ÜTS Registration by Device Class

The "Class" field you meet when registering your device in ÜTS looks like a single drop-down list. Yet the choice you make there determines which documents you can link and whether you must upload instructions for use

Device Class and ÜTS Registration: How Class Selection Affects the System

The "Class" field you meet when registering your device in ÜTS looks like nothing more than a single drop-down list. Yet the choice you make there determines which documents you can link, whether you must upload instructions for use, which additional fields open on the screen and with which application type you submit your application. A wrong class is not a single wrong field; it runs the entire application on the wrong track.

Based on the TİTCK Guide, this page explains the classes that can be selected in ÜTS under three frameworks (the former regulations, MDR and IVDR), the concrete effect of each class on the registration, and how a class change works in the system.

The Role of the Class Field in ÜTS

The class field works as a sequencing lock on the ÜTS registration screen. Article 25, paragraph 5, subparagraph (a), sub-item 3 of the Guide states that the device class must be selected before any document is chosen in the "Documents" field, and that documents are selected from those available for the device's class. The same provision is repeated word for word in paragraph 5(a) of Article 27 for MDR and Article 30 for IVDR. In other words, you cannot reach the document list without selecting a class; if you select the wrong class, you get the wrong set of documents.

How documents are linked to a device depends on logic called "document business rules". Article 63, paragraph 2 of the Guide states that linking under the MDD is carried out in ÜTS according to the document business rules determined by the device type and risk class and by the conformity assessment procedures in Article 58. The same structure is repeated in paragraph 2 of Article 64 for AIMDD, Article 65 for IVDD, Article 66 for MDR and Article 67 for IVDR. As a result, the class directly determines the document combination with which your device can be registered.

Classes Selectable in ÜTS under the Former Regulations

In registrations under the former Medical Device Regulations (MDD, AIMDD and IVDD), the class field combines medical device and in vitro diagnostic options in a single list. According to Article 25, paragraph 4, subparagraph (a), sub-item 3 of the Guide, this list has ten options, and the one that fits the device is selected.

  • Class I Other (non-sterile, no measuring function): MDD Class I devices that are not supplied sterile and have no measuring function.

  • Class Im (with measuring function): Class I devices with a measuring function.

  • Class Is (with sterile function): Class I devices supplied in a sterile condition.

  • Class IIa: the MDD medium-low risk group.

  • Class IIb: the MDD medium-high risk group.

  • Class III: the MDD highest risk group and all devices under AIMDD.

  • Annex II List A: high-risk IVD devices on List A of IVDD Annex II.

  • Annex II List B: IVD devices on List B of IVDD Annex II.

  • Self-Testing Devices: devices intended for self-testing under IVDD.

  • IVD Other: other in vitro diagnostic devices under IVDD that do not fall into these three groups.

Article 55, paragraph 3 of the Guide states expressly that devices under IVDD are classified in ÜTS as "Annex II List A", "Annex II List B", "Self-Testing Devices" and "IVD Other"; paragraph 2 of the same article lists the contents of Lists A and B item by item.

AIMDD Devices Are Registered in ÜTS as Class III

There is no need to look for a separate class option for active implantable devices. Article 25, paragraph 4, subparagraph (a), sub-item 4 of the Guide states directly that "Class III" is selected in the relevant field for devices under AIMDD.

The reason is set out in Article 54. Paragraph 1 of that article states that there is no classification rule specific to devices under AIMDD and that, taking into account the conformity assessment procedures referred to in Article 59 of the Guide, such devices are treated as belonging to the class equivalent to the highest risk class under the MDD (Class III). Article 64, paragraph 3 of the Guide confirms the same result for the document business rules.

The application side, however, differs. In the list of application types in Article 40, paragraph 3 of the Guide, "Active Implantable Medical Devices Application" is a separate option. So you select Class III in the class field, but you submit the application as an active implantable application, not as a Class 3 Products Application.

ÜTS Class Options under MDR

The MDR list is shorter than the one under the former regulations because the IVD side does not appear on this screen. According to Article 27, paragraph 4, subparagraph (a), sub-item 3 of the Guide, there are seven options under MDR, and the one that fits the device is selected.

  • Class I Other (non-sterile, no measuring function): the basic Class I group under MDR.

  • Class Ir (reusable surgical instruments): a sub-class that exists only on the MDR screen.

  • Class Im (with measuring function): Class I devices with a measuring function.

  • Class Is (with sterile function): Class I devices supplied sterile.

  • Class IIa: the MDR medium-low risk group.

  • Class IIb: the MDR medium-high risk group.

  • Class III: the MDR highest risk group.

Difference between Class I Other, Class Im, Class Is and Class Ir

These four are separate options in ÜTS; under MDR itself, all of them are Class I. Article 56, paragraph 1 of the Guide states this expressly: devices under MDR are divided into four classes, Class I, Class IIa, Class IIb and Class III, while ÜTS adds further classifications according to characteristic product features such as sterile packaging or measuring function. The example in that paragraph is clear: a device that is "Class Ir" in ÜTS and a device that is "Class I Other" in ÜTS are both Class I under MDR.

Class Ir is a category with no equivalent in the former regulations. Rule 6, found in Article 56, paragraph 4, subparagraph (b) of the Guide, states that all surgically invasive devices intended for transient use are classified as Class IIa, but that if they are reusable surgical instruments they are classified as Class I (Class Ir). If your device is a reusable surgical instrument and was registered as Class I Other in the MDD period, this field must be Class Ir in the MDR registration. The practical result of the distinction appears in Article 61, paragraph 4, subparagraph (c) of the Guide: where Class Ir is concerned, the notified body becomes involved, but its involvement is limited to aspects relating to the reuse of the device, in particular cleaning, disinfection, sterilisation, maintenance, functional testing and the related instructions for use.

Thirteen Class Options under IVDR

The IVDR screen has the longest class list in ÜTS. According to Article 30, paragraph 4, subparagraph (a), sub-item 3 of the Guide, the list has thirteen options.

  • Class A Sterile: Class A devices supplied sterile.

  • Class A: the lowest IVDR risk group.

  • Class B: the basic default IVDR group.

  • Class C: the IVDR medium-high risk group.

  • Class D: the highest IVDR risk group.

  • Self-Testing Device (Class B): devices intended for self-testing that fall into Class B.

  • Self-Testing Device (Class C): devices intended for self-testing that fall into Class C.

  • Self-Testing Device (Class D): devices intended for self-testing that fall into Class D.

  • Near-Patient Testing Device (Class B): Class B devices intended for near-patient testing.

  • Near-Patient Testing Device (Class C): Class C devices intended for near-patient testing.

  • Near-Patient Testing Device (Class D): Class D devices intended for near-patient testing.

  • Companion Diagnostic Device (Class C): Class C devices intended as companion diagnostics.

  • Companion Diagnostic Device (Class D): Class D devices intended as companion diagnostics.

These variants do not create a new risk class under IVDR. Article 57, paragraph 1 of the Guide states that devices under IVDR are divided into four classes, Class A, Class B, Class C and Class D, and that ÜTS adds further classifications according to characteristic product features and intended purposes. The example given in the paragraph is clear: a device that is a "Near-Patient Testing Device (Class C)" in ÜTS is classified as Class C under IVDR.

Registration of Class I and Class A Devices: What Applies When There Is No Notified Body?

Because no notified body certificate is required in the lowest risk groups, the weight in registration shifts to the declaration of conformity and the technical documentation; the fate of the application depends on the content of the declaration and the accuracy of the label.

  • MDD Class I Other: According to Article 58, paragraph 2 of the Guide, the procedures in MDD Annex VII are followed, technical documentation is prepared as specified in the same annex, and a declaration of conformity is drawn up in accordance with Article 10.

  • MDD Class Is and Class Im: According to the same paragraph, one of the procedures in MDD Annex II, Annex IV, Annex V or Annex VI is also carried out; the involvement of the notified body is limited, under subparagraph (a), to the conditions of sterility for Class Is and, under subparagraph (b), to conformity with metrological requirements for Class Im.

  • MDR Class I Other: According to Article 61, paragraph 4 of the Guide, technical documentation is prepared in accordance with MDR Annex II and Annex III, followed by an EU declaration of conformity under MDR Annex IV and Article 11.

  • IVDR Class A: According to Article 62, paragraph 4 of the Guide, technical documentation is prepared in accordance with IVDR Annex II and Annex III and an EU declaration of conformity under Annex IV; for Class A Sterile, one of the procedures in IVDR Annex IX (Chapters I and III) or Annex XI is also carried out.

  • IVDD "IVD Other": According to Article 60, paragraph 2 of the Guide, the procedures in IVDD Annex III are carried out, excluding devices for performance evaluation and the devices in IVDD Annex II, and a declaration of conformity is drawn up.

There is also no obligation to upload instructions for use in these groups. Article 17, paragraph 3 of the Guide states that uploading the instructions is optional for Class I Other, Class Im, Class Is, Class Ir, Class A, Class A Sterile and IVD Other devices, although the Agency may request that they be uploaded where it considers this necessary.

Registration of Class IIa Devices and the Instructions-for-Use Requirement

Class IIa is the group where the instructions-for-use rule is most often confused. Article 17, paragraph 2, subparagraph (a), sub-item 2 of the Guide states that, for Class IIa devices certified under MDR and MDD, uploading the instructions for use is mandatory if the manufacturer has prepared them. In other words, if instructions exist, you must upload them; there is no rule such as "instructions are not requested for Class IIa".

The same sub-item provides an exception, and this exception belongs only to the former regulations side: only for Class IIa devices under the MDD, if the manufacturer has not prepared instructions for use, an intended purpose form/statement referred to in Article 19 of the Guide, signed by the manufacturer or by the responsible manager of the importer company, is uploaded. Paragraph 2 of the same article states that when this document is uploaded to ÜTS, "Instructions for Use" is selected as the document type and "Yes" is ticked for the questions "Is the original document in Turkish?" and "Is it an Intended Purpose Form/Statement?"; paragraph 3 states that the document is for registration purposes only and is not physically included in the device packaging.

The rule changes for Class IIa devices that are accessories. Article 17, paragraph 4 of the Guide states that, for Class IIa devices that are accessories of another device, no obligation to upload instructions for use applies if, during device registration, the field "Is it a Component / Accessory of Another Medical Device?" is answered "Yes" in applications under the former regulations, or the field "Is it a component/accessory of a device?" is answered "Yes" in MDR and IVDR applications.

Registration of Class IIb Devices: Three Sub-Groups

Under MDR, Class IIb is not a single whole; the system runs three separate document business rules. Paragraph 5 of Article 66 of the Guide points to Annex 90 for Class IIb devices other than implantable devices or devices covered by MDR Annex VIII Rule 12; paragraph 6 points to Annex 91 for implantable Class IIb devices other than sutures, staples, dental fillings, dental braces, tooth crowns, screws, wedges, plates, wires, pins, clips and connectors; and paragraph 7 points to Annex 92 for Class IIb devices covered by Rule 12.

The counterpart of this distinction in conformity assessment is in Article 61 of the Guide. Paragraph 6(a) states that, for Class IIb devices, the technical documentation of at least one representative device per generic device group is assessed under MDR Annex IX (Chapter II, Section 4), while (b) states that for implantable Class IIb devices other than the products listed above, this assessment applies to every device. Paragraph 10 of the same article adds that, for Class IIb active devices covered by Rule 12, the MDR Annex IX (Section 5.1) or MDR Annex X (Section 6) procedure also applies. Rule 12, in Article 56, paragraph 5, subparagraph (ç) of the Guide, states that all active devices intended to administer and/or remove medicinal products, body liquids or other substances to or from the body are Class IIa, unless this is done in a manner that is potentially hazardous, taking into account the nature of the substances, the part of the body concerned and the mode of application, in which case they are Class IIb.

For Class IIb, instructions for use are undeniably mandatory. Article 17, paragraph 2, subparagraph (a), sub-item 1 of the Guide states that for Class III and Class IIb devices certified under MDR and MDD, uploading the instructions for use prepared by the manufacturer is mandatory.

Registration of Class III Devices and a Separate Application Type

Class III is the only risk class in ÜTS with its own application type. Article 40, paragraph 3 of the Guide states that the list in the "Create Medical Device Application" field currently includes the following device application types: "Optical Products Application", "Active Implantable Medical Devices Application", "Class 3 Products Application", "Other Products Application", "Custom-Made Medical Device Application", "System Procedure Pack Application" and "Products under 2023/KK-1 & 2023/KK-2 Application". Submitting your Class III device with an "Other Products Application" means opening the application with the wrong type.

Conformity assessment also narrows for Class III. Article 58, paragraph 5 of the Guide states that manufacturers of Class III devices under the MDD are subject either to the declaration of conformity procedure relating to the full quality assurance system in MDD Annex II, or to the type examination in MDD Annex III together with the EC verification in MDD Annex IV or the production quality assurance in MDD Annex V. The MDD Annex VI (product quality assurance) route, which is an option for Class IIa and Class IIb, does not appear in this paragraph for Class III. Under MDR, assessment goes down to the individual device: Article 61, paragraph 7, subparagraph (a) of the Guide states that manufacturers of Class III devices are subject to a conformity assessment under MDR Annex IX (Chapters I and III) and, for every device, MDR Annex IX (Chapter II).

Registration in IVD Classes: A, A Sterile, B, C, D and Variants

On the IVD side, the framework you are in completely changes the class names. In a former-regulation registration you choose among Annex II List A, Annex II List B, Self-Testing Devices and IVD Other; in an IVDR registration, the list with thirteen options appears. It is normal to see different class names in the IVDD registration and the IVDR registration of the same device.

The instructions-for-use requirement is defined in both frameworks. Article 17, paragraph 2, subparagraph (b) of the Guide states that uploading the instructions for use is mandatory for Class B, Class C and Class D devices certified under IVDR, as well as Self-Testing Devices (Class B/C/D), Near-Patient Testing Devices (Class B/C/D) and Companion Diagnostic Devices (Class C/D). Subparagraph (c) of the same paragraph introduces the same requirement for Annex II List A and List B devices and Self-Testing Devices under IVDD. Class A, Class A Sterile and IVD Other are in the optional group under paragraph 3.

Self-Testing, Near-Patient Testing and Companion Diagnostic Variants

In an IVDR registration there is another question alongside the class. Article 30, paragraph 7, subparagraph (t) of the Guide states that the field "Is it a device intended for self-testing or near-patient testing?" is a mandatory field in ÜTS; "Yes" is selected if the device is a Self-Testing Device or a Near-Patient Testing Device, and this information must appear on the label and/or sales packaging and in the instructions for use.

A separate label also comes into play on the certificate side. Article 62, paragraph 5 of the Guide states that for Class B devices that are Near-Patient Testing Devices or Self-Testing Devices, the manufacturer, in addition to the standard procedures, follows the procedure specified in IVDR Annex IX (Chapter II, Section 5.1), which is named in ÜTS "Annex IX Chapter II (including Section 5) Technical Documentation Assessment". Paragraph 6(b) repeats the same rule for Class C, (c) adds the Annex IX (Chapter II, Section 5.2) procedure for each device in the case of Class C Companion Diagnostic Devices, and paragraph 8 sets up the same structure for Class D.

The document business rules are also separated according to these variants. Article 67, paragraph 6 of the Guide points to Annex 97 for Self-Testing Device (Class B) and Near-Patient Testing Device (Class B), paragraph 7 to Annex 98 for Self-Testing Device (Class C) and (Class D), paragraph 8 to Annex 99 for Near-Patient Testing Device (Class C) and (Class D), and paragraph 9 to Annex 100 for Companion Diagnostic Device (Class C) and (Class D).

ÜTS Fields That Change by Class

The number of fields you see on screen is not fixed; it changes with the framework under which you register and with the characteristics of the device. The MDR registration screen carries several fields that the former-regulation screen does not have.

  • Nanomaterial Type: According to Article 27, paragraph 7, subparagraph (ğ) of the Guide, if the field "Does it contain nanomaterial?" is answered "Yes", this field appears and the appropriate one of "Particle", "Aggregate", "Agglomerate" and "Unknown" is selected.

  • Medicinal product derived from human blood or plasma: The field in subparagraph (ü) of the same paragraph is mandatory; according to sub-item 5, if the device contains such a product, the device is considered Class III.

  • Animal-origin tissues and cells: The field in subparagraph (y) of the same paragraph asks about contents covered by Regulation (EU) No 722/2012 and the Regulation on Medical Devices Manufactured Utilising Tissues of Animal Origin; according to sub-item 4, these devices are Class III, except those intended to come into contact only with intact skin.

The IVDR screen, on the other hand, brings fields not found under MDR. Article 30, paragraph 7, subparagraph (p) of the Guide makes the field "Does it contain cells or substances of microbial origin?" mandatory. The field "Is it a new type of device?" in subparagraph (ş) of the same paragraph rests on a definition: a device is considered a "new device" if such a device for the relevant analyte has not been continuously available in Türkiye and EU member states during the last 3 years, or if the procedure involves an analytical technology not continuously in use during the last three years.

What Happens If the Class Is Selected Incorrectly?

The most visible consequence of a wrong class is that the documents linked to the device become invalid. Article 43, paragraph 11 of the Guide states that if, as a result of updating the "Class" data field in accordance with the provisions of the same article, the documents linked to the relevant device records become incompatible with the device class, a revision is issued to the devices with the relevant barcode.

This means that when the class is corrected, the document business rule also changes and the previous document combination is no longer accepted. Your device record falls into revision and stays there until the missing documents are completed. In practice, this means that determining the class correctly from the start is cheaper than correcting it afterwards, because a revision affects not a single record but all devices with the same barcode.

A Class Change Request Always Goes to Agency Approval

Because records belonging to different companies can exist in ÜTS under the same barcode, some data fields cannot be changed by the decision of a single company. Article 43, paragraph 4, subparagraph (d) of the Guide states that, if all the companies concerned approve the request, fields other than those requiring Agency assessment, such as "Class", "GMDN" and "EMDN", are deemed unanimously accepted and are corrected automatically; subparagraph (e) states that requests concerning these three fields are submitted to the Agency's assessment in "Under Review" status.

Paragraph 7 of the same article closes the subject without leaving room for doubt: data change requests concerning these fields are submitted to the Agency's assessment and approval in every case referred to in paragraphs 4, 5 and 6 of the article. The result does not change even if the other companies accept the request unanimously. Paragraph 10 states that if the request is found unsuitable, it will be rejected with its justification and will move to "Rejected" status in ÜTS.

Where Are the Transition Dates by Class?

The transition periods are also structured by risk class; there is no single date. Article 45, paragraph 3 of the Guide sets 26.05.2026 as the last date for placing on the market for Class III implantable custom-made devices under the MDD; paragraph 4 sets 31.12.2027 for Class III devices under the MDD or AIMDD and for Class IIb implantable devices other than sutures, staples, dental fillings, dental braces, tooth crowns, screws, wedges, plates, wires, pins, clips and connectors; and paragraph 5 sets 31.12.2028 for Class IIb devices other than those covered by paragraph 4, and for Class IIa, Class Im and Class Is devices.

For IVDs, the timetable is in Article 48 of the Guide: paragraph 3 sets 31.12.2027 for Annex II List A, Annex II List B and Self-Testing Devices under IVDD; paragraph 4 sets, in subparagraph (a), 31.12.2027 for Class D and its variants, in subparagraph (b), 31.12.2028 for Class C and its variants, and in subparagraph (c), 31.12.2029 for Class B, Near-Patient Testing Device (Class B) and Class A Sterile. All these dates are subject to the conditions stated in the relevant Regulations being met, and they are the last dates for placing on the market that the legislation grants to devices; they do not show how long a procedure will take.

How Do I Determine My Device's Class?

Classification looks not at what the device is, but at what it is used for. Article 52, paragraph 1 of the Guide states that the application of classification rules for all Medical Device Regulations is based on the intended purpose of the devices. Paragraph 2 of the same article states that if a device is intended to be used in combination with another device, the rules apply to each device separately, and that accessories of a medical device are classified in their own right, separately from the device with which they are used. Paragraph 3 states that software which drives a device or influences the use of a device falls within the same class as the device, and that if the software is independent of any device, it is classified in its own right.

If more than one rule can apply at the same time, the decision is easy. Paragraph 5 of the same article states that if several rules, or several sub-rules within the same rule, apply to the same device based on its intended purpose, the strictest rule and sub-rule resulting in the higher classification apply. Paragraph 4 states that if the device is not intended to be used solely or principally in a specific part of the body, it is classified on the basis of the most critical specified use. On the IVD side, paragraph 8 adds that accessories are classified in their own right, paragraph 9 that calibrators fall in the same class as the device, and paragraph 10 that control materials with assigned values fall in the same class as the device.

For the full text of the rules, you can see our medical device classification rules page, and you can use the classification wizard in our MDR risk class article to see the estimated class of your device through questions.

Class-Based Registration Checklist

Once the class decision is made, the rest of the registration opens in a particular order. The list below shows at a glance the chain set out by the provisions cited above.

  • Framework first: The registration screen in Article 25, 27 or 30 of the Guide is selected according to the regulation under which the device is certified; the class list changes with this selection.

  • Then the class: The one that fits the device is selected among ten options in the former regulations, seven in MDR and thirteen in IVDR.

  • Then the documents: Documents can only be selected after the class is selected; the document business rule comes from the annexes in Articles 63 to 67.

  • Instructions-for-use decision: Under paragraph 2 of Article 17, instructions are mandatory for Class III, Class IIb, Class IIa with instructions, IVDR Class B/C/D and variants, and the IVDD Annex II lists and Self-Testing Devices.

  • Class-specific fields: Under MDR, the nanomaterial type, human blood/plasma derivative and animal-origin tissue fields; under IVDR, the microbial-origin substance field, "Is it a new type of device?" and the self-testing/near-patient testing fields are completed.

  • Application type: The correct one is selected from the seven application types in Article 40, paragraph 3; "Class 3 Products Application" is used for Class III and "Active Implantable Medical Devices Application" for AIMDD devices.

  • If a class correction is needed: An imported product data change request is opened under Articles 43 and 44, and the request goes to Agency approval in every case.

The class decision itself is an engineering and regulatory matter; the single line entered in ÜTS is only the result of that decision. Companies can also handle this work by opening a user authorisation for a consultant from their own ÜTS company page. With the "Medical Device Registration Officer" authorisation, the consultant records and confirms the class, the document linkage and the class-specific fields; the person who signs the application electronically remains the company's authorised signatory, and responsibility for the accuracy of the data rests with the company.

The screen-by-screen flow of the registration is covered on the ÜTS Product (Medical Device) Registration page, and the full set of transition dates on the ÜTS Registration in the MDR and IVDR Transition page. All 22 MDR and 7 IVDR classification rules are on the Medical Device Classification Rules page, conformity assessment routes on the Conformity Assessment Procedures and Notified Body Routes page, and the mandatory-document matrix of document types on the ÜTS Document Types page.

For the EU-side counterpart, our MDR classification tool page (in English) provides guidance.

ÜTS Registration

Medical Device ÜTS Registration Consultancy

Request an ÜTS registration consultancy quote

Ehlibeyt Mahallesi, Tekstilciler Caddesi No: 35/7 06520 Çankaya / Ankara - TÜRKİYE

Cevizli Mahallesi, Zuhal Caddesi Ritim İstanbul B Blok, No:44 Maltepe / İstanbul - TÜRKİYE

info@medexmedikal.com
+90 505 580 4622
+90 312 389 1354

ÜTS Guides · Legislation Calendar · Glossary · ÜTS Guide Index · Frequently Asked Questions

Services: ÜTS Consultancy · Who Carries Out the ÜTS Registration? · ÜTS Document Registration Consultancy · ÜTS Product Registration Consultancy · Importer ÜTS Registration Consultancy · Sales Centre Authorisation Certificate · EUDAMED Registration Consultancy · MDR/IVDR Declaration of Conformity · Technical File Preliminary Review · Label and Instructions for Use Check · PRRC Service · PMS and PSUR · Clinical Evaluation · EMDN and GMDN Code Determination · Certificate of Free Sale · EUDAMED XML Generator · ÜTS Registration Training · Correction of a Rejected ÜTS Application

KVKK and Privacy Policy | Cookie Policy | Accessibility Statement

© 2026 Medex Kurumsal Danışmanlık Ltd. Şti.

bottom of page
WhatsApp Hemen Ara