking
According to point (d) of Article 17(7) of the Manual, the instructions for use contain the CE marking together with the four-digit notified body number of the product. This rule has three extensions.
• Products for which the CE marking alone is sufficient: For Class I Other, Class A and IVD Other products, the CE marking alone is sufficient.
• SİP-specific instructions: The CE marking is not sought in instructions created specifically for the SİP.
• Products subject to more than one regulation: In the instructions and label of products that are subject to conformity assessment under other main regulations in addition to the medical device regulations (such as the Personal Protective Equipment or Machinery Safety regulations), more than one CE marking and notified body information may appear.
When the Print Date Does Not Specify a Day
In instructions for use, the print or update date is often written only as month and year. Article 17(9) of the Manual provides a single solution: if the date does not specify a day, "1" is entered as the day in ÜTS. If the document print and/or update date is stated as 05/2020, the date is entered in ÜTS as 01/05/2020.
It is a small detail, but it affects the outcome of the application; if a different day is selected, a mismatch arises between the date in ÜTS and the date in the document. Article 18(3) of the Manual also makes it mandatory, if the instructions have been updated, to enter at least one of the update date and/or the update information in ÜTS.
Electronic Instructions for Use (eIFU)
According to Article 17(11) of the Manual, the instructions for use may also be provided to the user in a format other than paper, for example electronically. For instructions provided in this way, too, a document registration or update notification is made in accordance with the provisions stated in Article 18 of the Manual.
An electronic manual does not remove the notification obligation. The route in Article 18 applies exactly: if the original document was not prepared in Turkish, the most current original manual in the technical file is uploaded to the "Original Document" field and its Turkish translation to the "Turkish Document" field. If the original document was prepared in multiple languages including Turkish, all pages of the multilingual manual are uploaded to a single field.
Intended Purpose Form/Declaration: Only MDD Class IIa
According to Article 19(1) of the Manual, for Class IIa devices under the MDD for which the manufacturer has not issued instructions for use, an intended purpose form/declaration is prepared by the manufacturer or importer in the format whose example is in Annex-15. This document is the only exception that takes the place of the instructions for use and is not used in any other framework or class.
• Choice of document type: When uploading to ÜTS, "Instructions for Use" is selected as the "Document type".
• Questions: A new document is added by answering the question "Is the original document in Turkish?" with "Yes", and the document is uploaded completely by answering the question "Is it an Intended Purpose Form/Declaration?" with "Yes".
• No alteration, and signature: The form is uploaded unaltered, wet-signed or e-signed; if it is e-signed, the original of the document must be uploaded so that verification can be made.
• Physical packaging and accuracy: Under paragraph 3 the form is for registration purposes only and does not physically appear on the device packaging; under paragraph 4 all the information in it must be correct and valid.
Quality Management Certificate – ISO 13485
According to Article 20 of the Manual, a document recorded in ÜTS under the name "Quality Management Certificate – (ISO-13485)" contains, at a minimum, the name and address of the manufacturer, the document number, the name and contact details of the certifying body, the document start and end dates and the product scope information.
• Product scope and language: If the original document is in a foreign language, the scope in the original document is entered in the "Product Scope in the Original Language" field and a consistent Turkish scope in the "Product Scope in Turkish" field.
• Start date and revision: If it has been revised, the revision date is entered as the start date; if the revision number is next to the certificate number, the document number is recorded with it.
• Validity period: Under paragraph 5, the validity period of such documents cannot exceed 3 years.
• Being optional: Under Article 21(9) of the Manual, this document is not mandatory; even if its registration is rejected, the linked devices remain registered.
Article 21 of the Manual also sets the certification route. In importer notifications, the apostille-certified copy of the document, the copy certified by a Republic of Türkiye embassy/consulate in the manufacturer's country, the verifiable e-apostille-certified original or the original of the certifying body's e-signed certificate is uploaded to the "Original Document" field. The same routes apply to documents that domestic manufacturers obtain from a body abroad; for documents obtained from a certification body established in Turkey, the original of the document or a copy certified by a Turkish notary is accepted.
Domestic Product Certificate
According to Article 22(1) of the Manual, the Domestic Product Certificate is issued, under Article 10 of the Domestic Goods Communiqué published in Official Gazette No. 32793 of 25/1/2025 pursuant to Article 63 of Public Procurement Law No. 4734, by the chamber or exchange affiliated with TOBB or TESK with which the manufacturer is registered.
According to paragraph 3, a standard form is used for the document. The form contains the date of issue and validity and the number of the document, the title and contact details of the manufacturer, registry numbers, capacity report information, product name and type, PRODTR codes, GTİP numbers, the domestic contribution rate, the document class (A, B, C, D, E), the technology level of the product, the product image together with label and technical information, and the seal and signature of the issuing chamber or exchange.
• Checked in the assessment: Under paragraph 4, the name and address of the manufacturer, the document number, the domestic contribution rate, the product name, the brand name, the document start and end dates, and the chamber's seal and signature are checked.
• Validity period: Under paragraph 5, the validity period is one year from the date of issue.
• Upload: Under paragraph 6, since the original document language is Turkish, the original of the document obtained from the chamber, or a copy certified by a Turkish notary, is uploaded with its annexes, if any, only to the "Original Document" field.
• Rejection: Under paragraph 9, even if the document registration is rejected, the linked devices remain registered.
When the Domestic Product Certificate Has No Brand Name
Article 22(4) of the Manual directly addresses a gap between the pieces of legislation. Although the brand name is a field that must be entered in ÜTS under the medical device regulations, under the current Domestic Goods Communiqué it is not required to appear on the domestic product certificate. Therefore, if the document has no brand name, it is mandatory to enter the brand name in the "Brand Name" ("Marka Adı") field on the Domestic Product Certificate screen in ÜTS, in accordance with the brand name provisions of Articles 25, 27 or 30 of the Manual. The brand name's absence from the document does not mean that the field may be left empty.
Authorised Distributorship Certificate
According to the definitions article of the Manual and Article 23(1), the Authorised Distributorship Certificate is a document issued by the manufacturer of the device on behalf of importer firms operating in our country, declaring that the importer concerned is authorised in Turkey for the devices covered by the document. According to paragraph 2, the four items that must be in the document and are checked during review are the following:
• Manufacturer information: The name or trade name and address of the manufacturer.
• Authorised firm: The name or trade name and address of the authorised firm.
• Product scope: The products covered by the document.
• Date and signature: The authorisation start date and, if any, end date, and the signature and stamp of the manufacturer's authorised person.
If the document has no authorisation end date, only the "Document Start Date" is entered in the system, using the "Is it unlimited?" ("Sınırsız süreli mi?") option. According to paragraph 9, this document is also optional; even if its registration is rejected, the linked devices remain registered.
Product Scope "All Products" / "Tüm Ürünler"
Article 23(3) of the Manual gives a ready solution for documents whose scope is not written in detail. If the original document is in a foreign language, the option "Is the original document in Turkish?" is selected as "No"; if the product scope is not explained in detail or the document covers all of the manufacturer's products, "All Products" is written in the "Product Scope in the Original Language" field and "Tüm Ürünler" in the "Product Scope in Turkish" field.
If the original document was prepared in Turkish or in multiple languages including Turkish, the same option is selected as "Yes" and the product information is written in Turkish; if the scope is not detailed, "Tüm Ürünler" is written in this field as well.
Catalogue (Product Catalogue)
According to Article 24(1) of the Manual, a Turkish catalogue containing images introducing the device, brief information and information such as reference, catalogue, version or model can be uploaded. This type goes through a different procedure.
• Review exemption: Under paragraph 2, uploaded catalogues are saved in ÜTS without being subject to review expert control.
• Obligation for SUT matching: Under Article 25 of the Manual, uploading a Turkish catalogue is mandatory for devices for which a SUT matching application will be made.
• Number: In the annexes to the document business rules, the catalogue is the only optional document that can be more than one.
If a Document Is Rejected, Does the Device Record Fall?
The effect of the rejection of a document registration on the device record varies by type; the Manual answers this explicitly for three types.
• Quality Management Certificate – ISO 13485: Under Article 21(9) of the Manual, even if the document registration is rejected, the linked devices remain registered.
• Domestic Product Certificate: The same provision is in Article 22(9) of the Manual.
• Authorised Distributorship Certificate: The same provision is in Article 23(9) of the Manual.
For the other document types the Manual contains no explicit provision in this direction. The logic in practice is this: if there is no record of a document to which the device record is bound under the document business rules (such as a declaration of conformity, a certificate or mandatory instructions for use), the link required for the device record cannot be established either. Therefore, for types other than the three optional documents, the status of the document directly concerns the device record.
The Difference Between the Importer and Domestic Manufacturer Routes
The same document type goes through a different procedure depending on who notifies it; Articles 8, 9, 12, 13, 15 and 16 of the Manual establish this distinction.
• Importer route: For documents belonging to a foreign manufacturer, the apostille-certified document, the embassy/consulate-certified document, the verifiable e-apostille-certified original or the original of the e-signed document is uploaded to the "Original Document" field. For certificates, a copy certified by a Turkish notary is also accepted if the document is not e-signed, provided that it can be verified with all its pages on the notified body's official query site or in EUDAMED.
• Sworn translation: On the importer route, if the original document is in a foreign language, a sworn translation is uploaded to the "Turkish Document" field; the document must contain the translator's name and surname, stamp/signature/seal and the declaration of faithfulness to the original. If the original document is in Turkish or in multiple languages including Turkish, the upload is made only to the "Original Document" field.
• Domestic manufacturer route: "e-Signed Document Application for Manufacturer Firms" is selected as the Document Application Type and the question "Are you the manufacturer of the products covered by the document?" is marked "Yes". For declarations, the original of the document is uploaded e-signed by the person registered as "Responsible Manager" ("Mesul Müdür") or "Signatory" ("İmza Yetkilisi").
• Annex-21 undertaking route: Under Article 9 of the Manual, in the EC/EU certificate notifications of domestic manufacturers, a copy of the certificate concerned is uploaded completely to the "Original Document" field, together with the e-signed original of the undertaking in Annex-21.
• Physical submission: All of these articles state that the uploaded documents do not need to be sent physically to the Agency; the application through ÜTS is sufficient.
Optional Documents and Quantity Limits
Articles 63 to 67 of the Manual tie which documents are to be associated with a device record, in five frameworks, to the document business rules, and show them by reference to the tables between Annex-81 and Annex-100. The "Optional Documents" column of these tables gives the documents that can be uploaded alongside the mandatory documents and the quantity limits.
• Quality Management Certificate ISO 13485: At most 1.
• Authorised Distributorship Certificate: At most 1.
• Domestic Product Certificate: At most 1.
• Technical File: At most 1.
• Catalogue: There can be more than one.
The noteworthy point is this: the text of the Manual has no separate document article headed "Technical File"; this item appears only in the optional document lists of the annexes. This is why we built the catalogue of document types on eleven headings.
Document Type Selection Guide
To avoid mistakes on the document side, the sequence needs to be set up from the start. Reading Articles 6, 7, 8 and 39 of the Manual together yields a four-step sequence.
• Step 1 - Determine the framework: Is your device under the MDR/IVDR or under the MDD/AIMDD/IVDD? This determines which type you will select on the certificate and declaration side.
• Step 2 - Select the document type: Under the old framework the EC Certificate and the Declaration of Conformity are used, and under the new framework the EU Certificate (MDR/IVDR) and the EU Declaration of Conformity (MDR/IVDR). Instructions for use, ISO 13485, Domestic Product, Authorised Distributorship and Catalogue are used under the same name in both frameworks.
• Step 3 - Select the application type: The types in Article 39(4) of the Manual are "Document Application for Importer Firms", "Document Application for Other Types for Importer or Manufacturer Firms", "e-Signed Document Application for Manufacturer Firms" and "Document Application under Announcement No. 2023/KK-1". Certificates and declarations go through the first type, instructions for use and optional documents through the second, and domestic manufacturer notifications through the third.
• Step 4 - Select the approval route: On the importer route, the apostille, embassy/consulate certification, e-apostille or e-signed original; on the domestic manufacturer route, the e-signed application and, if required, the Annex-21 undertaking.
According to Article 39 of the Manual, a document application contains at most 5 document notifications, and only the firm's users of the "Medical Device Registration Operator" type can perform the "Approve" action. This structure is the basis of the consultancy model: the firm opens this authority to its consultant from its own page in ÜTS, the records are created with this authority, the firm's signatory signs the application with the e-signature, and the accuracy of the data remains the firm's responsibility.
If you also offer your device to the EU market, take a look at our English guide on the MDR technical file.